Technology in Cancer Research & Treatment | 2021

Influence of Modulation Factor on Treatment Plan Quality and Irradiation Time in Hippocampus-Sparing Whole-Brain Radiotherapy Using Tomotherapy

 
 
 
 
 

Abstract


Objectives: Hippocampus-sparing whole-brain radiotherapy (HS-WBRT) using tomotherapy is known to provide a better dose distribution than volumetric-modulated arc therapy but requires an extended irradiation time. The present study aimed to investigate whether irradiation time can be shortened by reducing the modulation factor (MF) without losing the target dose distribution. Methods: Using six tilted computed tomography images in the head area, the planning target volume (PTV) and hippocampal doses, and the irradiation time was investigated with a jaw width of 1\u2005cm, a pitch of 0.200, and the MF changed from 3.0 to 2.6, 2.2, 1.8, and 1.4. Results: No significant changes in the PTV or hippocampus were found with MF in the range from 3.0 to 1.8, but marked deterioration was found with that of 1.4. The irradiation time showed a linear relationship with the MF within the range from 3.0 to 1.8, with 1334, 1158, 986, and 817\u2005s at modulation factors of 3.0, 2.6, 2.2, and 1.8, respectively. However, when the MF was 1.4, the irradiation time was 808\u2005s. Conclusions: When HS-WBRT is performed with a tilted body position and a jaw width of 1\u2005cm, with a MF of 1.8, a favorable balance between dose parameters and irradiation time is achieved, whereas with a MF of 1.4, the quality of the radiotherapy plan deteriorates, and the irradiation time is approximately the same as that with a MF of 1.8.

Volume 20
Pages None
DOI 10.1177/15330338211045497
Language English
Journal Technology in Cancer Research & Treatment

Full Text