Nature Communications | 2021

Photon quantum entanglement in the MeV regime and its application in PET imaging

 
 
 
 
 
 
 
 
 

Abstract


Positron Emission Tomography (PET) is a widely-used imaging modality for medical research and clinical diagnosis. Imaging of the radiotracer is obtained from the detected hit positions of the two positron annihilation photons in a detector array. The image is degraded by backgrounds from random coincidences and in-patient scatter events which require correction. In addition to the geometric information, the two annihilation photons are predicted to be produced in a quantum-entangled state, resulting in enhanced correlations between their subsequent interaction processes. To explore this, the predicted entanglement in linear polarisation for the two photons was incorporated into a simulation and tested by comparison with experimental data from a cadmium zinc telluride (CZT) PET demonstrator apparatus. Adapted apparati also enabled correlation measurements where one of the photons had undergone a prior scatter process. We show that the entangled simulation describes the measured correlations and, through simulation of a larger preclinical PET scanner, illustrate a simple method to quantify and remove the unwanted backgrounds in PET using the quantum entanglement information alone. Gamma photons used in positron emission tomography are predicted to be produced in an entangled state. Here, the authors simulate the effects of entanglement and test them through comparison with experimental data from a PET demonstrator apparatus, showing the potential gains in background suppression.

Volume 12
Pages None
DOI 10.1038/s41467-021-22907-5
Language English
Journal Nature Communications