Micromachines | 2021

In Vivo Microelectrode Arrays for Detecting Multi-Region Epileptic Activities in the Hippocampus in the Latent Period of Rat Model of Temporal Lobe Epilepsy

 
 
 
 
 
 
 
 

Abstract


Temporal lobe epilepsy (TLE) is a form of refractory focal epilepsy, which includes a latent period and a chronic period. Microelectrode arrays capable of multi-region detection of neural activities are important for accurately identifying the epileptic focus and pathogenesis mechanism in the latent period of TLE. Here, we fabricated multi-shank MEAs to detect neural activities in the DG, hilus, CA3, and CA1 in the TLE rat model. In the latent period in TLE rats, seizures were induced and changes in neural activities were detected. The results showed that induced seizures spread from the hilus and CA3 to other areas. Furthermore, interneurons in the hilus and CA3 were more excited than principal cells and exhibited rhythmic oscillations at approximately 15 Hz in grand mal seizures. In addition, the power spectral density (PSD) of neural spikes and local field potentials (LFPs) were synchronized in the frequency domain of the alpha band (9–15 Hz) after the induction of seizures. The results suggest that fabricated MEAs have the advantages of simultaneous and precise detection of neural activities in multiple subregions of the hippocampus. Our MEAs promote the study of cellular mechanisms of TLE during the latent period, which provides an important basis for the diagnosis of the lesion focus of TLE.

Volume 12
Pages None
DOI 10.3390/mi12060659
Language English
Journal Micromachines