The Review of scientific instruments | 2021

Performance of microchannel plate based detectors for <25 keV x rays: Monte Carlo simulations and comparisons with experimental results.

 
 
 

Abstract


We present the results of Monte Carlo simulations of the microchannel plate (MCP) response to x rays in the 250\xa0eV to 25 keV energy range as a function of both x-ray energy and impact angle and their comparisons with the experimental results from the X8A beamline at the National Synchrotron Light Source at Brookhaven National Laboratory. Incoming x rays interact with the lead glass of the microchannel plate, producing photoelectrons. Transport of the photoelectrons is neglected in this model, and it is assumed that photoelectrons deposit all their energy at the point they are created. This deposition leads to the generation of many secondary electrons, some fraction of which diffuse to the MCP pore surface where they can initiate secondary electron cascades in the pore under an external voltage bias. X-ray penetration through multiple MCP pore walls is increasingly important above 5 keV, and the effect of this penetration on MCP performance is studied. In agreement with past measurements, we find that the dependence of MCP sensitivity with angle relative to the pore bias changes from a cotangent dependence to angular independence and then proceeds to a secant dependence as the x-ray energy increases. We also find that with the increasing x-ray energy, the MCP gain sensitivity as a function of bias voltage decreases. The simulations also demonstrate that for x rays incident normal to the MCP surface, spatial resolution shows little dependence on the x-ray energy but degrades with the increasing x-ray energy as the angle of incidence relative to the surface normal increases. This agrees with experimental measurements. Simulation studies have also been completed for MCPs gated with a subnanosecond voltage pulse. We find that the optical gate profile width increases as the x-ray energy is increased above 5 keV, a consequence of increased x-ray penetration at energies >5 keV. Simulations of the pulsed dynamic range show that the dynamic range varies between ∼100 and 1000 depending on x-ray energy and peak voltage.

Volume 92 4
Pages \n 043101\n
DOI 10.1063/5.0030514
Language English
Journal The Review of scientific instruments

Full Text