Radiation protection dosimetry | 2021

The Potential to Use Tld Measurements to Validate the Occupational Radiation Protection at the Department of Nuclear Medicine.

 
 
 
 

Abstract


Diagnostic examinations in nuclear medicine increase. This can cause a higher radiation burden to the personnel. The aim of this study was to create and apply a method to validate the occupational radiation protection at the Department of Nuclear Medicine using thermoluminescence dosemeters (TLD). Occupational doses in- and outside of the department and the dose rate distribution around the positron emission tomography/computed tomography (PET/CT) were validated in terms of Hp(10) using TLD-100. Backscatter correction factors were estimated to implement the TLD measurements on different materials. After correction for backscatter, the result indicates that doses in- and outside of the department are below recommended values and dose limits. The highest and the lowest dose rate were measured beside the 18F-autoinjector and the PET/CT gantry, respectively. The method of using TLD gives a good estimate of how the occupational radiation protection performs at the Department of Nuclear Medicine and can thus help in the optimization of radiation protection.

Volume None
Pages None
DOI 10.1093/rpd/ncab085
Language English
Journal Radiation protection dosimetry