JA Clinical Reports | 2021

Right ventricular perforation, pneumothorax, and a pneumatocele by a pacemaker lead: a case report

 
 
 
 
 
 

Abstract


Background Perforation of the right ventricle by a pacemaker lead is a rare and potentially life-threatening complication. We present a patient who developed right ventricular perforation, pneumothorax, and a cyst and underwent partial lung resection. Case presentation A 94-year-old woman was diagnosed with sick sinus syndrome and underwent a dual-chamber permanent pacemaker implantation. The next day, pacing failed and chest radiography showed that the right ventricular lead was outside the cardiac silhouette. Computed tomography revealed that the lead had perforated the right ventricular apex, causing a left-sided pneumothorax and a cystic lesion at the site of pulmonary injury by the pacemaker lead. The patient underwent lung resection and a right ventricular lead extraction. Pathological analysis revealed the cystic lesion to be an acute pneumatocele. Conclusions Pneumothorax and pneumatocele associated with right ventricular pacemaker lead perforation is extremely rare. In our case, a radical surgical intervention provided an excellent outcome.

Volume 7
Pages None
DOI 10.1186/s40981-021-00470-8
Language English
Journal JA Clinical Reports

Full Text