American journal of surgery | 2021

Comparing intraoperative parathyroid identification based on surgeon experience versus near infrared autofluorescence detection - A surgeon-blinded multi-centric study.

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Abstract


BACKGROUND\nNear infrared autofluorescence (NIRAF) detection has previously demonstrated significant potential for real-time parathyroid gland identification. However, the performance of a NIRAF detection device - PTeye® - remains to be evaluated relative to a surgeon s own ability to identify parathyroid glands.\n\n\nMETHODS\nPatients eligible for thyroidectomy and/or parathyroidectomy were enrolled under 6 endocrine surgeons at 3 high-volume institutions. Participating surgeons were categorized based on years of experience. All surgeons were blinded to output of PTeye® when identifying tissues. The surgeon s performance for parathyroid discrimination was then compared with PTeye®. Histology served as gold standard for excised specimens, while expert surgeon s opinion was used to validate in-situ tissues.\n\n\nRESULTS\nPTeye® achieved 92.7% accuracy across 167 patients recruited. Junior surgeons (<5 years of experience) were found to have lower confidence in parathyroid identification and higher tissue misclassification rate per specimen when compared to PTeye® and senior surgeons (>10 years of experience).\n\n\nCONCLUSIONS\nNIRAF detection with PTeye® can be a valuable intraoperative adjunct technology to aid in parathyroid identification for surgeons.

Volume None
Pages None
DOI 10.1016/j.amjsurg.2021.05.001
Language English
Journal American journal of surgery