The Journal of International Medical Research | 2019

Long-term outcome of first 300 implanted Absorb bioresorbable vascular scaffolds in an all-comers Middle East population

 
 
 
 
 
 
 
 
 

Abstract


Objectives To evaluate long-term clinical outcomes of the Absorb bioresorbable vascular scaffold (BVS) system (Abbott Vascular) in an all-comers Middle East population. Methods This prospective registry study included an initial set of patients with coronary lesions treated using Absorb BVS. Patients were followed for target vessel failure (TVF) including cardiac death, target vessel myocardial infarction (MI), and target lesion revascularization. Results A total of 217 patients (age, 55\u2009±\u200911 years; male, 169) with 300 treated lesions were included (median follow-up, 36 months [range, 26–41 months]; complete follow-up, 201 patients). Diabetes mellitus and acute coronary syndrome were present in 50% and 57% of patients, respectively. TVF rate was 32/201 (15.9%), including cardiac death in 10 (5%), target vessel MI in 13 (6.5%), and target lesion revascularization in 22 patients (10.9%). Definite or probable device thrombosis occurred in 11/201 patients (5.5%). TVF was associated with heart failure, worse ejection fraction, multi-vessel BVS, multi BVS in lesion, and total BVS length >50 mm. Conclusions Long-term outcome following Absorb BVS implantation in a population with high prevalence of high-risk and complex patients is acceptable, but heart failure, worse ejection fraction, and multi-vessel or long BVS implantation were associated with worse outcomes.

Volume 47
Pages 173 - 187
DOI 10.1177/0300060518798994
Language English
Journal The Journal of International Medical Research

Full Text