Network


Latest external collaboration on country level. Dive into details by clicking on the dots.

Hotspot


Dive into the research topics where Florian Cremers is active.

Publication


Featured researches published by Florian Cremers.


Physics in Medicine and Biology | 2007

Dosimetric evaluation of a 2D pixel ionization chamber for implementation in clinical routine

Julia Herzen; Manuel Todorovic; Florian Cremers; V Platz; Dirk Albers; A. Bartels; Rainer Schmidt

In this paper we present the results of a dosimetric evaluation of a 2D ionization chamber array with the objective of its implementation for quality assurance in clinical routine. The pixel ionization chamber MatriXX (Scanditronix Wellhofer, Germany) consists of 32x32 chambers with a distance of 7.6 mm between chamber centres. The effective depth of measurement under the surface of the detector was determined. The dose and energy dependence, the behaviour of the device during its initial phase and its time stability as well as the lateral response of a single chamber of the detector in cross-plane and diagonal directions were analysed. It could be shown, that the detectors response is linear with dose and energy independent. Taking the lateral response into account, two different dose profiles, for a pyramidal and an IMRT dose distribution, were applied to compare the data generated by a treatment planning system with measurements. From these investigations it can be concluded that the detector is a suitable device for quality assurance and 2D dose verifications.


Medical Physics | 2006

Evaluation of GafChromic EBT prototype B for external beam dose verification

Manuel Todorovic; M Fischer; Florian Cremers; E. Thom; Rainer Schmidt

The capability of the new GafChromic EBT prototype B for external beam dose verification is investigated in this paper. First the general characteristics of this film (dose response, postirradiation coloration, influence of calibration field size) were derived using a flat-bed scanner. In the dose range from 0.1 to 8 Gy, the sensitivity of the EBT prototype B film is ten times higher than the response of the GafChromic HS, which so far was the GafChromic film with the highest sensitivity. Compared with the Kodak EDR2 film, the response of the EBT is higher by a factor of 3 in the dose range from 0.1 to 8 Gy. The GafChromic EBT almost does not show a temporal growth of the optical density and there is no influence of the chosen calibration field size on the dose response curve obtained from this data. A MatLab program was written to evaluate the two-dimensional dose distributions from treatment planning systems and GafChromic EBT film measurements. Verification of external beam therapy (SRT, IMRT) using the above-mentioned approach resulted in very small differences between the planned and the applied dose. The GafChromic EBT prototype B together with the flat-bed scanner and MatLab is a successful approach for making the advantages of the GafChromic films applicable for verification of external beam therapy.


Physics in Medicine and Biology | 2006

Design of a computer-controlled multileaf collimator for advanced electron radiotherapy

T. Gauer; Dirk Albers; Florian Cremers; R Harmansa; R Pellegrini; Rainer Schmidt

A multileaf collimator for electrons (eMLC) has been designed that fulfils the technical requirements for providing advanced irradiation techniques with electrons. In the present work, the basic design parameters of leaf material, leaf height, leaf width and number of leaves as well as leaf overtravel and leaf shape were determined such that an eMLC with motorized leaves can be manufactured by a company specialized in MLC technology. For this purpose, a manually driven eMLC with variable source-to-collimator distance (SCD) was used to evaluate the chosen leaf specification and investigate the impact of the SCD on the off-axis dose distribution. In order to select the final SCD of the eMLC, a compromise had to be found between maximum field size, minimum beam penumbra and necessary distance between eMLC and isocentre to eliminate patient realignments during gantry rotation. As a result, the eMLC is placed according to the target position at 72 and 84 cm SCD, respectively. This feature will be achieved by interchangeable distance holders. At these SCDs, the corresponding maximum field sizes at 100 cm source-to-isocentre distance are 20 x 20 cm and 17 x 17 cm, respectively. Finally, the off-axis dose distribution at the maximum opening of the eMLC was improved by fine-tuning the settings of the accelerator jaws and introducing trimmer bars above the eMLC. Following this optimization, a prototype eMLC consisting of 2 x 24 computer-controlled brass leaves is manufactured by 3D Line Medical Systems.


Physics in Medicine and Biology | 2008

Characterization of an add-on multileaf collimator for electron beam therapy

T. Gauer; J. Sokoll; Florian Cremers; R Harmansa; M Luzzara; Rainer Schmidt

An add-on multileaf collimator for electrons (eMLC) has been developed that provides computer-controlled beam collimation and isocentric dose delivery. The design parameters result from the design study by Gauer et al (2006 Phys. Med. Biol. 51 5987-6003) and were configured such that a compact and light-weight eMLC with motorized leaves can be industrially manufactured and stably mounted on a conventional linear accelerator. In the present study, the efficiency of an initial computer-controlled prototype was examined according to the design goals and the performance of energy- and intensity-modulated treatment techniques. This study concentrates on the attachment and gantry stability as well as the dosimetric characteristics of central-axis and off-axis dose, field size dependence, collimator scatter, field abutment, radiation leakage and the setting of the accelerator jaws. To provide isocentric irradiation, the eMLC can be placed either 16 or 28 cm above the isocentre through interchangeable holders. The mechanical implementation of this feature results in a maximum field displacement of less than 0.6 mm at 90 degrees and 270 degrees gantry angles. Compared to a 10 x 10 cm applicator at 6-14 MeV, the beam penumbra of the eMLC at a 16 cm collimator-to-isocentre distance is 0.8-0.4 cm greater and the depth-dose curves show a larger build-up effect. Due to the loss in energy dependence of the therapeutic range and the much lower dose output at small beam sizes, a minimum beam size of 3 x 3 cm is necessary to avoid suboptimal dose delivery. Dose output and beam symmetry are not affected by collimator scatter when the central axis is blocked. As a consequence of the broader beam penumbra, uniform dose distributions were measured in the junction region of adjacent beams at perpendicular and oblique beam incidence. However, adjacent beams with a high difference in a beam energy of 6 to 14 MeV generate cold and hot spots of approximately 15% in the abutting region. In order to improve uniformity, the energy of adjacent beams must be limited to 6 to 10 MeV and 10 to 14 MeV respectively. At the maximum available beam energy of 14 MeV, radiation leakage results mainly from the intraleaf leakage of approximately 2.5% relative dose which could be effectively eliminated at off-axis distances remote from the field edge by adjusting the jaw field size to the respective opening of the eMLC. Additionally, the interleaf and leaf-end leakage could be reduced by using a tongue-and-groove leaf shape and adjoining the leaf-ends off-axis respectively.


Zeitschrift Fur Medizinische Physik | 2012

Towards accurate dose accumulation for Step-&-Shoot IMRT: Impact of weighting schemes and temporal image resolution on the estimation of dosimetric motion effects.

René Werner; Jan Ehrhardt; Alexander Schmidt-Richberg; Dirk Albers; Thorsten Frenzel; Cordula Petersen; Florian Cremers; Heinz Handels

PURPOSE Breathing-induced motion effects on dose distributions in radiotherapy can be analyzed using 4D CT image sequences and registration-based dose accumulation techniques. Often simplifying assumptions are made during accumulation. In this paper, we study the dosimetric impact of two aspects which may be especially critical for IMRT treatment: the weighting scheme for the dose contributions of IMRT segments at different breathing phases and the temporal resolution of 4D CT images applied for dose accumulation. METHODS Based on a continuous problem formulation a patient- and plan-specific scheme for weighting segment dose contributions at different breathing phases is derived for use in step-&-shoot IMRT dose accumulation. Using 4D CT data sets and treatment plans for 5 lung tumor patients, dosimetric motion effects as estimated by the derived scheme are compared to effects resulting from a common equal weighting approach. Effects of reducing the temporal image resolution are evaluated for the same patients and both weighting schemes. RESULTS The equal weighting approach underestimates dosimetric motion effects when considering single treatment fractions. Especially interplay effects (relative misplacement of segments due to respiratory tumor motion) for IMRT segments with only a few monitor units are insufficiently represented (local point differences >25% of the prescribed dose for larger tumor motion). The effects, however, tend to be averaged out over the entire treatment course. Regarding temporal image resolution, estimated motion effects in terms of measures of the CTV dose coverage are barely affected (in comparison to the full resolution) when using only half of the original resolution and equal weighting. In contrast, occurence and impact of interplay effects are poorly captured for some cases (large tumor motion, undersized PTV margin) for a resolution of 10/14 phases and the more accurate patient- and plan-specific dose accumulation scheme. CONCLUSIONS Radiobiological consequences of reported single fraction local point differences >25% of the prescribed dose are widely unclear and should be subject to future investigation. Meanwhile, if aiming at accurate and reliable estimation of dosimetric motion effects, precise weighting schemes such as the presented patient- and plan-specific scheme for step-&-shoot IMRT and full available temporal 4D CT image resolution should be applied for IMRT dose accumulation.


Journal of Applied Clinical Medical Physics | 2016

Assessment of image quality and dose calculation accuracy on kV CBCT, MV CBCT, and MV CT images for urgent palliative radiotherapy treatments.

M Held; Florian Cremers; Penny K. Sneed; Steve Braunstein; Shannon Fogh; Igor J. Barani; A Perez-Andujar; Jean Pouliot; Olivier Morin

A clinical workflow was developed for urgent palliative radiotherapy treatments that integrates patient simulation, planning, quality assurance, and treatment in one 30‐minute session. This has been successfully tested and implemented clinically on a linac with MV CBCT capabilities. To make this approach available to all clinics equipped with common imaging systems, dose calculation accuracy based on treatment sites was assessed for other imaging units. We evaluated the feasibility of palliative treatment planning using on‐board imaging with respect to image quality and technical challenges. The purpose was to test multiple systems using their commercial setup, disregarding any additional in‐house development. kV CT, kV CBCT, MV CBCT, and MV CT images of water and anthropomorphic phantoms were acquired on five different imaging units (Philips MX8000 CT Scanner, and Varian TrueBeam, Elekta VersaHD, Siemens Artiste, and Accuray Tomotherapy linacs). Image quality (noise, contrast, uniformity, spatial resolution) was evaluated and compared across all machines. Using individual image value to density calibrations, dose calculation accuracies for simple treatment plans were assessed for the same phantom images. Finally, image artifacts on clinical patient images were evaluated and compared among the machines. Image contrast to visualize bony anatomy was sufficient on all machines. Despite a high noise level and low contrast, MV CT images provided the most accurate treatment plans relative to kV CT‐based planning. Spatial resolution was poorest for MV CBCT, but did not limit the visualization of small anatomical structures. A comparison of treatment plans showed that monitor units calculated based on a prescription point were within 5% difference relative to kV CT‐based plans for all machines and all studied treatment sites (brain, neck, and pelvis). Local dose differences >5% were found near the phantom edges. The gamma index for 3%/3 mm criteria was ≥95% in most cases. Best dose calculation results were obtained when the treatment isocenter was near the image isocenter for all machines. A large field of view and immediate image export to the treatment planning system were essential for a smooth workflow and were not provided on all devices. Based on this phantom study, image quality of the studied kV CBCT, MV CBCT, and MV CT on‐board imaging devices was sufficient for treatment planning in all tested cases. Treatment plans provided dose calculation accuracies within an acceptable range for simple, urgently planned palliative treatments. However, dose calculation accuracy was compromised towards the edges of an image. Feasibility for clinical implementation should be assessed separately and may be complicated by machine specific features. Image artifacts in patient images and the effect on dose calculation accuracy should be assessed in a separate, machine‐specific study. PACS number(s): 87.55.D‐, 87.57.C‐, 87.57.Q‐A clinical workflow was developed for urgent palliative radiotherapy treatments that integrates patient simulation, planning, quality assurance, and treatment in one 30-minute session. This has been successfully tested and implemented clinically on a linac with MV CBCT capabilities. To make this approach available to all clinics equipped with common imaging systems, dose calculation accuracy based on treatment sites was assessed for other imaging units. We evaluated the feasibility of palliative treatment planning using on-board imaging with respect to image quality and technical challenges. The purpose was to test multiple systems using their commercial setup, disregarding any additional in-house development. kV CT, kV CBCT, MV CBCT, and MV CT images of water and anthropomorphic phantoms were acquired on five different imaging units (Philips MX8000 CT Scanner, and Varian TrueBeam, Elekta VersaHD, Siemens Artiste, and Accuray Tomotherapy linacs). Image quality (noise, contrast, uniformity, spatial resolution) was evaluated and compared across all machines. Using individual image value to density calibrations, dose calculation accuracies for simple treatment plans were assessed for the same phantom images. Finally, image artifacts on clinical patient images were evaluated and compared among the machines. Image contrast to visualize bony anatomy was sufficient on all machines. Despite a high noise level and low contrast, MV CT images provided the most accurate treatment plans relative to kV CT-based planning. Spatial resolution was poorest for MV CBCT, but did not limit the visualization of small anatomical structures. A comparison of treatment plans showed that monitor units calculated based on a prescription point were within 5% difference relative to kV CT-based plans for all machines and all studied treatment sites (brain, neck, and pelvis). Local dose differences >5% were found near the phantom edges. The gamma index for 3%/3 mm criteria was ≥95% in most cases. Best dose calculation results were obtained when the treatment isocenter was near the image isocenter for all machines. A large field of view and immediate image export to the treatment planning system were essential for a smooth workflow and were not provided on all devices. Based on this phantom study, image quality of the studied kV CBCT, MV CBCT, and MV CT on-board imaging devices was sufficient for treatment planning in all tested cases. Treatment plans provided dose calculation accuracies within an acceptable range for simple, urgently planned palliative treatments. However, dose calculation accuracy was compromised towards the edges of an image. Feasibility for clinical implementation should be assessed separately and may be complicated by machine specific features. Image artifacts in patient images and the effect on dose calculation accuracy should be assessed in a separate, machine-specific study. PACS number(s): 87.55.D-, 87.57.C-, 87.57.Q.


Medical Physics | 2014

Suitability of markerless EPID tracking for tumor position verification in gated radiotherapy

Marco Serpa; Kurt Baier; Florian Cremers; Matthias Guckenberger; Juergen Meyer

PURPOSE To maximize the benefits of respiratory gated radiotherapy (RGRT) of lung tumors real-time verification of the tumor position is required. This work investigates the feasibility of markerless tracking of lung tumors during beam-on time in electronic portal imaging device (EPID) images of the MV therapeutic beam. METHODS EPID movies were acquired at ∼2 fps for seven lung cancer patients with tumor peak-to-peak motion ranges between 7.8 and 17.9 mm (mean: 13.7 mm) undergoing stereotactic body radiotherapy. The external breathing motion of the abdomen was synchronously measured. Both datasets were retrospectively analyzed in PortalTrack, an in-house developed tracking software. The authors define a three-step procedure to run the simulations: (1) gating window definition, (2) gated-beam delivery simulation, and (3) tumor tracking. First, an amplitude threshold level was set on the external signal, defining the onset of beam-on/-off signals. This information was then mapped onto a sequence of EPID images to generate stamps of beam-on/-hold periods throughout the EPID movies in PortalTrack, by obscuring the frames corresponding to beam-off times. Last, tumor motion in the superior-inferior direction was determined on portal images by the tracking algorithm during beam-on time. The residual motion inside the gating window as well as target coverage (TC) and the marginal target displacement (MTD) were used as measures to quantify tumor position variability. RESULTS Tumor position monitoring and estimation from beams-eye-view images during RGRT was possible in 67% of the analyzed beams. For a reference gating window of 5 mm, deviations ranging from 2% to 86% (35% on average) were recorded between the reference and measured residual motion. TC (range: 62%-93%; mean: 77%) losses were correlated with false positives incidence rates resulting mostly from intra-/inter-beam baseline drifts, as well as sudden cycle-to-cycle fluctuations in exhale positions. Both phenomena can lead to considerable deviations (with MTD values up to a maximum of 7.8 mm) from the intended tumor position, and in turn may result in a marginal miss. The difference between tumor traces determined within the gating window against ground truth trajectory maps was 1.1 ± 0.7 mm on average (range: 0.4-2.3 mm). CONCLUSIONS In this retrospective analysis of motion data, it is demonstrated that the system is capable of determining tumor positions in the plane perpendicular to the beam direction without the aid of fiducial markers, and may hence be suitable as an online verification tool in RGRT. It may be possible to use the tracking information to enable on-the-fly corrections to intra-/inter-beam variations by adapting the gating window by means of a robotic couch.


Zeitschrift Fur Medizinische Physik | 2015

Design, performance characteristics and application examples of a new 4D motion platform.

Carsten Grohmann; Thorsten Frenzel; René Werner; Florian Cremers

In this publication, a three-dimensionally movable motion phantom is described and its performance characteristics are evaluated. The intended primary fields of application for the phantom are the quality assurance (QA) of respiratory motion management devices in radiation therapy (RT) like gating or tumour tracking systems, training for clinical use of these techniques, and related 4DRT research. Considering especially the QA aspect, the phantom was designed as a motion platform that can be equipped with an appropriate add-on like standard QA phantoms for dosimetric measurements. The platform is driven by three computer-controlled independent linear motors (motion range: 40 × 50 × 50 mm in anterior-posterior/superior-inferior/lateral direction; max. velocity: 3.9 m/s; max. acceleration: 10 m/s(2)), which allow the simulation of normal breathing patterns as well as arbitrary trajectories and anomalous events like coughing or baseline drift. For normal breathing patterns (here: sinusoidal curves with an amplitude of 20mm and a period of 3 s/6 s), the accuracy of the simulated motion paths was measured to be within 0,521 mm even for the ArcCHECK (weight: 20 kg) as a platform load - values that we consider to be sufficient for the intended fields of application. The respective use of the motion phantom is illustrated.


Journal of Applied Clinical Medical Physics | 2010

Commissioning of a double-focused micro multileaf collimator (μMLC)

Marcus Fischer; Manuel Todorovic; Eva Drud; Florian Cremers

Double‐focused μMLCs are able to create fields with steeper dose gradients at the field edges and are, therefore, an advancement in delivering stereotactic treatments. A double‐focused μMLC has been installed at a Siemens Primus linear accelerator (linac) as a first research installation in Europe. The basic dosimetric parameters, such as leakage, output factors, depth‐dose curves and penumbra, have been measured in 6 and 15 MV‐mode by use of radiochromic films (GafChromic EBT), ionization chambers and our solid water QA‐phantom (Easy Cube). The leakage between the leaves is minimal and lower than that of most commercially available MLCs. Therefore, the field size of the linac can be kept constant while the leaves of the μMLC are creating different aperture shapes. Percentage depth doses (PDDs) generated by the double‐focused μMLC are equal to depth‐dose curves of the original linac. That means the μMLC affects only the off‐axis ratio (OAR). Based on the fact that the μMLC is double‐focused and the source‐to‐collimator distance is larger, the penumbra is sharper than that for fields defined by the original linac MLC. The mechanical and dosimetric investigations show the benefit of the double‐focused μMLC attached to a Siemens Primus linear accelerator. PACS number: 87.10.+e


Journal of Applied Clinical Medical Physics | 2017

The effects of metal artifact reduction on the retrieval of attenuation values

Christian Ziemann; Maik Stille; Florian Cremers; Dirk Rades; Thorsten M. Buzug

Abstract Background The quality of CT slices can be drastically reduced in the presence of high‐density objects such as metal implants within the patients’ body due to the occurrence of streaking artifacts. Consequently, a delineation of anatomical structures might not be possible, which strongly influences clinical examination. Purpose The aim of the study is to clinically evaluate the retrieval of attenuation values and structures by the recently proposed Augmented Likelihood Image Reconstruction (ALIR) and linear interpolation in the presence of metal artifacts. Material and Methods A commercially available phantom was equipped with two steel inserts. At a position between the metal rods, which shows severe streaking artifacts, different human tissue‐equivalent inserts are alternately mounted. Using a single‐source computer tomograph, raw data with and without metal rods are acquired for each insert. Images are reconstructed using the ALIR algorithm and a filtered back projection with and without linear interpolation. Mean and standard deviation are compared for a region of interest in the ALIR reconstructions, linear interpolation results, uncorrected images with metal rods, and the images without metal rods, which are used as a reference. Furthermore, the reconstructed shape of the inserts is analyzed by comparing different profiles of the image. Results The measured mean and standard deviation values show that for all tissue classes, the metal artifacts could be reduced using the ALIR algorithm and the linear interpolation. Furthermore, the HU values for the different classes could be retrieved with errors below the standard deviation in the reference image. An evaluation of the shape of the inserts shows that the reconstructed object fits the shape of the insert accurately after metal artifact correction. Moreover, the evaluation shows a drop in the standard deviation for the ALIR reconstructed images compared to the reference images while reducing artifacts and keeping the shape of the inserts, which indicates a noise reduction ability of the ALIR algorithm. Conclusion HU values, which are distorted by metal artifacts, can be retrieved accurately with the ALIR algorithm and the linear interpolation approach. After metal artifact correction, structures, which are not perceptible in the original images due to streaking artifacts, are reconstructed correctly within the image using the ALIR algorithm. Furthermore, the ALIR produced images with a reduced noise level compared to reference images and artifact images. Linear interpolation results in a distortion of the investigated shapes and features remaining streaking artifacts.

Collaboration


Dive into the Florian Cremers's collaboration.

Top Co-Authors

Avatar
Top Co-Authors

Avatar
Top Co-Authors

Avatar
Top Co-Authors

Avatar
Top Co-Authors

Avatar
Top Co-Authors

Avatar

T. Gauer

University of Hamburg

View shared research outputs
Top Co-Authors

Avatar
Top Co-Authors

Avatar
Top Co-Authors

Avatar
Top Co-Authors

Avatar
Researchain Logo
Decentralizing Knowledge