Nuclear Instruments & Methods in Physics Research Section A-accelerators Spectrometers Detectors and Associated Equipment | 2021

Performance of the Secondary Electrons Profile Monitor for the scanning gantry in the Heavy Ion Medical Machine

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Abstract


Abstract A Secondary Electrons Profile Monitor (SEPM) is developed as part of the beam dose delivery system in a clinical gantry of the Heavy Ion Medical Machine (HIMM). With the purpose of less disturbance to the beam, a thin foil with a thickness of 0 . 7 μ m and an accelerating grid with a diameter of 15 μ m are employed to produce and accelerate the secondary electrons, and two microchannel plates (MCPs) are used to multiply the electrons. Providing the real-time information of beam size, position, flatness, and symmetry in front of the scanning magnet, the beam monitor features an active area of 989 mm 2 , beam intensities between 1.0 × 1 0 3 to 1.0 × 1 0 8 particles per pulse (ppp) and a wide range of beam energies. A prototype is tested at the horizontal treatment room of the HIMM. Using different beam positions, sizes, and energies of 12C 6 + beams, test experiments show that the beam monitor maintains good consistency with a Multi-Strip Ionization Chamber (MSIC). Compared to the MSIC, the SEPM shows a better signal-to-noise ratio (SNR) and higher sensitivity to the variation of the beam.

Volume 999
Pages 165212
DOI 10.1016/J.NIMA.2021.165212
Language English
Journal Nuclear Instruments & Methods in Physics Research Section A-accelerators Spectrometers Detectors and Associated Equipment

Full Text