Asian Journal of Pharmaceutical Analysis | 2021

Validated RP-HPLC method for the estimation of Amiloride and hydrochlorothiazide in combined tablet dosage form

 
 
 
 

Abstract


The aim of this work is to build up a fast, exact, precise and touchy reverse phase liquid chromatographic method for the synchronous assessment of amiloride and hydrochlorothiazide in tablet dose structure. The chromatographic strategy was normalized utilizing Hypersil ODS segment (250×4.6mm, 5μm molecule size) with UV discovery at 210nm and stream pace of 1ml/min. The portable stage includes phosphate buffer (pH acclimated to 2.5 with dilute Ortho Phosphoric acid) and acetonitrile in the proportion of 60:40 v/v. The linearity of proposed technique was examined in the scope of 5-30μg/ml (R²=0.999) for amiloride and 50-300μg/ml (R²=0.999) for Hydrochlorothiazide appropriately. The limit of detection (LOD) was discovered to be 0.10μg/ml and 0.40μg/ml for Amiloride and Hydrochlorothiazide appropriately. The limit of quantitation (LOQ) was discovered to be 0.30μg/ml and 1.20μg/ml for Amiloride and Hydrochlorothiazide separately. The retention times of Amiloride and Hydrochlorothiazide were found to be 3.258min and 2.383min separately. The technique was truly recommended and %RSD was found to be under 2 demonstrating serious level of exactness and accuracy. Subsequently proposed strategy can be effectively evaluated for the synchronous assessment of Amiloride and Hydrochlorothiazide in promoted formulations.

Volume None
Pages None
DOI 10.52711/2231-5675.2021.00037
Language English
Journal Asian Journal of Pharmaceutical Analysis

Full Text