A size-specific effective dose for patients undergoing CT examinations

 
 
 
 
 
 

Abstract


This study aims to develop a simple method for estimating the size-specific effective dose using the ImPACT software. The size-specific effective dose was calculated from the images of patients who underwent CT examinations of the thorax and abdomen regions. Volume computed tomography dose index (CTDIvol) and mAs data were obtained from the dose report. The average of mAs and CTDIvol were used to determine the normalized CTDIvol (nCTDIvol). Patient size was expressed in effective diameter (Deff), and was measured at nine slices along the z-axis. The normalized size-specific dose estimate (nSSDE) was then calculated. The normalized size-specific effective dose was obtained in the effective dose text-box by inputting an arbitrary value in the CTDI (air) text-box so that the CTDIvol text-box showed a similar value to the nSSDE. The results show that the normalized size-specific effective dose decreases exponentially with increasing patient size. The results are consistent with results of Sahbaee formula to within 20%.

Volume 1204
Pages 12002
DOI 10.1088/1742-6596/1204/1/012002
Language English
Journal None