Clinical Chemistry and Laboratory Medicine (CCLM) | 2021
Analytical and clinical performances of a SARS-CoV-2 S-RBD IgG assay: comparison with neutralization titers
Abstract
Abstract Objectives SARS-CoV-2 serology presents an important role in several aspects of COVID-19 pandemic. Immunoassays performances have to be accurately evaluated and correlated with neutralizing antibodies. We investigated the analytical and clinical performances of a SARS-CoV-2 RBD IgG assay, automated on a high throughput platform, and the correlation of the antibodies (Ab) levels with the plaque reduction neutralization (PRNT50) Ab titers. Methods A series of 546 samples were evaluated by SARS-CoV-2 RBD IgG assay (Snibe diagnostics), including 171 negative and 168 positive SARS-CoV-2 subjects and a further group of 207 subjects of the COVID-19 family clusters follow-up cohort. Results Assay imprecision ranged from 3.98 to 12.18% being satisfactory at low and medium levels; linearity was excellent in all the measurement range. Considering specimens collected after 14 days post symptoms onset, overall sensitivity and specificity were 99.0 and 92.5%, respectively. A total of 281 leftover samples results of the PRNT50 test were available. An elevated correlation was obtained between the SARS-CoV-2 RBD IgG assay and the PRNT50 titer at univariate (ρ=0.689) and multivariate (ρ=0.712) analyses. Conclusions SARS-CoV-2 S-RBD IgG assay shows satisfactory analytical and clinical performances, and a strong correlation with sera neutralizing activity.