[Determination of Bone SPECT Image Reconstruction Conditions in the Head and Neck Region].

 
 
 
 
 
 

Abstract


PURPOSE\nQuantitative analysis using a standardized uptake value (SUV) has become possible for single-photon emission computed tomography-computed tomography (SPECT-CT) of bone. However, previous research was targeted to the trunk area, and there are few studies for the head and neck region. Therefore, the purpose of this study was to determine the optimal image reconstruction conditions for bone SPECT of the head and neck using a phantom study.\n\n\nMETHOD\nThe radioactivity concentration of the 99mTc solution enclosed in the cylindrical phantom was set to the same count rate as in clinical cases, and six hot spheres (10, 13, 17, 22, 28, 37 mm) with four times the concentration were placed within it. The image reconstruction was 3D-OSEM, and the reconstruction conditions were varied by the number of iterative updates and the width of the Gaussian filter. Quantitative evaluations of the image quality were performed using the % contrast, background variability, and SUV for the hot spheres and background. A visual evaluation was performed by four observers to determine the optimal image reconstruction conditions for bone SPECT of the head and neck region.\n\n\nRESULT\nThe concentration of the 99mTc solution enclosed in the phantom was 6.95 (kBq/ml). Based on the results of the quantitative and visual evaluations, the optimal image reconstruction conditions were iterative updates=60 (subset: 10, iteration: 6) and a Gaussian filter of 7.8 mm.\n\n\nCONCLUSION\nThe optimal image reconstruction conditions were subset=10, iterations=6, and a Gaussian filter of 7.8 mm.

Volume 77 7
Pages \n 700-709\n
DOI 10.6009/jjrt.2021_JSRT_77.7.700
Language English
Journal Nihon Hoshasen Gijutsu Gakkai zasshi

Full Text