Biomedical optics express | 2019

Parallel detection of Jones-matrix elements in polarization-sensitive optical coherence tomography.

 
 
 
 

Abstract


The polarization properties of a sample can be characterized using a Jones matrix. To measure the Jones matrix without assumptions of the sample, two different incident states of polarization are usually used. This requirement often causes certain drawbacks in polarization-sensitive optical coherence tomography (PS-OCT), e.g., a decrease in the effective A-scan rate or axial depth range, if a multiplexing scheme is used. Because both the A-scan rate and axial depth range are important for clinical applications, including the imaging of an anterior eye segment, a new PS-OCT method that does not have these drawbacks is desired. Here, we present a parallel-detection approach that maintains the same A-scan rate and axial measurement range as conventional OCT. The interferometer consists of fiber-optic components, most of which are polarization-maintaining components with fast-axis blocking free from polarization management. When a parallel detection is implemented using swept-source OCT (SS-OCT), synchronization between the A-scans and synchronization between the detection channels have critical effects on the Jones-matrix measurement. Because it is difficult to achieve perfect synchronization using only hardware, we developed a solution using a numerical correction with signals from a static mirror. Using the developed system, we demonstrate the imaging of an anterior eye segment from the cornea to the back surface of the crystalline lens.

Volume 10 5
Pages \n 2318-2336\n
DOI 10.1364/BOE.10.002318
Language English
Journal Biomedical optics express