Network


Latest external collaboration on country level. Dive into details by clicking on the dots.

Hotspot


Dive into the research topics where C. Huet is active.

Publication


Featured researches published by C. Huet.


Medical Physics | 2013

Small fields output factors measurements and correction factors determination for several detectors for a CyberKnife® and linear accelerators equipped with microMLC and circular cones.

C. Bassinet; C. Huet; S. Derreumaux; G. Brunet; M. Chéa; M. Baumann; T. Lacornerie; S. Gaudaire-Josset; F. Trompier; P. Roch; G. Boisserie; I. Clairand

PURPOSE The use of small photon fields is now an established practice in stereotactic radiosurgery and radiotherapy. However, due to a lack of lateral electron equilibrium and high dose gradients, it is difficult to accurately measure the dosimetric quantities required for the commissioning of such systems. Moreover, there is still no metrological dosimetric reference for this kind of beam today. In this context, the first objective of this work was to determine and to compare small fields output factors (OF) measured with different types of active detectors and passive dosimeters for three types of facilities: a CyberKnife(®) system, a dedicated medical linear accelerator (Novalis) equipped with m3 microMLC and circular cones, and an adaptive medical linear accelerator (Clinac 2100) equipped with an additional m3 microMLC. The second one was to determine the kQclin,Qmsr (fclin,fmsr) correction factors introduced in a recently proposed small field dosimetry formalism for different active detectors. METHODS Small field sizes were defined either by microMLC down to 6 × 6 mm(2) or by circular cones down to 4 mm in diameter. OF measurements were performed with several commercially available active detectors dedicated to measurements in small fields (high resolution diodes: IBA SFD, Sun Nuclear EDGE, PTW 60016, PTW 60017; ionizing chambers: PTW 31014 PinPoint chamber, PTW 31018 microLion liquid chamber, and PTW 60003 natural diamond). Two types of passive dosimeters were used: LiF microcubes and EBT2 radiochromic films. RESULTS Significant differences between the results obtained by several dosimetric systems were observed, particularly for the smallest field size for which the difference in the measured OF reaches more than 20%. For passive dosimeters, an excellent agreement was observed (better than 2%) between EBT2 and LiF microcubes for all OF measurements. Moreover, it has been shown that these passive dosimeters do not require correction factors and can then be used as reference dosimeters. Correction factors for the active detectors have then been determined from the mean experimental OF measured by the passive dosimeters. CONCLUSIONS Four sets of correction factors needed to apply the new small field dosimetry formalism are provided for several active detectors. A protocol for small photon beams OF determination based on passive dosimeters measurements has been recently proposed to French radiotherapy treatment centers.


Radiation Protection Dosimetry | 2008

Lessons from recent accidents in radiation therapy in France

S. Derreumaux; C. Etard; C. Huet; F. Trompier; I. Clairand; J.-F. Bottollier-Depois; B. Aubert; Patrick Gourmelon

Many accidents in radiotherapy have been reported in France over the last years. This is due to the recent legal obligation to declare to the national safety authorities any significant incident relative to the use of ionising radiation including medical applications. The causes and consequences of the most serious events in radiotherapy are presented in this paper. Lessons can be learned from possible technical dysfunctions, from human errors or organisational weaknesses as to how such events can be prevented. The technical aspects are addressed here: in particular, dosimetric issues.


Physics in Medicine and Biology | 2011

Construction of an extended library of adult male 3D models: rationale and results.

David Broggio; J Beurrier; M Bremaud; Aurélie Desbrée; J. Farah; C. Huet; D. Franck

In order to best cover the possible extent of heights and weights of male adults the construction of 25 whole body 3D models has been undertaken. Such a library is thought to be useful to specify the uncertainties and relevance of dosimetry calculations carried out with models representing individuals of average body heights and weights. Representative 3D models of Caucasian body types are selected in a commercial database according to their height and weight, and 3D models of the skeleton and internal organs are designed using another commercial dataset. A review of the literature enabled one to fix volume or mass target values for the skeleton, soft organs, skin and fat content of the selected individuals. The composition of the remainder tissue is fixed so that the weight of the voxel models equals the weight of the selected individuals. After mesh and NURBS modelling, volume adjustment of the selected body shapes and additional voxel-based work, 25 voxel models with 109 identified organs or tissue are obtained. Radiation transport calculations are carried out with some of the developed models to illustrate potential uses. The following points are discussed throughout this paper: justification of the fixed or obtained models’ features regarding available and relevant literature data; workflow and strategy for major modelling steps; advantages and drawbacks of the obtained library as compared with other works. The construction hypotheses are explained and justified in detail since future calculation results obtained with this library will depend on them.


Medical Physics | 2014

Determination of the correction factors for detectors used with an 800 MU/min CyberKnife® system equipped with fixed collimators and a study of detector response to small photon beams using a Monte Carlo method

C. Moignier; C. Huet; L. Makovicka

PURPOSE In a previous work, output ratio (ORdet) measurements were performed for the 800 MU/min CyberKnife(®) at the Oscar Lambret Center (COL, France) using several commercially available detectors as well as using two passive dosimeters (EBT2 radiochromic film and micro-LiF TLD-700). The primary aim of the present work was to determine by Monte Carlo calculations the output factor in water (OFMC,w) and the [Formula: see text] correction factors. The secondary aim was to study the detector response in small beams using Monte Carlo simulation. METHODS The LINAC head of the CyberKnife(®) was modeled using the PENELOPE Monte Carlo code system. The primary electron beam was modeled using a monoenergetic source with a radial gaussian distribution. The model was adjusted by comparisons between calculated and measured lateral profiles and tissue-phantom ratios obtained with the largest field. In addition, the PTW 60016 and 60017 diodes, PTW 60003 diamond, and micro-LiF were modeled. Output ratios with modeled detectors (ORMC,det) and OFMC,w were calculated and compared to measurements, in order to validate the model for smallest fields and to calculate [Formula: see text] correction factors, respectively. For the study of the influence of detector characteristics on their response in small beams; first, the impact of the atomic composition and the mass density of silicon, LiF, and diamond materials were investigated; second, the material, the volume averaging, and the coating effects of detecting material on the detector responses were estimated. Finally, the influence of the size of silicon chip on diode response was investigated. RESULTS Looking at measurement ratios (uncorrected output factors) compared to the OFMC,w, the PTW 60016, 60017 and Sun Nuclear EDGE diodes systematically over-responded (about +6% for the 5 mm field), whereas the PTW 31014 Pinpoint chamber systematically under-responded (about -12% for the 5 mm field). ORdet measured with the SFD diode and PTW 60003 diamond detectors were in good agreement with OFMC,w except for the 5 mm field size (about -7.5% for the diamond and +3% for the SFD). A good agreement with OFMC,w was obtained with the EBT2 film and micro-LiF dosimeters (deviation less than 1.4% for all fields investigated). [Formula: see text] correction factors for several detectors used in this work have been calculated. The impact of atomic composition on the dosimetric response of detectors was found to be insignificant, unlike the mass density and size of the detecting material. CONCLUSIONS The results obtained with the passive dosimeters showed that they can be used for small beam OF measurements without correction factors. The study of detector response showed that ORdet is depending on the mass density, the volume averaging, and the coating effects of the detecting material. Each effect was quantified for the PTW 60016 and 60017 diodes, the micro-LiF, and the PTW 60003 diamond detectors. None of the active detectors used in this work can be recommended as a reference for small field dosimetry, but an improved diode detector with a smaller silicon chip coated with tissue-equivalent material is anticipated (by simulation) to be a reliable small field dosimetric detector in a nonequilibrium field.


Health Physics | 2009

SESAME: a software tool for the numerical dosimetric reconstruction of radiological accidents involving external sources and its application to the accident in Chile in December 2005.

C. Huet; A. Lemosquet; I. Clairand; J B. Rioual; D. Franck; L. de Carlan; I. Aubineau-Lanièce; J.-F. Bottollier-Depois

Abstract—Estimating the dose distribution in a victim’s body is a relevant indicator in assessing biological damage from exposure in the event of a radiological accident caused by an external source. This dose distribution can be assessed by physical dosimetric reconstruction methods. Physical dosimetric reconstruction can be achieved using experimental or numerical techniques. This article presents the laboratory-developed SESAME—Simulation of External Source Accident with MEdical images—tool specific to dosimetric reconstruction of radiological accidents through numerical simulations which combine voxel geometry and the radiation-material interaction MCNP(X) Monte Carlo computer code. The experimental validation of the tool using a photon field and its application to a radiological accident in Chile in December 2005 are also described.


Medical Physics | 2015

Characterization of XR-RV3 GafChromic® films in standard laboratory and in clinical conditions and means to evaluate uncertainties and reduce errors

J. Farah; Annalisa Trianni; Olivera Ciraj-Bjelac; I. Clairand; C. De Angelis; S. delle Canne; L. Hadid; C. Huet; H. Järvinen; A. Negri; L. Novák; M. Pinto; T. Siiskonen; M. J. Waryn; Željka Knežević

PURPOSE To investigate the optimal use of XR-RV3 GafChromic(®) films to assess patient skin dose in interventional radiology while addressing the means to reduce uncertainties in dose assessment. METHODS XR-Type R GafChromic films have been shown to represent the most efficient and suitable solution to determine patient skin dose in interventional procedures. As film dosimetry can be associated with high uncertainty, this paper presents the EURADOS WG 12 initiative to carry out a comprehensive study of film characteristics with a multisite approach. The considered sources of uncertainties include scanner, film, and fitting-related errors. The work focused on studying film behavior with clinical high-dose-rate pulsed beams (previously unavailable in the literature) together with reference standard laboratory beams. RESULTS First, the performance analysis of six different scanner models has shown that scan uniformity perpendicular to the lamp motion axis and that long term stability are the main sources of scanner-related uncertainties. These could induce errors of up to 7% on the film readings unless regularly checked and corrected. Typically, scan uniformity correction matrices and reading normalization to the scanner-specific and daily background reading should be done. In addition, the analysis on multiple film batches has shown that XR-RV3 films have generally good uniformity within one batch (<1.5%), require 24 h to stabilize after the irradiation and their response is roughly independent of dose rate (<5%). However, XR-RV3 films showed large variations (up to 15%) with radiation quality both in standard laboratory and in clinical conditions. As such, and prior to conducting patient skin dose measurements, it is mandatory to choose the appropriate calibration beam quality depending on the characteristics of the x-ray systems that will be used clinically. In addition, yellow side film irradiations should be preferentially used since they showed a lower dependence on beam parameters compared to white side film irradiations. Finally, among the six different fit equations tested in this work, typically used third order polynomials and more rational and simplistic equations, of the form dose inversely proportional to pixel value, were both found to provide satisfactory results. Fitting-related uncertainty was clearly identified as a major contributor to the overall film dosimetry uncertainty with up to 40% error on the dose estimate. CONCLUSIONS The overall uncertainty associated with the use of XR-RV3 films to determine skin dose in the interventional environment can realistically be estimated to be around 20% (k = 1). This uncertainty can be reduced to within 5% if carefully monitoring scanner, film, and fitting-related errors or it can easily increase to over 40% if minimal care is not taken. This work demonstrates the importance of appropriate calibration, reading, fitting, and other film-related and scan-related processes, which will help improve the accuracy of skin dose measurements in interventional procedures.


Physica Medica | 2016

Study of commercial detector responses in non-equilibrium small photon fields of a 1000 MU/min CyberKnife system

C. Huet; Cyril Moignier; Victor Barraux; Cédric Loiseau; Karine Sebe-Mercier; Alain Batalla; I. Clairand

PURPOSE The purpose of this study was to analyze the detector responses in non-equilibrium small photon fields. METHODS Five detectors (PTW 31014 ionization chamber, PTW 60016, PTW 60017 and Sun Nuclear EDGE diodes and PTW 60003 diamond detector) and one passive dosimeter (Harshaw micro-LiF) as well as a 1000MU/min CyberKnife were modeled with the PENELOPE Monte Carlo code. Field factors, [Formula: see text] were calculated and perturbations due to volume averaging effect, active material effect and coating effect were quantified for the five detectors and passive dosimeter. RESULTS The PTW 31014 ionization chamber under-response is mainly due to the fluence perturbation caused by the presence of air as detecting material. Regarding diodes, the high density materials used in their active volume and in their coating is responsible for their over-response. Regarding the PTW 60003 diamond, its under-response for the 5mm field size is due to a large volume averaging effect whereas for largest field sizes a nearly perfect compensation between the volume averaging effect and the material effect due to the diamond density occurs. Despite its small size, a volume averaging effect was observed for the micro-LiF for the 5mm field size. CONCLUSION Perturbations due to volume averaging effect, active material effect and coating effect were investigated and quantified for five active detectors. Since these perturbations can cause opposite effects, wrong conclusions may be drawn regarding the radiological water-equivalence of detectors. Thus, we recommend performing such a study for each novel detector available on the market.


Journal of Applied Physics | 2015

Why diamond dimensions and electrode geometry are crucial for small photon beam dosimetry

F. Marsolat; D. Tromson; N. Tranchant; M. Pomorski; C. Bassinet; C. Huet; S. Derreumaux; M. Chea; K. Cristina; G. Boisserie; I. Buchheit; V. Marchesi; S. Gaudaire-Josset; A. Lisbona; D. Lazaro; R. Hugon; P. Bergonzo

Recent use of very small photon beams (down to 4 mm) in stereotactic radiotherapy requires new detectors to accurately determine the delivered dose. Diamond detectors have been presented in the literature as an attractive candidate for this application, due to their small detection volume and the diamond atomic number (Z = 6) which is close to water effective atomic number (Zeff ∼ 7.42). However, diamond exhibits a density 3.51 times greater than that of water and recent studies using Monte Carlo simulations have demonstrated the drawback of a high-density detector on small beam output factors. The current study focuses on geometrical parameters of diamond detector, namely, the diamond dimensions and the electrode geometry, in order to solve the dosimetric issues still observed in small photon beams with diamond detectors. To give better insights to these open questions, we have used both computational method and experimental analysis. This study highlighted that reducing diamond dimensions is crucial for...


Medical Physics | 2011

SU‐E‐T‐163: Characterization of the Response of Active Detectors and Passive Dosemeters Used for Dose Measurement in Small Photon Beams

S. Derreumaux; C. Bassinet; C. Huet; M. Chéa; G. Boisserie; G. Brunet; M. Baumann; F. Trompier; P. Roch; I. Clairand

Purpose: To characterize the response of detectors used for the dosimetry in small photon beams, especially for stereotactic radiation therapy. Methods: Measurements were performed with active detectors dedicated to the dosimetry in small fields (high resolution diodes IBA SFD, Sun Nuclear EDGE, PTW 60016 and PTW 60017; PTW 31014 0.015 cm3 PinPoint chamber, PTW 31018 microLion liquid chamber and PTW 60003 natural diamond) and with passive dosemeters suitable for small fields (TLD micro‐cubes, EBT2 radiochromic films). They were irradiated in a Cobalt 60 beam for investigation of reproducibility and drift and for study of the response as a function of the dose, in X‐ray linac beams ranging from 4 to 18 MV for study of the response as a function of the beam energy and as a function of the dose rate, and in small and large 6 MV X‐ray beams of different irradiating systems for study of the influence of beam size. Results: Regarding the reproducibility of the detector response, diodes give the best results, especially the EDGE and the PTW 60016 diodes with reproducibility better than 0.3%. The SFD diode reveals problem of stability, with sudden unexplainable large drift of the response. Both the PTW diodes undergo a slight variation of the response as a function of the dose rate. The detectors showing the highest variation of the response as a function of beam energy are the unshielded diodes. On the contrary, shielded diodes reveal more important impact of their non tissue equivalent composition on the measurement of output factors in very small fields (Huet et al., 2011; Bassinet et al., 2011). Correction factors depend on the measurement depth. Conclusions: This study will help to determine an optimized methodology for the measurement of the data characterizing the small photon beams used in radiotherapy.


Medical Physics | 2014

Determination of the kQclin,Qmsrfclin,fmsr correction factors for detectors used with an 800 MU/min CyberKnife® system equipped with fixed collimators and a study of detector response to small photon beams using a Monte Carlo method

C. Moignier; C. Huet; L. Makovicka

PURPOSE In a previous work, output ratio (ORdet) measurements were performed for the 800 MU/min CyberKnife(®) at the Oscar Lambret Center (COL, France) using several commercially available detectors as well as using two passive dosimeters (EBT2 radiochromic film and micro-LiF TLD-700). The primary aim of the present work was to determine by Monte Carlo calculations the output factor in water (OFMC,w) and the [Formula: see text] correction factors. The secondary aim was to study the detector response in small beams using Monte Carlo simulation. METHODS The LINAC head of the CyberKnife(®) was modeled using the PENELOPE Monte Carlo code system. The primary electron beam was modeled using a monoenergetic source with a radial gaussian distribution. The model was adjusted by comparisons between calculated and measured lateral profiles and tissue-phantom ratios obtained with the largest field. In addition, the PTW 60016 and 60017 diodes, PTW 60003 diamond, and micro-LiF were modeled. Output ratios with modeled detectors (ORMC,det) and OFMC,w were calculated and compared to measurements, in order to validate the model for smallest fields and to calculate [Formula: see text] correction factors, respectively. For the study of the influence of detector characteristics on their response in small beams; first, the impact of the atomic composition and the mass density of silicon, LiF, and diamond materials were investigated; second, the material, the volume averaging, and the coating effects of detecting material on the detector responses were estimated. Finally, the influence of the size of silicon chip on diode response was investigated. RESULTS Looking at measurement ratios (uncorrected output factors) compared to the OFMC,w, the PTW 60016, 60017 and Sun Nuclear EDGE diodes systematically over-responded (about +6% for the 5 mm field), whereas the PTW 31014 Pinpoint chamber systematically under-responded (about -12% for the 5 mm field). ORdet measured with the SFD diode and PTW 60003 diamond detectors were in good agreement with OFMC,w except for the 5 mm field size (about -7.5% for the diamond and +3% for the SFD). A good agreement with OFMC,w was obtained with the EBT2 film and micro-LiF dosimeters (deviation less than 1.4% for all fields investigated). [Formula: see text] correction factors for several detectors used in this work have been calculated. The impact of atomic composition on the dosimetric response of detectors was found to be insignificant, unlike the mass density and size of the detecting material. CONCLUSIONS The results obtained with the passive dosimeters showed that they can be used for small beam OF measurements without correction factors. The study of detector response showed that ORdet is depending on the mass density, the volume averaging, and the coating effects of the detecting material. Each effect was quantified for the PTW 60016 and 60017 diodes, the micro-LiF, and the PTW 60003 diamond detectors. None of the active detectors used in this work can be recommended as a reference for small field dosimetry, but an improved diode detector with a smaller silicon chip coated with tissue-equivalent material is anticipated (by simulation) to be a reliable small field dosimetric detector in a nonequilibrium field.

Collaboration


Dive into the C. Huet's collaboration.

Top Co-Authors

Avatar

I. Clairand

Institut de radioprotection et de sûreté nucléaire

View shared research outputs
Top Co-Authors

Avatar

C. Bassinet

Institut de radioprotection et de sûreté nucléaire

View shared research outputs
Top Co-Authors

Avatar

F. Trompier

Institut de radioprotection et de sûreté nucléaire

View shared research outputs
Top Co-Authors

Avatar

S. Derreumaux

Institut de radioprotection et de sûreté nucléaire

View shared research outputs
Top Co-Authors

Avatar

C. Moignier

Institut de radioprotection et de sûreté nucléaire

View shared research outputs
Top Co-Authors

Avatar

M. Baumann

Institut de radioprotection et de sûreté nucléaire

View shared research outputs
Top Co-Authors

Avatar

L. Makovicka

University of Franche-Comté

View shared research outputs
Top Co-Authors

Avatar

P. Roch

Institut de radioprotection et de sûreté nucléaire

View shared research outputs
Top Co-Authors

Avatar

D. Franck

Institut de radioprotection et de sûreté nucléaire

View shared research outputs
Top Co-Authors

Avatar

J.-F. Bottollier-Depois

Institut de radioprotection et de sûreté nucléaire

View shared research outputs
Researchain Logo
Decentralizing Knowledge