arXiv: Instrumentation and Detectors | 2019

Feasibility study of a compact Neutron Resonance Transmission Analysis instrument.

 
 
 

Abstract


Neutron Resonance Transmission Analysis (NRTA) uses resonant absorption of neutrons to infer the absolute isotopic composition of a target object, enabling applications in a broad range of fields such as archaeology, materials analysis of nuclear fuel, and arms control treaty verification. In the past, NRTA involved large user facilities and complex detector systems. However, recent advances in the intensity of compact neutron sources have made compact neutron imaging designs increasingly feasible. This work describes the Monte Carlo (MC) based design of a compact epithermal NRTA radiographic instrument which uses a moderated, compact deuterium-tritium (DT) neutron source and an epithermal neutron detector. Such an instrument would have a wide range of applications, and would be especially impactful for such scenarios as nuclear inspection and arms control verification exercises, where system cost and mobility may be of critical importance. The MC simulations presented in this work demonstrate accurate time-of-flight (TOF) reconstructions for transmitted neutron energies, capable of differentiating isotopic compositions of nuclear material with high levels of accuracy. A new generation of miniaturized and increasingly more intense neutron sources will allow this technique to achieve measurements with greater precision and speed, with significant impact on a variety of engineering and societal problems.

Volume None
Pages None
DOI 10.1063/1.5129961
Language English
Journal arXiv: Instrumentation and Detectors

Full Text