Talanta | 2019

Discrete sampling based-flow injection as an introduction system in ICP-MS for the direct analysis of low volume human serum samples.

 
 
 
 

Abstract


Discrete sampling based on the flow injection options offered by advanced autosamplers has been tested and applied for the direct analysis of low volume samples (human serum) for multi-elemental purposes (simultaneous Al, Be, Ca, Cd, Co, Cr, Cu, Fe, K, Li, Mg, Mn, Mo, Ni, Pb, Rb, Se, V, and Zn assessment). Serum samples (200\u202fµL aliquots) were directly diluted to 2.0\u202fmL with 1% (v/v) HNO3, and discrete sampling operating conditions were optimized to allow cleaning, loading (300\u202fµL injection loop), and measuring steps by using a volume lower than 2.0\u202fmL (diluted serum sample). Matrix effect has been overcome by using the standard addition technique, and accurate results implied the use of kinetic energy discrimination (KED) mode (He as a collision gas) for measurements. The proposed method has been found to be reliable for serum samples analysis because of the low sample volume requirements, the minimal sample preparation required, and high sampling rate (each replicate analysis takes 2.50\u202fmin). In addition, validation results show good precision and accuracy (both analytical recovery and analysis of certified reference materials), and acceptable sensitivity. Applicability of the proposed method has been finally demonstrated by analysing several serum samples from healthy adults.

Volume 199
Pages \n 220-227\n
DOI 10.1016/j.talanta.2019.02.050
Language English
Journal Talanta