Applied ergonomics | 2021

Usability of smart infusion pumps: A heuristic evaluation.

 
 
 
 
 
 

Abstract


Although smart infusion pumps were built to eliminate medication errors, new types of usability errors have arisen. The purposes of this study were to determine potential risks when using smart pumps during secondary medication administration and to identify opportunities for design improvements. We observed and analyzed nurses when they interacted with smart pumps and heuristically evaluated the smart pump to identify usability problems. Forty-three usability problems were identified with the smart pump. The usability problems have the potential to create high cognitive burden on nurses and to increase the likelihood of mistakes. We discuss design and process improvement recommendations for each major finding from this study.

Volume 98
Pages \n 103584\n
DOI 10.1016/j.apergo.2021.103584
Language English
Journal Applied ergonomics