GMS Hygiene and Infection Control | 2021

Preclinical evaluation of passive disinfection caps with a long-term catheter for the prevention of catheter-related bloodstream infection in pediatric cancer patients

 
 
 

Abstract


The use of passive disinfection devices (disinfection caps) may be a beneficial part of a maintenance care bundle, aiming at the prevention of catheter-related bloodstream infections in pediatric cancer patients. This preclinical in vitro investigation tested the visual and mechanical integrity of a Broviacâ„¢ catheter hub after simulation testing with 122 3Mâ„¢ CurosTM Stopper Disinfection Caps for Open Female Luers repeatedly attached and removed over 6 months. We found that these catheter hubs were compatible, fully operational, and airtight with use of 3M Curos stopper caps after 6 months of use with 122 caps per catheter hub.

Volume 16
Pages None
DOI 10.3205/dgkh000391
Language English
Journal GMS Hygiene and Infection Control