Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine | 2021

A kinetic model for the thermoluminescent high dose response of LiF:Mg,Cu,P (MCP-N).

 
 
 

Abstract


This contribution describes a kinetic model attempting to reproduce the response of the thermoluminescent material LiF:Mg,Cu,P when it is irradiated to absorbed dose values in the kGy range. The modelling is based on the hypothesis of a relationship between the irradiation time (i.e. the absorbed dose) and the density of trapping/recombination centres. X-ray diffraction and thermal X-ray diffraction measurements have been performed to investigate the potential radiation and thermal damage on the structure of the material, including the possibility of partial phases. The proposed kinetic model qualitatively reproduces the observed changes in the TL glow curve for temperatures above the main peak as well as the two observed regions of absorbed dose response: linear and sub-linear.

Volume 170
Pages \n 109634\n
DOI 10.1016/j.apradiso.2021.109634
Language English
Journal Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine