Anthony O. Okorodudu
University of Texas Medical Branch
Network
Latest external collaboration on country level. Dive into details by clicking on the dots.
Publication
Featured researches published by Anthony O. Okorodudu.
Clinical Chemistry and Laboratory Medicine | 2006
Paul D'Orazio; Robert W. Burnett; Niels Fogh-Andersen; Ellis Jacobs; Katsuhiko Kuwa; Wolf R. Külpmann; Lasse Larsson; Andrzej Lewenstam; Anton H. J. Maas; Gerhard Mager; Jerzy W. Naskalski; Anthony O. Okorodudu
Abstract In current clinical practice, plasma and blood glucose are used interchangeably with a consequent risk of clinical misinterpretation. In human blood, glucose is distributed, like water, between erythrocytes and plasma. The molality of glucose (amount of glucose per unit water mass) is the same throughout the sample, but the concentration is higher in plasma, because the concentration of water and therefore glucose is higher in plasma than in erythrocytes. Different devices for the measurement of glucose may detect and report fundamentally different quantities. Different water concentrations in the calibrator, plasma, and erythrocyte fluid can explain some of the differences. Results for glucose measurements depend on the sample type and on whether the method requires sample dilution or uses biosensors in undiluted samples. If the results are mixed up or used indiscriminately, the differences may exceed the maximum allowable error for glucose determinations for diagnosing and monitoring diabetes mellitus, thus complicating patient treatment. The goal of the International Federation of Clinical Chemistry and Laboratory Medicine, Scientific Division, Working Group on Selective Electrodes and Point of Care Testing (IFCC-SD-WG-SEPOCT) is to reach a global consensus on reporting results. The document recommends reporting the concentration of glucose in plasma (in the unit mmol/L), irrespective of sample type or measurement technique. A constant factor of 1.11 is used to convert concentration in whole blood to the equivalent concentration in plasma. The conversion will provide harmonized results, facilitating the classification and care of patients and leading to fewer therapeutic misjudgments. Clin Chem Lab Med 2006;44:1486–90.
Critical Care Medicine | 1998
Gerald J. Kost; Huynh Troung Vu; Judith H. Lee; Peggy Bourgeois; Frederick L. Kiechle; Carol Martin; Sam S. Miller; Anthony O. Okorodudu; John J. Podczasy; Robert Webster; Karen J. Whitlow
OBJECTIVES Existing handheld glucose meters are glucose oxidase (GO)-based. Oxygen side reactions can introduce oxygen dependency, increase potential error, and limit clinical use. Our primary objectives were to: a) introduce a new glucose dehydrogenase (GD)-based electrochemical biosensor for point-of-care testing; b) determine the oxygen-sensitivity of GO- and GD-based electrochemical biosensor test strips; and c) evaluate the clinical performance of the new GD-based glucose meter system in critical care/hospital/ambulatory patients. DESIGN Multicenter study sites compared glucose levels determined with GD-based biosensors to glucose levels determined in whole blood with a perchloric acid deproteinization hexokinase reference method. One site also studied GO-based biosensors and venous plasma glucose measured with a chemistry analyzer. Biosensor test strips were used with a handheld glucose monitoring system. Bench and clinical oxygen sensitivity, hematocrit effect, and precision were evaluated. SETTING The study was performed at eight U.S. medical centers and one Canadian medical center. PATIENTS There were 1,248 patients. RESULTS The GO-based biosensor was oxygen-sensitive. The new GD-based biosensor was oxygen-insensitive. GD-based biosensor performance was acceptable: 2,104 (96.1%) of 2,189 glucose meter measurements were within +/-15 mg/dL (+/-0.83 mmol/L) for glucose levels of < or = 100 mg/dL (< or = 5.55 mmol/L) or within +/-15% for glucose levels of > 100 mg/dL, compared with the whole-blood reference method results. With the GD-based biosensor, the percentages of glucose measurements that were not within the error tolerance were comparable for different specimen types and clinical groups. Bracket predictive values were acceptable for glucose levels used in therapeutic management. CONCLUSIONS The performance of GD-based, oxygen-insensitive, handheld glucose testing was technically suitable for arterial specimens in critical care patients, cord blood and heelstick specimens in neonates, and capillary and venous specimens in other patients. Multicenter findings benchmark the performance of bedside glucose testing devices. With the new +/-15 mg/dL --> 100 mg/dL --> +/-15% accuracy criterion, point-of-care systems for handheld glucose testing should score 95% (or better), as compared with the recommended reference method. Physiologic changes, preanalytical factors, confounding variables, and treatment goals must be taken into consideration when interpreting glucose results, especially in critically ill patients, for whom arterial blood glucose measurements will reflect systemic glucose levels.
Clinica Chimica Acta | 2003
John R. Petersen; Anthony O. Okorodudu; Amin A. Mohammad; Deborah A. Payne
Over the past 10 years, capillary electrophoresis (CE) is an analytical tool that has shown great promise in replacing many conventional clinical laboratory methods, especially electrophoresis and high performance liquid chromatography (HPLC). The main attraction of CE was that it was fast, used small amounts of sample and reagents, and was extremely versatile, being able to separate large and small analytes, both neutral and charged. Because of this versatility, numerous methods for clinically relevant analytes have been developed. However, with the exception of the molecular diagnostic and forensic laboratories CE has not had a major impact. A possible reason is that CE is still perceived as requiring above-average technical expertise, precluding its use in a laboratory workforce that is less technically adept. With the introduction of multicapillary instruments that are more automated, less technique-dependent, in addition to the availability of commercial and cost effective test kit methods, CE may yet be accepted as a instrument routinely used in the clinical laboratories. Thus, this review will focus on the areas where CE shows the most potential to have the greatest impact on the clinical laboratory. These include analysis of proteins found in serum, urine, CSF and body fluids, immunosubstraction electrophoresis, hemoglobin variants, lipoproteins, carbohydrate-deficient transferrin (CDT), forensic and therapeutic drug screening, and molecular diagnostics.
Clinica Chimica Acta | 2008
John R. Petersen; Donna F. Graves; Danyel H. Tacker; Anthony O. Okorodudu; Amin A. Mohammad; Victor J. Cardenas
BACKGROUND Point of care (POC) glucose meters are routinely used to monitor glucose levels for patients on tight glycemic control therapy. We determined if glucose values were different for a POC glucose meter as compared to the main clinical laboratory for medical intensive care unit patients on a tight glycemic protocol and whether the site of blood sampling had a significant impact on glucose values. METHODS Eighty-four patients (114 paired samples) who were on a tight glycemic protocol in the period November 2005 through August 2006 were enrolled. After simultaneous blood draws, we compared the glucose levels for the glucose meter (arterial/venous/capillary), blood gas (arterial/venous), and central clinical laboratory (serum/plasma from arterial/venous samples). RESULTS The mean glucose levels of all arterial/venous/fingerstick samples using the glucose meter demonstrated a positive bias of 0.7-0.9 mmol/l (12.6-16.2 mg/dl) (p<0.001) relative to central laboratory venous plasma. There was also a smaller positive (0.1-0.3 mmol/l or 1.8-5.4 mg/dl, p<0.05) bias for arterial/venous blood gas samples and laboratory arterial serum/plasma glucose samples. Using Parkes error grid analysis we were able to show that the bias for arterial or venous POC glucose results would have not impacted clinical care. This was not the case, however, for fingerstick sampling where a high bias could have significantly impacted clinical care. Additionally, in 3 fingerstick samples a severe underestimation (<46% of the central laboratory plasma result) was found. CONCLUSION Glucose meters using arterial/venous whole blood may be utilized in the MICU; however, due to the increased variability of results we do not recommend the routine use of capillary blood sampling for monitoring glucose levels in the MICU setting.
Clinical Chemistry and Laboratory Medicine | 2008
Mohammed C. Ben Rayana; Robert W. Burnett; Arthur K. Covington; Paul D'Orazio; Niels Fogh-Andersen; Ellis Jacobs; Wolf R. Külpmann; Katsuhiko Kuwa; Lasse Larsson; Andrzej Lewenstam; Anton H. J. Maas; Gerhard Mager; Jerzy W. Naskalski; Anthony O. Okorodudu; Christoph Ritter; Andrew St John
Abstract Analyzers with ion-selective electrodes (ISEs) for ionized magnesium (iMg) should yield comparable and unbiased results for iMg. This IFCC guideline on sampling, measuring and reporting iMg in plasma provides a prerequisite to achieve this goal [in this document, “plasma” refers to circulating plasma and the forms in which it is sampled, namely the plasma phase of anticoagulated whole blood (or “blood”), plasma separated from blood cells, or serum]. The guideline recommends measuring and reporting ionized magnesium as a substance concentration relative to the substance concentration of magnesium in primary aqueous calibrants with magnesium, sodium, and calcium chloride of physiological ionic strength. The recommended name is “the concentration of ionized magnesium in plasma”. Based on this guideline, results will be approximately 3% higher than the true substance concentration and 4% lower than the true molality in plasma. Calcium ions interfere with all current magnesium ion-selective electrodes (Mg-ISEs), and thus it is necessary to determine both ions simultaneously in each sample and correct the result for Ca2+ interference. Binding of Mg in plasma is pH-dependent. Therefore, pH should be measured simultaneously with iMg to allow adjustment of the result to pH 7.4. The concentration of iMg in plasma may be physiologically and clinically more relevant than the concentration of total magnesium. Furthermore, blood-gas analyzers or instruments for point-of-care testing are able to measure plasma iMg using whole blood (with intact blood cells) as the sample, minimizing turn-around time compared to serum and plasma, which require removal of blood cells. Clin Chem Lab Med 2008;46:21–6.
Clinica Chimica Acta | 2000
Lokinendi V. Rao; Anthony O. Okorodudu; John R. Petersen; M.T Elghetany
Prothrombin time (PT) and activated partial thromboplastin time (aPTT) are common laboratory tests that are useful in the diagnosis of coagulation disorders and monitoring anticoagulant therapy. Recent expansions in the outreach laboratory services at our institution prompted us to investigate the shipping limitations for some tests, including PT and aPTT. Although we followed NCCLS guidelines for the collection of blood specimens, we observed falsely elevated PT and aPTT values due to the different storage conditions. The objective of this study is to determine the effect of conditions and duration of storage on PT and aPTT tests using plasma and whole blood samples, respectively. For this study, 36 plasma samples with normal and prolonged PT and aPTT were exposed to different storage conditions. Blood was centrifuged immediately and plasma was stored at room temperature (RT), refrigerated at 4 degrees C, or frozen at -20 degrees C. The samples were analyzed at 0 h and repeated at 6, 12 and 24 h under various conditions. Although statistically significant differences were observed for plasma samples for normal PT tests after 12 h at refrigerated and frozen storage conditions, the differences would not change the clinical interpretation of the results. On the other hand, samples stored refrigerated or at RT showed significant differences for aPTT at 24 h. These differences would change clinical interpretation, especially for samples with normal or near normal aPTT times. Interestingly, aPTT was significantly higher for samples stored frozen when compared to refrigerated and RT conditions at 6 h. Similar patterns were also observed on ten whole blood samples with normal PT and aPTT values. In conclusion, either plasma or whole blood samples can be accepted for PT testing up to 24 h and for aPTT testing up to 12 h only, when transported either at RT or at 4 degrees C.
Pediatrics | 1998
Terho Heikkinen; Faryal Ghaffar; Anthony O. Okorodudu; Tasnee Chonmaitree
Background. Increasing prevalence of antibiotic-resistant bacteria is a serious clinical problem that calls for reduction of unnecessary use of antibiotics. Acute otitis media (AOM) is the most common reason for antibiotic therapy in the United States. Approximately 30% of AOM cases do not have a bacterial etiology. Rapid identification of these cases could help withhold unnecessary antibiotic treatment. Objective. To determine the usefulness of serum levels of interleukin-6 (IL-6), an acute phase cytokine shown to be a reliable marker of neonatal bacterial infection, in differentiation between bacterial and nonbacterial AOM in children. Study Design. IL-6 was measured in stored serum samples from 184 children (mean age, 22 months) with AOM who were enrolled in antibiotic efficacy trials at our department. The samples were obtained at enrollment and at 9 to 12 days after initiation of antibiotic therapy. Sera from 21 uninfected children (mean age, 23 months) were used as controls. The etiology of AOM was determined by bacterial and viral cultures as well as respiratory syncytial virus antigen detection in the middle ear fluids obtained by tympanocentesis. Results. Bacterial etiology of AOM was confirmed in 125 children (68%), whereas in 59 children (32%) no bacterial pathogen could be detected in the middle ear fluid. Children with bacterial AOM had significantly higher IL-6 levels than those with nonbacterial AOM (median, 11.5 vs 3.7 pg/mL). However, this difference was almost entirely attributable to pneumococcal AOM specifically. IL-6 levels in children with AOM caused by Streptococcus pneumoniae were significantly higher (median, 40.1 pg/mL) than in AOM caused byHaemophilus influenzae (7.3 pg/mL) or Moraxella catarrhalis (6.8 pg/mL). At the cutoff value of 30 pg/mL, the specificity of IL-6 for detection of pneumococcal AOM was 91% with a sensitivity of 61%, but its sensitivity for detection of bacterial AOM in general was only 27%. Conclusions. Low levels of IL-6 do not rule out bacterial etiology of AOM in general; therefore, IL-6 is not sensitive enough as a marker of bacterial AOM. Surprisingly, serum IL-6 levels in pneumococcal AOM were significantly higher than the levels associated with other bacterial AOM, and serum IL-6 levels of >30 pg/mL were highly specific for pneumococcal AOM. These findings suggest a distinctive role for S pneumoniae in the pathogenesis of AOM.
Journal of Chromatography B: Biomedical Sciences and Applications | 1999
Lokinendi V. Rao; John R. Petersen; Michael G. Bissell; Anthony O. Okorodudu; Amin A. Mohammad
In clinical practice, the measurement of urinary free cortisol (UFC) provides the most sensitive and specific diagnostic information for excess adrenal production of cortisol. The existing methodologies (RIA and HPLC) are time consuming, costly, involve tedious extractions, derivatizations and problems with non-specific interactions with cortisol metabolites in urine. In the present study, we describe the development of an SPE-CE method for the rapid analysis of UFC. UFC was concentrated using SPE C18 cartridges (3M Empore) under a vacuum and eluted with acetonitrile-SDS. The use of 10% acetone to wash cartridges before final elution with acetonitrile-SDS showed significant improvements in the free cortisol recovery. The complete extraction was accomplished in 10-15 min with a recovery of 89-94%. CE analysis was done on a Beckman P/ACE 5010 with detection at 254 nm using a neutral capillary. Detection limits of free cortisol in urine was improved to 10 microg/l with SPE compared to 500 microg/l without SPE. No interferences either from BSA or other urinary cortisol metabolites affected the free cortisol determinations. The results showed the feasibility of a rapid UFC detection with improved sample handling capacity.
Journal of Tropical Pediatrics | 2002
Dorothy J. VanderJagt; Jaimie Shores; Anthony O. Okorodudu; Seline N. Okolo; Robert H. Glew
Reports of circulating lipids of children with sickle cell disease (SCD) in Nigeria disagree on the question of whether these children are at increased risk of cardiovascular disease (CVD). We therefore analyzed the serum of 40 females and 37 males with SCD, age 5-19 years, and equal numbers of age-matched controls for total cholesterol, HDL-cholesterol, LDL-cholesterol, triglycerides, and homocysteine. Using bioelectrical impedance analysis, we documented a significant reduction in the per cent fat-free mass in the SCD males and increases in the per cent body fat in both the male and female children with SCD. Marked hypocholesterolemia was present in both genders (means, 100-102 mg/dl) and the LDL-cholesterol levels of the male and female subjects with SCD (54 mg/dl) were below the lower limit of the reference range (59-137 mg/dl). Serum triglycerides in the SCD subjects were in the middle of the reference range for children. Although the mean HDL-cholesterol levels of the SCD males (23.1 mg/dl) and females (24.5 mg/dl) were well below the lower limit of the reference range (35-84 mg/dl), respectively, the LDL-cholesterol/HDL-cholesterol ratios of the SCD subjects were not abnormal. The mean serum homocysteine concentrations of the male and female SCD subjects (9.4-9.6 micromol/l) were at the high end of the normal range. Collectively, these results indicate that children with SCD in northern Nigeria are not at increased risk of CVD. However, their marked hypocholesterolemia should be a cause of concern about the overall mortality and general well-being.
Clinica Chimica Acta | 2010
John R. Petersen; Amin A. Mohammad; Thomas J. Shea; Anthony O. Okorodudu; Hyunsu Ju
BACKGROUND Glycosylated hemoglobin evaluation is very important for assessing the control of diabetes. Since the use of point-of-care (POC) devices for monitoring HbA1c is increasing, it is important to determine how these devices compare in relation to instrumentation used in the central laboratory (CL). METHODS Eighty-eight randomly selected samples previously analyzed using the Bio-Rad Variant™ II Hemoglobin Testing System were run on three POC Analyzers (Siemens DCA Vantage™ Analyzer, Axis-Shield Afinion™ AS100 Analyzer, and Bio-Rad In2it™ Analyzer). RESULTS All POC instruments showed good correlation to the CL method (R(2)>0.95 for all methods). HbA1c levels obtained using Variant II (mean=7.9; 95% CI=7.5-8.3%) and In2it (mean=7.9; 95% C.I.=7.5-8.2%) instruments were found to have no statistical mean difference (p=0.21), while the values obtained using DCA Vantage (mean=7.2% C.I.=6.9-7.5%) and Afinion (mean=7.3% C.I.=7.0-7.6%) instruments were different (p<0.001) from those of the CL method. The Afinion and DCA Vantage instruments increasingly underestimated the HbA1c compared to the CL as the HbA1c values increased. These differences were even more striking when the estimated average glucose is calculated. CONCLUSIONS Despite significant variation of results among the POC instruments evaluated relative to the CL method and pending resolution of HbA1c standardization issues, we conclude that all of the POC instruments can be used for HbA1c determination if clinicians are given instrument specific reference ranges.