EJNMMI Physics | 2021

Performance evaluation of a PET insert for preclinical MRI in stand-alone PET and simultaneous PET–MRI modes

 
 
 
 
 
 
 
 
 
 
 

Abstract


Background This study aimed to evaluate the performance of a preclinical PET insert in three configurations: as a stand-alone unit outside the MRI bore, inside the bore of a cryogen-free 3T MRI and, finally, while performing simultaneous PET/MRI studies. Methods The PET insert consists of two rings of six detectors, each detector comprising 8\u2009×\u200912 SiPMs reading out dual offset layers of pixelated LYSO crystals with a 1.4-mm pitch. The inner diameter is 60\xa0mm, transaxial field of view (FoV) 40\xa0mm and axial FoV 98\xa0mm. Evaluation was based on NEMA NU 4-2008 guidelines with appropriate modifications. Spatial resolution and sensitivity were measured inside and outside the MR bore. Image quality, count rate and quantitative performance were measured in all three configurations. The effect of temperature stability on PET sensitivity during fast spin echo sequences was also evaluated. B 0 field homogeneity and T1 and T2 relaxation times were measured using a water-filled phantom, with and without simultaneous PET operation. Finally, PET and MRI scans of a mouse injected with 10\xa0MBq [ 18 F]NaF and a mouse injected with 16\xa0MBq [ 18 F]FDG were performed in sequential and simultaneous modes. Results Peak absolute sensitivity was 10.15% with an energy window of 250–750\xa0keV. Absolute sensitivity values outside and inside the MR bore with MR idle agreed to within 0.1%. Outside the MR bore, spatial resolution was 1.21/1.59\xa0mm FWHM (radial/tangential) 5\xa0mm from the centre of the FoV which compared well with 1.19/1.26\xa0mm FWHM inside the MR bore. There were no substantial differences between all three scan configurations in terms of peak NEC rate (175 kcps at 17\xa0MBq), scatter or random fractions. Uniformity and recovery coefficients were also consistent between scanning modes. B 0 field homogeneity and T1 and T2 relaxation times were unaltered by the presence of the PET insert. No significant differences were observed between sequential and simultaneous scans of the animals. Conclusions We conclude that the performance of the PET insert and MRI system is not significantly affected by the scanning mode.

Volume 8
Pages None
DOI 10.1186/s40658-021-00415-1
Language English
Journal EJNMMI Physics

Full Text