Journal of Cardiovascular Magnetic Resonance | 2021

Quantitative assessment of velocity and flow using compressed SENSE in children and young adults with adequate acquired temporal resolution

 
 
 
 

Abstract


Background Phase contrast (PC) cardiovascular magnetic resonance (CMR) imaging with parallel imaging acceleration is established and validated for measuring velocity and flow. However, additional acceleration to further shorten acquisition times would be beneficial in patients with complex vasculature who need multiple PC-CMR measurements, especially pediatric patients with higher heart rates. Methods PC-CMR images acquired with compressed sensitivity encoding (C-SENSE) factors of 3 to 6 and standard of care PC-CMR with sensitivity encoding (SENSE) factor of 2 (S2) acquired as part of clinical CMR examinations performed between November 2020 and January 2021 were analyzed retrospectively. The velocity and flow through the ascending aorta (AAo), descending aorta (DAo), and superior vena cava (SVC) in a transverse plane at the level of pulmonary artery bifurcation were compared. Additionally, frequency power distribution and dynamic time warp distance were calculated for these acquisitions. To further validate the adequate temporal resolution requirement, patients with S2 PC-CMR in the same acquisition plane were added in frequency power distribution analysis. Results Twenty-eight patients (25 males; 15.9\u2009±\u20091.9\xa0years; body surface area (BSA) 1.7\u2009±\u20090.2\xa0m 2 ; heart rate 81\u2009±\u200916\xa0bpm) underwent all five PC-CMR acquisitions during the study period. An additional 22 patients (16 males; 17.5\u2009±\u20097.7\xa0years; BSA 1.6\u2009±\u20090.5\xa0m 2 ; heart rate 91\u2009±\u200916\xa0bpm) were included for frequency power spectrum analysis. As expected, scan time decreased with increasing C-SENSE acceleration factor\u2009=\u20093 (37.5\u2009±\u20096.5\xa0s, 26.4\u2009±\u20097.6%), 4 (28.1\u2009±\u20094.9\xa0s, 44.7\u2009±\u20095.6%), 5 (21.6\u2009±\u20093.6\xa0s, 57.6\u2009±\u20094.4%), and 6 (19.1\u2009±\u20093.2\xa0s, 62.3\u2009±\u20094.2%) relative to SENSE\u2009=\u20092 (51.3\u2009±\u200910.1\xa0s) PC-CMR acquisition. Mean peak velocity, net flow, and cardiac output were comparable ( p \u2009>\u20090.87) between the five PC-CMR acquisitions with mean differences less than\u2009<\u20094%,\u2009<\u20092%, and\u2009<\u20093% respectively. All individual blood vessels showed a non-significant dependence of difference in f max99 (<\u20094\xa0Hz, p \u2009>\u20090.2), and dynamic time warp distance ( p \u2009>\u20090.3) on the C-SENSE acceleration factor used. There was a strongly correlated (r\u2009=\u20090.74) increase in f max99 (10.5\u2009±\u20092.2, range: 7.1–16.4\xa0Hz) with increasing heart rate. The computed minimum required cardiac phase number was 15\u2009±\u20092.0 (range: 11–20) over the heart rate of 86\u2009±\u200915\xa0bpm (range: 58–113\xa0bpm). Conclusions Stroke volume, cardiac output, and mean peak velocity measurements using PC-CMR with C-SENSE of up to 6 agree with measurements by standard of care PC-CMR with SENSE\u2009=\u20092 and resulted in up to a 65% reduction in acquisition time. Adequate temporal sampling can be ensured by acquiring 20 cardiac phases throughout the entire cardiac cycle over a wide range of pediatric and young adult heart rates.

Volume 23
Pages None
DOI 10.1186/s12968-021-00811-7
Language English
Journal Journal of Cardiovascular Magnetic Resonance

Full Text