Performance of popular pulse oximeters compared with simultaneous arterial oxygen saturation or clinical-grade pulse oximetry: a cross-sectional validation study in intensive care patients

 
 
 
 
 
 
 

Abstract


Objectives To evaluate the performance of direct-to-consumer pulse oximeters under clinical conditions, with arterial blood gas measurement (SaO2) as reference standard. Design Cross-sectional, validation study. Setting Intensive care. Participants Adult patients requiring SaO2-monitoring. Interventions The studied oximeters are top-selling in Europe/USA (AFAC FS10D, AGPTEK FS10C, ANAPULSE ANP 100, Cocobear, Contec CMS50D1, HYLOGY MD-H37, Mommed YM101, PRCMISEMED F4PRO, PULOX PO-200 and Zacurate Pro Series 500\u2009DL). Directly after collection of a SaO2 blood sample, we obtained pulse oximeter readings (SpO2). SpO2-readings were performed in rotating order, blinded for SaO2 and completed <10\u2009min after blood sample collection. Outcome measures Bias (SpO2–SaO2) mean, root mean square difference (ARMS), mean absolute error (MAE) and accuracy in identifying hypoxaemia (SaO2 ≤90%). As a clinical index test, we included a hospital-grade SpO2-monitor (Philips). Results In 35 consecutive patients, we obtained 2258 SpO2-readings and 234 SaO2-samples. Mean bias ranged from −0.6 to −4.8. None of the pulse oximeters met ARMS ≤3%, the requirement set by International Organisation for Standardisation (ISO)-standards and required for Food and Drug Administration (FDA) 501(k)-clearance. The MAE ranged from 2.3 to 5.1, and five out of ten pulse oximeters met the requirement of ≤3%. For hypoxaemia, negative predictive values were 98%–99%. Positive predictive values ranged from 11% to 30%. Highest accuracy (95%\u2009CI) was found for Contec CMS50D1; 91% (86–94) and Zacurate Pro Series 500\u2009DL; 90% (85–94). The hospital-grade SpO2-monitor had an ARMS of 3.0% and MAE of 1.9, and an accuracy of 95% (91%–97%). Conclusion Top-selling, direct-to-consumer pulse oximeters can accurately rule out hypoxaemia, but do not meet ISO-standards required for FDA-clearance

Volume 8
Pages None
DOI 10.1136/bmjresp-2021-000939
Language English
Journal BMJ Open Respiratory Research

Full Text