RADIATION EXPOSURE TO THE BACK WITH DIFFERENT TYPES OF APRONS.

 
 
 

Abstract


Physicians and nurses stand with their back towards the C-arm fluoroscope when using the computer, taking things out of closets and preparing drugs for injection or instruments for intervention. This study was conducted to investigate the relationship between the type of lead apron and radiation exposure to the backs of physicians and nurses while using C-arm fluoroscopy. We compared radiation exposure to the back in the three groups: no lead apron (group C), front coverage type (group F) and wrap-around type (group W). The other wrap-around type apron was put on the bed instead of on a patient. We ran C-arm fluoroscopy 40 times for each measurement. We collected the air kerma (AK), exposure time (ET) and effective dose (ED) of the bedside table, upper part and lower part of apron. We measured these variables 30 times for each location. In group F, ED of the upper part was the highest (p\xa0<\xa00.001). ED of the lower part in group C and F was higher than that in group W (p\xa0=\xa00.012). The radiation exposure with a front coverage type apron is higher than that of the wrap-around type and even no apron at the neck or thyroid. For reducing radiation exposure to the back of physician or nurse, the wrap-around type apron is recommended. This type of apron can reduce radiation to the back when the physician turns away from the patient or C-arm fluoroscopy.