Influence of Connector Diameter on Fracture Load of CAD/CAM-Processed Monolithic Lithium Disilicate Fixed Partial Dentures.

 
 
 
 
 

Abstract


PURPOSE\nTo investigate the influence of connector diameter on the mechanical load to fracture in monolithic three-unit lithium disilicate fixed partial dentures (FPDs).\n\n\nMATERIALS AND METHODS\nA total of 24 FPDs were designed and manufactured using computer-aided design/computer-assisted manufacturing (CAD/CAM) with connector diameters of 16 mm2, 12 mm2, or 9 mm2 (Groups A, B, and C, respectively; n = 8 for each group). After thermal and mechanical aging, the FPDs were subjected to mechanical load-to-fracture assessment.\n\n\nRESULTS\nFracture loads of Groups B (834 ± 105 N) and C (796 ± 41 N) were significantly lower compared to Group A (990 ± 65 N).\n\n\nCONCLUSION\nConnector dimensions proved to be crucial for fracture resistance of monolithic lithium disilicate FPDs.