The Journal of Applied Laboratory Medicine | 2021

Implementation of an Extraction-Free COVID Real-Time PCR Workflow in a Pediatric Hospital Setting

 
 
 
 
 
 

Abstract


Abstract Background This study outlines the development, implementation, and impact of a laboratory-developed, extraction-free real-time PCR assay as the primary diagnostic test for severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) in a pediatric hospital. Methods Clinical specimens from both upper and lower respiratory tract sources were validated, including nasopharyngeal aspirates, nasopharyngeal swabs, anterior nares swabs, and tracheal aspirates (n\u2009=\u2009333 clinical samples). Testing volumes and laboratory turnaround times were then compared before and after implementation to investigate effects of the workflow changes. Results Compared to magnetic-bead extraction platforms, extraction-free real-time PCR demonstrated ≥95% positive agreement and ≥97% negative agreement across all tested sources. Implementation of this workflow reduced laboratory turnaround time from an average of 8.8 (+/−5.5) h to 3.6 (+/−1.3) h despite increasing testing volumes (from 1515 to 4884 tests per week over the reported period of testing). Conclusions The extraction-free workflow reduced extraction reagent cost for SARS-CoV-2 testing by 97%, shortened sample handling time, and significantly alleviated supply chain scarcities due to the elimination of specialized extraction reagents for routine testing. Overall, this assay is a viable option for laboratories to increase efficiency and navigate reagent shortages for SARS-CoV-2 diagnostic testing.

Volume None
Pages None
DOI 10.1093/jalm/jfab079
Language English
Journal The Journal of Applied Laboratory Medicine

Full Text