Dento maxillo facial radiology | 2019

Comparison of manual and dose reduction modes of a MORITA R100 CBCT.

 
 
 
 
 

Abstract


OBJECTIVE:\nTo compare effective dose differences when acquiring (1) dose reduction mode and (2) manual mode in a MORITA R100 CBCT.\n\n\nMETHODS:\n24 exposure protocols with different technique factors were performed in both the dose reduction mode and the manual mode in a Veraviewepocs 3D R100 cone beam CT device. 48 TLD were placed in a RANDO head phantom at 24 different sites. Effective doses were calculated according to the formalism published in the 103rd ICRP report.\n\n\nRESULTS:\nEffective doses for the dose reduction mode protocols ranged from 14 to 156 µSv [mean = 57 µSv, standard deviation (SD) = 37 µSv], whereas effective doses for the manual mode protocols ranged from 22 to 267 µSv (mean = 94 µSv, SD = 65 µSv). Thus, across all protocols, the dose reduction mode leads to a drop of the effective dose by 38 % (SD = 6 %).\n\n\nCONCLUSIONS:\nThe estimated effective doses are significantly lower if dose reduction protocols are acquired.

Volume 48 2
Pages \n 20180009\n
DOI 10.1259/dmfr.20180009
Language English
Journal Dento maxillo facial radiology