PLoS ONE | 2019

High resolution continuous arterial spin labeling of human cerebral perfusion using a separate neck tagging RF coil

 
 
 

Abstract


For standard clinical applications, ASL images are typically acquired with 4–8 mm thick slices and 3–4 mm in-plane resolution. However, in this paper we demonstrate that high-resolution continuous arterial spin labeling (CASL) perfusion images can be acquired in a clinically relevant scan time using current MRI technology. CASL was implemented with a separate neck coil for labeling the arterial blood on a 4.7T MRI using standard axial 2D GE-EPI. Typical-resolution to high-resolution (voxels of 95, 60, 45, 27, or 7 mm3) images were compared for qualitative and quantitative cerebral blood flow analysis (CBF) in nine healthy volunteers (ages: 24–32 years). The highest resolution (1.5x1.5x3 = 7 mm3) CASL implementation yielded perfusion images with improved cortex depiction and increased cortical CBF measurements (53 ± 8 ml/100g/min), consistent with reduced partial volume averaging. The 7 mm3 voxel images were acquired with 6 cm brain coverage in a clinically relevant scan of 6 minutes. Improved spatial resolution facilitates CBF measurement with reduced partial volume averaging and may be valuable for the detection of perfusion deficits in small lesions and perfusion measurement in small brain regions.

Volume 14
Pages None
DOI 10.1371/journal.pone.0215998
Language English
Journal PLoS ONE

Full Text