Abstract


Abstract This study constitutes a feasibility assessment of dynamic conformal arc (DCA) therapy as an alternative to volumetric‐modulated arc therapy (VMAT) for stereotactic body radiation therapy (SBRT) of lung cancer. The rationale for DCA is lower geometric complexity and hence reduced risk for interplay errors induced by respiratory motion. Forward planned DCA and inverse planned DCA based on segment‐weight optimization were compared to VMAT for single arc treatments of five lung patients. Analysis of dose‐volume histograms and clinical goal fulfillment revealed that DCA can generate satisfactory and near equivalent dosimetric quality to VMAT, except for complex tumor geometries. Segment‐weight optimized DCA provided spatial dose distributions qualitatively similar to those for VMAT. Our results show that DCA, and particularly segment‐weight optimized DCA, may be an attractive alternative to VMAT for lung SBRT treatments if the patient anatomy is favorable.