Journal of synchrotron radiation | 2019

The U125 insertion device beamline at the Metrology Light Source.

 
 
 

Abstract


At the Metrology Light Source, an electron storage ring dedicated to metrological applications, the U125 insertion device beamline utilizes undulator radiation for various applications over a broad spectral range. Using a hybrid normal-incidence and grazing-incidence in-vacuum switchable plane-grating monochromator, a spectral region ranging from the near-infrared to soft X-ray is covered. The beamline is dedicated to surface-analytical methods, e.g. ellipsometry, photoelectron spectroscopy or photoemission tomography. The traceability of radiometric quantities, i.e. quantitative determination of the available radiant power (or photon flux), is required for some of these applications to support the metrological aspect of the measurements. In particular, attention is paid to the suppression of unwanted spectral contributions from higher diffraction orders, and to the monitoring of the radiation intensity during the measurements. With the results from the beamline commissioning, an uncertainty budget for all relevant radiometric quantities was\xa0established.

Volume 26 Pt 2
Pages \n 535-542\n
DOI 10.1107/S1600577518018428
Language English
Journal Journal of synchrotron radiation

Full Text