Electrical Activity of the Diaphragm in a Small Cohort of Preterm Infants on Noninvasive Neurally Adjusted Ventilatory Assist and Continuous Positive Airway Pressure: A Prospective Comparative Pilot Study

 
 
 
 
 
 

Abstract


Objective: To compare the electrical activity of the diaphragm (Edi) of premature infants placed on continuous positive airway pressure (CPAP) with the Edi of premature infants placed on noninvasive neurally adjusted ventilatory assist (NIV NAVA). The secondary aim was to evaluate the feasibility of the use of NIV NAVA mode in the busy tertiary neonatal unit. Study Design: This was a prospective crossover pilot study of premature infants requiring noninvasive respiratory support. Infants were randomized to initially receive either CPAP/biphasic (group 1) or NIV NAVA (group 2) and were then crossed over to the alternate group. Continuous Edi signals were recorded for 24 h, with 12 h each on CPAP/biphasic, and NIV NAVA along with other clinical and respiratory parameters. Results: Ten infants with a mean gestation age of 29 weeks (range 25-34 weeks) were enrolled, with a total cumulative Edi monitoring of 240 h. The average Edi peak on the biphasic/CPAP group (15.6 ± 7 mcV) was significantly higher (P < 0.005), compared to the Edi Peak on the NIV NAVA group (10.8 ± 3.3 mcV). The Edi min values were 3.23 ± 1.1 mcV, and 3.07 ± 0.5 mcV on CPAP/biphasic and NIV NAVA (P = 0.69) respectively. There were no significant differences in other clinical parameters between the two groups. No major adverse events were recorded during Edi catheter monitoring. Conclusion: The Edi peak values were significantly lower in NIV NAVA mode compared to CPAP/biphasic mode. The Edi catheter and NIV NAVA may also be used safely in premature infants.

Volume 11
Pages None
DOI 10.7759/cureus.6291
Language English
Journal Cureus

Full Text