Journal of synchrotron radiation | 2021

Exploiting spatio-spectral aberrations for rapid synchrotron infrared imaging.

 
 
 
 
 
 
 
 
 

Abstract


The Infrared Microspectroscopy Beamline at the Australian Synchrotron is equipped with a Fourier transform infrared (FTIR) spectrometer, which is coupled with an infrared (IR) microscope and a choice of two detectors: a single-point narrow-band mercury cadmium telluride (MCT) detector and a 64 × 64 multi-pixel focal plane array (FPA) imaging detector. A scanning-based point-by-point mapping method is commonly used with a tightly focused synchrotron IR beam at the sample plane, using an MCT detector and a matching 36× IR reflecting objective and condenser (NA = 0.5), which is time consuming. In this study, the beam size at the sample plane was increased using a 15× objective and the spatio-spectral aberrations were investigated. A correlation-based semi-synthetic computational optical approach was applied to assess the possibilities of exploiting the aberrations to perform rapid imaging rather than a mapping approach.

Volume 28 Pt 5
Pages \n 1616-1619\n
DOI 10.1107/S1600577521007104
Language English
Journal Journal of synchrotron radiation

Full Text