Patient body motion correction for dynamic cardiac PET‐CT by attenuation‐emission alignment according to projection consistency conditions

 
 
 
 

Abstract


INTRODUCTION\nPatient body motion is known to cause large deviations in the determination of myocardial blood flow (MBF) with errors exceeding 300%. Accurate correction for patient whole-body motion is still a largely unsolved problem in cardiac positron emission tomography (PET) imaging.\n\n\nOBJECTIVE\nThis study evaluated the efficacy of using Natterer s formulation of the Helgason-Ludwig consistency conditions on the two-dimensional Radon transform to align computed tomography to PET projection data in multiple time frames of a dynamic sequence for the purpose of frame-by-frame correction of rigid whole-body motion.\n\n\nMETHODS\nThe correction algorithm was evaluated with digital NCAT phantoms using realistic noise added by the analytical simulator. Count rates used in the simulation were derived from clinical patient data. In addition, a proof of concept test using measured data with a cardiac torso phantom was conducted.\n\n\nRESULTS\nMotion correction resulted in significant improvement in the accuracy of MBF estimates, especially for high count-rate acquisitions. Maximum errors for 2\xa0cm of motion dropped from 325% to 25% and from 250% to 25% using global and regional partial-volume correction, respectively. Median MBF errors dropped from 33% to 4.5% and 27% to 3.8%, respectively. Importantly, the correction algorithm performed equally well to compensate for body motion in both early and late time frames.\n\n\nCONCLUSION\nCardiac PET-CT data used for attenuation correction (CTAC) alignment using projection consistency conditions was effective for reducing errors in MBF measurements due to simulated patient motion, and can be integrated into the image reconstruction workflow.

Volume 46
Pages 1697–1706
DOI 10.1002/mp.13419
Language English
Journal Medical Physics

Full Text