Biomedical Optics Express | 2021

Clinical integration of fast Raman spectroscopy for Mohs micrographic surgery of basal cell carcinoma

 
 
 
 
 
 
 
 
 
 
 
 
 

Abstract


We present the first clinical integration of a prototype device based on integrated auto-fluorescence imaging and Raman spectroscopy (Fast Raman device) for intra-operative assessment of surgical margins during Mohs micrographic surgery of basal cell carcinoma (BCC). Fresh skin specimens from 112 patients were used to optimise the tissue pre-processing and the Fast Raman algorithms to enable an analysis of complete Mohs layers within 30 minutes. The optimisation allowed >95% of the resection surface area to be investigated (including the deep and epidermal margins). The Fast Raman device was then used to analyse skin layers excised from the most relevant anatomical sites (nose, temple, eyelid, cheek, forehead, eyebrow and lip) and to detect the three main types of BCC (nodular, superficial and infiltrative). These results suggest that the Fast Raman technique is a promising tool to provide an objective diagnosis “tumour clear yes/no” during Mohs surgery of BCC. This clinical integration study is a key step towards a larger scale diagnosis test accuracy study to reliably determine the sensitivity and specificity in a clinical setting.

Volume 12
Pages 2015 - 2026
DOI 10.1364/BOE.417896
Language English
Journal Biomedical Optics Express

Full Text