Alexey A. Sysoev
National Research Nuclear University MEPhI
Network
Latest external collaboration on country level. Dive into details by clicking on the dots.
Publication
Featured researches published by Alexey A. Sysoev.
European Journal of Mass Spectrometry | 2002
Alexander A. Sysoev; Alexey A. Sysoev
At the beginning of the age of laser-ionisation mass spectrometry (LIMS) increasing numbers of publications were observed. However, later the method began to run into obstacles associated with poor reproducibility of analysis and large variations in elemental sensitivities so that the wide interest of the scientific community in the method faded away. However, the results described here show that the current knowledge of laser plasma processes, together with modern technical solutions to ion separation and quantification with time-of-flight (ToF) mass spectrometry, allow one to overcome the above-mentioned obstacles in LIMS. Thus, the performance in direct-sampling solid analysis demonstrated by the LAMAS-10M instrument is similar to that typically obtained by laser ablation/inductively-coupled plasma mass spectrometry (LA-ICP-MS) and glow-discharge mass spectrometry (GD-MS) methods. At the same time, there are additional advantages, including compactness of the instrument, absence of the need for expensive consumables and freedom from mass line interferences. Direct-sampling elemental LIMS is discussed as a promising alternative to LA-ICP-MS and GD-MS. Existing and prospective approaches to designing direct-sampling laser-ionisation mass spectrometers are theoretically justified. Factors affecting the main performance criteria, such as reproducibility, correctness, variations of relative sensitivity factors, linear dynamic range and resolution are considered. The demonstrated reproducibility, resolution, low-ppb limit of detection and one order-of-magnitude variation in elemental sensitivity are not the limit for direct-sampling laser-ionisation mass spectrometry of solid samples. Ways of improving LI-ToF-MS instrumental performance are discussed and theoretically justified.
Talanta | 2008
A.-K. Viitanen; Timo Mauriala; T. Mattila; Alexey Adamov; Christian Schack Pedersen; Jyrki M. Mäkelä; M. Marjamäki; Alexey A. Sysoev; J. Keskinen; Tapio Kotiaho
Performance of several time-of-flight (TOF) type ion mobility spectrometers (IMS) was compared in a joint measurement campaign and their mobility scales were adjusted based on the measurements. A standard reference compound 2,6-di-tert butylpyridine (2,6-DtBP) was used to create a single peak ion mobility distribution with a known mobility value. The effective length of the drift tube of each device, considered here as an instrument constant, was determined based on the measurements. Sequentially, two multi-peaked test compounds, DMMP and DIMP, were used to verify the performance of the adjustment procedure in a wider mobility scale. By determining the effective drift tube lengths using 2,6-DtBP, agreement between the devices was achieved. The determination of effective drift tube lengths according to standard reference compound was found to be a good method for instrument inter-comparison. The comparison procedure, its benefits and shortcomings as well as dependency on accuracy of literature value are discussed along with the results.
Analytical Chemistry | 2013
Alexey A. Sysoev; D.M. Chernyshev; S.S. Poteshin; A.V. Karpov; Oleg I. Fomin; Alexander A. Sysoev
Recently developed ion mobility mass spectrometer is described. The instrument is based on a drift tube ion mobility spectrometer and an orthogonal acceleration electrostatic sector time-of-flight mass analyzer. Data collection is performed using a specially developed fast ADC-based recorder that allows real-time data integration in an interval between 3 and 100 s. Primary tests were done with positive ion electrospray. The tests have shown obtaining 100 ion mobility resolving power and 2000 mass resolving power. Obtained for 2,6-di-tert-butylpyridine in electrosprayed liquid samples during 100 s analysis and full IMS/MS data collection mode were 4 nM relative limits of detection and a 1 pg absolute limit of detection (S/N=3). Characteristic ion mobility/mass distributions were recorded for selected antibiotics, including amoxicillin, ampicillin, lomefloxacin, and ofloxacin. At studied conditions, lomefloxacin forms only a protonated molecule-producing reduced ion mobility peak at 1.082 cm(2)/(V s). Both amoxicillin and ampicillin produce [M + H](+), [M + CH3OH + H](+), and [M + CH3CN + H](+). Amoxicillin shows two peaks at 0.909 cm(2)/(V s) and 0.905 cm(2)/(V s). Ampicillin shows one peak at 0.945 cm(2)/(V s). Intensity of protonated methanol containing cluster for both ampicillin and amoxicillin has a clear tendency to rise with sample keeping time. Ofloxacin produces two peaks in the ion mobility distribution. A lower ion mobility peak at 1.051 cm(2)/(V s) is shown to be formed by [M + H](+) ions. A higher ion mobility peak appearing for samples kept more than 48 h is shown to be formed by both [M + H](+) ion and a component identified as the [M + 2H + M](+2) cluster. The cluster probably partly dissociates in the interface producing the [M + H](+) ion.
Journal of the American Society for Mass Spectrometry | 2010
Jaakko Laakia; Alexey Adamov; Matti Jussila; Christian Schack Pedersen; Alexey A. Sysoev; Tapio Kotiaho
This study demonstrates how positive ion atmospheric pressure photoionization-ion mobility spectrometry-mass spectrometry (APPI-IMS-MS) can be used to produce different ionic forms of an analy te and how these can be separated. When hexane:toluene (9:1) is used as a solvent, 2,6-di-tert-butylpyridine (2,6-DtBPyr) and 2,6-di-tert-4-methylpyridine (2,6-DtB-4-MPyr) efficiently produce radical cations [M]+ and protonated [M + H]+ molecules, whereas, when the sample solvent is hexane, protonated molecules are mainly formed. Interestingly, radical cations drift slower in the drift tube than the protonated molecules. It was observed that an oxygen adduct ion, [M + O2]+, which was clearly seen in the mass spectra for hexane:toluene (9:1) solutions, shares the same mobility with radical cations, [M]+. Therefore, the observed mobility order is most likely explained by oxygen adduct formation, i.e., the radical cation forrning a heavier adduct. For pyridine and 2-tert-butylpyridine, only protonated molecules could be efficiently formed in the conditions used. For 1- and 2-naphthol it was observed that in hexane the protonated molecule typically had a higher intensity than the radical cation, whereas in hexane:toluene (9:1) the radical cation [M]+ typically had a higher intensity than the protonated molecule [M + H]+. Interestingly, the latter drifts slower than the radical cation [M]+, which is the opposite of the drift pattern seen for 2,6-DtBPyr and 2,6-DtB-4-MPyr.
International Journal of Mass Spectrometry | 2001
Alexey A. Sysoev; Raimo A. Ketola; Ismo Mattila; Virpi Tarkiaine; Tapio Kotiaho
Abstract Permeation parameters of several organic compounds were determined with a numerical simulation model developed earlier. Diffusivities were determined by calculating permeation curves at various diffusivity values and searching for a value of diffusivity that gave the best correlation of the theoretical curve with the experimental permeation curves. The method also allows determination of error in diffusivity calculation. Error is mainly caused by scattering of experimental data and adding organic interaction with a polymer into the character of a permeation curve. Gas phase distribution ratios were determined from literature data, and permeation selectivities were derived from comparison of experimental data measured by Membrane Inlet Mass Spectrometry (MIMS) and direct inlet measurements. Finding the highest concentration in gas and liquid phase at which the signal still behaves linearly allows estimation of the highest point of linear partitioning. The highest concentration at which diffusivity still remains constant can be estimated by finding the point at which the errors caused by the scattering of the experimental data and the contribution of organic interaction with a polymer into character of a permeation curve are equal. The model was postulated to be applicable over a concentration range in which membrane transport obeys ideal diffusion law, and there is linear sample/membrane partitioning. Calculated membrane diffusivities, water diffusivities, and water/membrane distribution ratios from literature sources were used to simulate permeation fluxes of organic compounds from aqueous phase as a function of the sample flow rate. Comparison of the simulated results showed generally good agreement with the experimental ones, and the expected behavior of permeating flux as a function of sample flow rate was observed.
Rapid Communications in Mass Spectrometry | 2009
Jaakko Laakia; Christian Schack Pedersen; Alexey Adamov; Jyrki Viidanoja; Alexey A. Sysoev; Tapio Kotiaho
Negative corona discharge atmospheric pressure chemical ionization (APCI) was used to investigate phenols with varying numbers of tert-butyl groups using ion mobility spectrometry-mass spectrometry (IMS-MS). The main characteristic ion observed for all the phenolic compounds was the deprotonated molecule [M-H](-). 2-tert-Butylphenol showed one main mobility peak in the mass-selected mobility spectrum of the [M-H](-) ion measured under nitrogen atmosphere. When air was used as a nebulizer gas an oxygen addition ion was seen in the mass spectrum and, interestingly, this new species [M-H+O](-) had a shorter drift time than the lighter [M-H](-) ion. Other phenolic compounds primarily produced two IMS peaks in the mass-selected mobility spectra measured using the [M-H](-) ion. It was also observed that two isomeric compounds, 2,4-di-tert-butylphenol and 2,6-di-tert-butylphenol, could be separated with IMS. In addition, mobilities of various characteristic ions of 2,4,6-trinitrotoluene were measured, since this compound was previously used as a mobility standard. The possibility of using phenolic compounds as mobility standards is also discussed.
European Journal of Mass Spectrometry | 2014
Alexey A. Sysoev; S.S. Poteshin; D.M. Chernyshev; A.V. Karpov; Yuriy B. Tuzkov; Vyacheslav V. Kyzmin; Alexander A. Sysoev
Characteristic ion mobility mass spectrometry data, reduced mobility, and limits of detection (signal-to-noise ratio = 3) were determined for six synthetic drugs and cocaine by ion mobility time-of-flight mass spectrometry (IM-TOF-MS) with electrospray ionization (ESI) and atmospheric pressure chemical ionization (APCI). The studied synthetic illicit drugs recently appeared on the recreational drug market as designer drugs and were methylone, 4-MEC (4′-methylethcathinone), 3,4-MDPV (3,4-methylenedioxypyrovalerone), JWH-210 [4-ethylnaphthalen-1-yl-(1-pentylindol-3-yl)methanone], JWH-250 [2-(2-methoxyphenyl)-1-(1-pentyl-1H-indol-3-yl)ethanone], and JWH-203 [1-pentyl-3-(2′-chlorophenylacetyl) indole]. Absolute reduced mobilities in nitrogen were 1.35, 1.28, 1.41, 1.30, 1.18, 0.98, 1.09, and 1.07 cm2 V−1 s−1, for methylone [M–H]+, methylone [M+H]+, 4-MEC [M–H]+, 4-MEC [M+H]+, 3,4-MDPV [M+H]+, JWH-210 [M+H]+, JWH-250 [M+H]+, and JWH-203 [M+H]+, respectively. Selected illicit drugs are easily identified by IM-TOF-MS during a 100 s analysis. Relative limits of detection ranged from 4 to 400 nM are demonstrated for these compounds. Such relative limits of detection correspond to 14 pg to 2 ng absolute limits of detection. Better detection limits are obtained in APCI mode for all the illicit drugs except cocaine. ESI mode was found to be preferable for the IM-TOF-MS detection of cocaine at trace levels. A single sample analysis is completed in an order of magnitude less time than that for conventional liquid chromatography/mass spectrometry approach. The application allows one to consider IM-TOF-MS as a good candidate for a method to determine quickly the recently surfaced designer drugs marketed on the internet as “bath salts,” “spice,” and “herbal blends”.
Journal of the American Society for Mass Spectrometry | 2008
Christian Schack Pedersen; Frants R. Lauritsen; Alexey A. Sysoev; A.-K. Viitanen; Jyrki M. Mäkelä; Alexey Adamov; Jaakko Laakia; Timo Mauriala; Tapio Kotiaho
Ionized acetates were used as model compounds to describe gas-phase behavior of oxygen containing compounds with respect to their formation of dimers in ion mobility spectrometry (IMS). The ions were created using corona discharge at atmospheric pressure and separated in a drift tube before analysis of the ions by mass spectrometry. At the ambient operational temperature and pressure used in our instrument, all acetates studied formed dimers. Using a homolog series of n-alkyl-acetates, we found that the collision cross section of a dimer was smaller than that of a monomer with the same reduced mass. Our experiments also showed that the reduced mobility of acetate dimers with different functional groups increased in the order n-alkyl ≤ branched chain alkyl ≤ cyclo alkyl < aromat. For mixed n-alkyl dimers we found that the reduced mobility of acetate dimers having the same number of carbons, for example a dimer of acetyl acetate and hexyl acetate has the same reduced mobility as a dimer composed of two butyl acetates. The fundamental behavior of acetate monomers and dimers described in this paper will assist in a better understanding of the influence of dimer formation in ion mobility spectrometry.
Review of Scientific Instruments | 2007
Alexey Adamov; Jyrki Viidanoja; Esko Kärpänoja; Heikki Paakkanen; Raimo A. Ketola; Risto Kostiainen; Alexey A. Sysoev; Tapio Kotiaho
This article presents the combination of an aspiration-type ion mobility spectrometer with a mass spectrometer. The interface between the aspiration ion mobility spectrometer and the mass spectrometer was designed to allow for quick mounting of the aspiration ion mobility spectrometer onto a Sciex API-300 triple quadrupole mass spectrometer. The developed instrumentation is used for gathering fundamental information on aspiration ion mobility spectrometry. Performance of the instrument is demonstrated using 2,6-di-tert-butyl pyridine and dimethyl methylphosphonate.
European Journal of Mass Spectrometry | 2015
Alexey A. Sysoev; V. I. Troyan; Peter V. Borisyuk; Andrey V. Krasavin; Oleg S. Vasiliev; Vitaly G. Palchikov; Ivan A. Avdeev; D.M. Chernyshev; S.S. Poteshin
There is a growing need for the development of atomic and nuclear frequency standards because of the important contribution of methods for precision time and frequency measurements to the development of fundamental science, technology and the economy. It is also conditioned by their potential use in optical clocks and quantum logic applications. It is especially important to develop a universal method that could allow one to use ions of most elements effectively (including ones that are not easily evaporated) proposed for the above-mentioned applications. A linear quadrupole ion trap for the optical spectroscopy of electron and nuclear transitions has been developed and evaluated experimentally. An ion source construction is based on an ultra-high vacuum evaporator in which a metal sample is subjected to an electron beam of energy up to 1 keV, resulting in the appearance of gaseous atoms and ions of various charge state. The linear ion trap consists of five successive quadrupole sections including an entrance quadrupole section, quadrupole mass filter, quadrupole ion guide, ion-trap section and exit quadrupole section. The same radiofrequency but a different direct current voltage feeds the quadrupole sections. The instrument allows the mass- and energy-selected trapping of ions from ion beams of various intensities and their localization in the area of laser irradiation. The preliminary results presented show that the proposed instrument and methods allow one to produce effectively up to triply charged thorium ions as well as to trap ions for future spectroscopic study. The instrument is proposed for future use in optical clocks and quantum logic application development.