Abstract


This paper presents the broadband antenna for the microwave radiometric sensing of internal body temperature. For broadband operation, the bow-tie antenna was designed and backed with a cylindrical cavity, which decreased environmental electromagnetic interference and also improved the directivity of the antenna. The broadband impedance-transforming balun in microstrip form was also designed to feed the bow-tie antenna, and was located inside the cavity. An impedance-matching dielectric layer (IMDL) was introduced on top of the bow-tie antenna, for impedance match with the human body with high permittivity. The fabricated antenna was measured in free space with the IMDL removed, showing an input reflection coefficient lower than −10 dB from 2.64 to > 3.60 GHz with antenna gain over 6.0 dBi and radiation efficiency over 74.7% from 2.7 to 3.5 GHz. The IMDL was re-installed on the cavity-backed bow-tie antenna to measure the antenna performance for the human head with relative permittivity of about 40. The measured reflection coefficient was as low as −28.9 dB at 2.95 GHz and lower than −10 dB from 2.65 to > 3.5 GHz. It was also shown that the designed antenna recovered a good impedance match by adjusting the permittivity and thickness of the IMDL for the different parts of the human body with different permittivities.