Radiation Measurements | 2019

Assessing 131I in thyroid by non-spectroscopic instruments - A European intercomparison exercise

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Abstract


Abstract One of the issues of the Open Project for the European Radiation Research Area (OPERRA) was human thyroid monitoring in case of a large scale nuclear accident. This issue was covered in task 5.4 as project “CaThyMARA” (Child and Adult Thyroid Monitoring After Reactor Accident), which included several aspects of thyroid monitoring, e.g. screening of facilities able to perform thyroid monitoring in the European countries, dose estimation, modelling of detector response, and two intercomparison exercises. The intercomparison described in this paper focused on thyroid monitoring by non-spectrometric instruments, including gamma cameras and other instruments that were considered available for measurements made by members of the public. A total of 12 facilities from 7 European countries have participated and 43 various measuring devices have been evaluated. The main conclusion of this intercomparison is that the ability to make assessments of 131I activity in the thyroid to the exposed population after an accidental release must, on the average, be considered as good among the European laboratories taking part in this study. This intercomparison also gave the participants the possibility to calibrate the measuring devices for thyroid measurements of children where this procedure was not available before. A comprehensive report of the intercomparison is given.

Volume 128
Pages 106115
DOI 10.1016/J.RADMEAS.2019.04.018
Language English
Journal Radiation Measurements

Full Text