Journal of synchrotron radiation | 2019

CT dosimetry at the Australian Synchrotron for 25-100\u2005keV photons and 35-160\u2005mm-diameter biological specimens.

 
 
 
 

Abstract


The dose length product (DLP) method for medical computed tomography (CT) dosimetry is applied on the Australian Synchrotron Imaging and Medical Beamline (IMBL). Beam quality is assessed from copper transmission measurements using image receptors, finding near 100% (20\u2005keV), 3.3% (25\u2005keV) and 0.5% (30-40\u2005keV) relative contributions from third-harmonic radiation. The flat-panel-array medical image receptor is found to have a non-linear dose response curve. The amount of radiation delivered during an axial CT scan is measured as the dose in air alone, and inside cylindrical PMMA phantoms with diameters 35-160\u2005mm for mono-energetic radiation 25-100\u2005keV. The radiation output rate for the IMBL is comparable with that used for medical CT. Results are presented as the ratios of CT dose indices (CTDI) inside phantoms to in air with no phantom. Ratios are compared for the IMBL against medical CT where bow-tie filters shape the beam profile to reduce the absorbed dose to surface organs. CTDI ratios scale measurements in air to estimate the volumetric CTDI representing the average dose per unit length, and the dose length product representing the absorbed dose to the scanned volume. Medical CT dose calculators use the DLP, beam quality, axial collimation and helical pitch to estimate organ doses and the effective dose. The effective dose per unit DLP for medical CT is presented as a function of body region, beam energy and sample sizes from neonate to adult.

Volume 26 Pt 2
Pages \n 517-527\n
DOI 10.1107/S1600577518018015
Language English
Journal Journal of synchrotron radiation

Full Text