Hali Morrison
University of Alberta
Network
Latest external collaboration on country level. Dive into details by clicking on the dots.
Publication
Featured researches published by Hali Morrison.
Medical Physics | 2014
Hali Morrison; Geetha Menon; Ron S. Sloboda
PURPOSE Radiochromic film dosimetry is typically performed for high energy photons and moderate doses characterizing external beam radiotherapy (XRT). The purpose of this study was to investigate the accuracy of previously established film calibration procedures used in XRT when applied to low-energy, seed-based brachytherapy at higher doses, and to determine necessary modifications to achieve similar accuracy in absolute dose measurements. METHODS Gafchromic EBT3 film was used to measure radiation doses upwards of 35 Gy from 75 kVp, 200 kVp, 6 MV, and (∼28 keV) I-125 photon sources. For the latter irradiations a custom phantom was built to hold a single I-125 seed. Film pieces were scanned with an Epson 10000XL flatbed scanner and the resulting 48-bit RGB TIFF images were analyzed using both FilmQA Pro software andMATLAB. Calibration curves relating dose and optical density via a rational functional form for all three color channels at each irradiation energy were determined with and without the inclusion of uncertainties in the measured optical densities and dose values. The accuracy of calibration curve variations obtained using piecewise fitting, a reduced film measurement area for I-125 irradiation, and a reduced number of dose levels was also investigated. The energy dependence of the film lot used was also analyzed by calculating normalized optical density values. RESULTS Slight differences were found in the resulting calibration curves for the various fitting methods used. The accuracy of the calibration curves was found to improve at low doses and worsen at high doses when including uncertainties in optical densities and doses, which may better represent the variability that could be seen in film optical density measurements. When exposing the films to doses > 8 Gy, two-segment piecewise fitting was found to be necessary to achieve similar accuracies in absolute dose measurements as when using smaller dose ranges. When reducing the film measurement area for the I-125 irradiations, the accuracy of the calibration curve was degraded due to the presence of localized film heterogeneities. No degradation in the calibration curves was found when reducing the number of calibration points down to only 4, but with piecewise fitting, 6 calibration points as well as a blank film are required. Variations due to photon energy in film optical density of up to 3% were found above doses of 2 Gy. CONCLUSIONS A modified procedure for performing EBT3 film calibration was established for use with low-energy brachytherapy seeds and high dose exposures. The energy dependence between 6 MV and I-125 photons is significant such that film calibrations should be done with an appropriately low-energy source when performing low-energy brachytherapy dose measurements. Two-segment piecewise fitting with the inclusion of errors in measured optical density as well as dose was found to result in the most accurate calibration curves. Above doses of 1 Gy, absolute dose measurements can be made with an accuracy of 1.6% for 6 MV beams and 5.7% for I-125 seed exposures if using the I-125 source for calibration, or 2.3% if using the 75 kVp photon beam for calibration.
Journal of Applied Clinical Medical Physics | 2017
Brie Cawston-Grant; Hali Morrison; Geetha Menon; Ron S. Sloboda
&NA; Model‐based dose calculation algorithms have recently been incorporated into brachytherapy treatment planning systems, and their introduction requires critical evaluation before clinical implementation. Here, we present an experimental evaluation of Oncentra® Brachy Advanced Collapsed‐cone Engine (ACE) for a multichannel vaginal cylinder (MCVC) applicator using radiochromic film. A uniform dose of 500 cGy was specified to the surface of the MCVC using the TG‐43 dose formalism under two conditions: (a) with only the central channel loaded or (b) only the peripheral channels loaded. Film measurements were made at the applicator surface and compared to the doses calculated using TG‐43, standard accuracy ACE (sACE), and high accuracy ACE (hACE). When the central channel of the applicator was used, the film measurements showed a dose increase of (11 ± 8)% (k = 2) above the two outer grooves on the applicator surface. This increase in dose was confirmed with the hACE calculations, but was not confirmed with the sACE calculations at the applicator surface. When the peripheral channels were used, a periodic azimuthal variation in measured dose was observed around the applicator. The sACE and hACE calculations confirmed this variation and agreed within 1% of each other at the applicator surface. Additionally for the film measurements with the central channel used, a baseline dose variation of (10 ± 4)% (k = 2) of the mean dose was observed azimuthally around the applicator surface, which can be explained by offset source positioning in the central channel.
Journal of Contemporary Brachytherapy | 2017
Ron S. Sloboda; Hali Morrison; Brie Cawston-Grant; Geetha Menon
Model-based dose calculation algorithms (MBDCAs) have recently emerged as potential successors to the highly practical, but sometimes inaccurate TG-43 formalism for brachytherapy treatment planning. So named for their capacity to more accurately calculate dose deposition in a patient using information from medical images, these approaches to solve the linear Boltzmann radiation transport equation include point kernel superposition, the discrete ordinates method, and Monte Carlo simulation. In this overview, we describe three MBDCAs that are commercially available at the present time, and identify guidance from professional societies and the broader peer-reviewed literature intended to facilitate their safe and appropriate use. We also highlight several important considerations to keep in mind when introducing an MBDCA into clinical practice, and look briefly at early applications reported in the literature and selected from our own ongoing work. The enhanced dose calculation accuracy offered by a MBDCA comes at the additional cost of modelling the geometry and material composition of the patient in treatment position (as determined from imaging), and the treatment applicator (as characterized by the vendor). The adequacy of these inputs and of the radiation source model, which needs to be assessed for each treatment site, treatment technique, and radiation source type, determines the accuracy of the resultant dose calculations. Although new challenges associated with their familiarization, commissioning, clinical implementation, and quality assurance exist, MBDCAs clearly afford an opportunity to improve brachytherapy practice, particularly for low-energy sources.
Medical Physics | 2016
Hali Morrison; Geetha Menon; Matthew P. Larocque; Hans-Sonke Jans; Ezekiel Weis; Ron S. Sloboda
PURPOSE To estimate the total dosimetric uncertainty at the tumor apex for ocular brachytherapy treatments delivered using 16 mm Collaborative Ocular Melanoma Study (COMS) and Super9 plaques loaded with (125)I seeds in order to determine the size of the apex margin that would be required to ensure adequate dosimetric coverage of the tumor. METHODS The total dosimetric uncertainty was assessed for three reference tumor heights: 3, 5, and 10 mm, using the Guide to the expression of Uncertainty in Measurement/National Institute of Standards and Technology approach. Uncertainties pertaining to seed construction, source strength, plaque assembly, treatment planning calculations, tumor height measurement, plaque placement, and plaque tilt for a simple dome-shaped tumor were investigated and quantified to estimate the total dosimetric uncertainty at the tumor apex. Uncertainties in seed construction were determined using EBT3 Gafchromic film measurements around single seeds, plaque assembly uncertainties were determined using high resolution microCT scanning of loaded plaques to measure seed positions in the plaques, and all other uncertainties were determined from the previously published studies and recommended values. All dose calculations were performed using plaque simulator v5.7.6 ophthalmic treatment planning system with the inclusion of plaque heterogeneity corrections. RESULTS The total dosimetric uncertainties at 3, 5, and 10 mm tumor heights for the 16 mm COMS plaque were 17.3%, 16.1%, and 14.2%, respectively, and for the Super9 plaque were 18.2%, 14.4%, and 13.1%, respectively (all values with coverage factor k = 2). The apex margins at 3, 5, and 10 mm tumor heights required to adequately account for these uncertainties were 1.3, 1.3, and 1.4 mm, respectively, for the 16 mm COMS plaque, and 1.8, 1.4, and 1.2 mm, respectively, for the Super9 plaque. These uncertainties and associated margins are dependent on the dose gradient at the given prescription depth, thus resulting in the changing uncertainties and margins with depth. CONCLUSIONS The margins determined in this work can be used as a guide for determining an appropriate apex margin for a given treatment, which can be chosen based on the tumor height. The required margin may need to be increased for more complex scenarios (mushroom shaped tumors, tumors close to the optic nerve, oblique muscle related tilt, etc.) than the simple dome-shaped tumor examined and should be chosen on a case-by-case basis. The sources of uncertainty contributing most significantly to the total dosimetric uncertainty are seed placement within the plaques, treatment planning calculations, tumor height measurement, and plaque tilt. This work presents an uncertainty-based, rational approach to estimating an appropriate apex margin.
Medical Physics | 2018
Hali Morrison; Geetha Menon; Matthew P. Larocque; Bob van Veelen; Yury Niatsetski; Ezekiel Weis; Ron S. Sloboda
Purpose To investigate the dose calculation accuracy of the Advanced Collapsed cone Engine (ACE) algorithm for ocular brachytherapy using a COMS plaque loaded with I‐125 seeds for two heterogeneous patient tissue scenarios. Methods The Oncura model 6711 I‐125 seed and 16 mm COMS plaque were added to a research version (v4.6) of the Oncentra® Brachy (OcB) treatment planning system (TPS) for dose calculations using ACE. Treatment plans were created for two heterogeneous cases: (a) a voxelized eye phantom comprising realistic eye materials and densities and (b) a patient CT dataset with variable densities throughout the dataset. ACE dose calculations were performed using a high accuracy mode, high‐resolution calculation grid matching the imported CT datasets (0.5 × 0.5 × 0.5 mm3), and a user‐defined CT calibration curve. The accuracy of ACE was evaluated by replicating the plan geometries and comparing to Monte Carlo (MC) calculated doses obtained using MCNP6. The effects of the heterogeneous patient tissues on the dose distributions were also evaluated by performing the ACE and MCNP6 calculations for the same scenarios but setting all tissues and air to water. Results Average local percent dose differences between ACE and MC within contoured structures and at points of interest for both scenarios ranged from 1.2% to 20.9%, and along the plaque central axis (CAX) from 0.7% to 7.8%. The largest differences occurred in the plaque penumbra (up to 17%), and at contoured structure interfaces (up to 20%). Other regions in the eye agreed more closely, within the uncertainties of ACE dose calculations (˜5%). Compared to that, dose differences between water‐based and fully heterogeneous tissue simulations were up to 27%. Conclusions Overall, ACE dosimetry agreed well with MC in the tumor volume and along the plaque CAX for the two heterogeneous tissue scenarios, indicating that ACE could potentially be used for clinical ocular brachytherapy dosimetry. In general, ACE data matched the fully heterogeneous MC data more closely than water‐based data, even in regions where the ACE accuracy was relatively low. However, depending on the plaque position, doses to critical structures near the plaque penumbra or at tissue interfaces were less accurate, indicating that improvements may be necessary. More extensive knowledge of eye tissue compositions is still required.
Medical Physics | 2016
Brie Cawston-Grant; Hali Morrison; Ron S. Sloboda; Geetha Menon
Purpose: To present an investigation of the Advanced Collapsed-cone Engine (ACE) in Oncentrae Brachy (OcB) v4.5 using a tissue equivalent phantom modeling scalp brachytherapy (BT) treatments. Methods: A slab phantom modeling the skin, skull, brain and mold was used. A dose of 400cGy was prescribed to just above the skull layer using TG-43 and was delivered using an HDR afterloader. Measurements were made using Gafchromic™ EBT3 film at four depths within the phantom. The TG-43 planned and film measured doses were compared to the standard (sACE) and high (hACE) accuracy ACE options in OcB between the surface and below the skull. Results: The average difference between the TG-43 calculated and film measured doses was −11.25±3.38% when there was no air gap between the mold and skin; sACE and hACE doses were on average lower than TG-43 calculated doses by 3.41±0.03% and 2.45±0.03%, respectively. With a 3mm air gap between the mold and skin, the difference between the TG-43 calculated and measured doses was −8.28±5.76%; sACE and hACE calculations yielded average doses 1.87±0.03% and 1.78±0.04% greater than TG-43, respectively. Conclusions: TG-43, sACE, and hACE were found to overestimate doses below the skull layer compared to film. With a 3mm air gap between the mold and skin, sACE and hACE more accurately predicted the film dose to the skin surface than TG-43. More clinical variations and their implications are currently being investigated.
Medical Physics | 2016
Hali Morrison; Geetha Menon; Ron S. Sloboda
Purpose: To investigate the accuracy of model-based dose calculations using a collapsed-cone algorithm for COMS eye plaques loaded with I-125 seeds. Methods: The Nucletron SelectSeed 130.002 I-125 seed and the 12 mm COMS eye plaque were incorporated into a research version of the Oncentra® Brachy v4.5 treatment planning system which uses the Advanced Collapsed-cone Engine (ACE) algorithm. Comparisons of TG-43 and high-accuracy ACE doses were performed for a single seed in a 30×30×30 cm3 water box, as well as with one seed in the central slot of the 12 mm COMS eye plaque. The doses along the plaque central axis (CAX) were used to calculate the carrier correction factor, T(r), and were compared to tabulated and MCNP6 simulated doses for both the SelectSeed and IsoAid IAI-125A seeds. Results: The ACE calculated dose for the single seed in water was on average within 0.62 ± 2.2% of the TG-43 dose, with the largest differences occurring near the end-welds. The ratio of ACE to TG-43 calculated doses along the CAX (T(r)) of the 12 mm COMS plaque for the SelectSeed was on average within 3.0% of previously tabulated data, and within 2.9% of the MCNP6 simulated values. The IsoAid and SelectSeed T(r) values agreed within 0.3%. Conclusions: Initial comparisons show good agreement between ACE and MC doses for a single seed in a 12 mm COMS eye plaque; more complicated scenarios are being investigated to determine the accuracy of this calculation method.
Medical Physics | 2015
Hali Morrison; Geetha Menon; Hans-Sonke Jans; Matthew P. Larocque; Ron S. Sloboda
Purpose: To determine the effect of geometric uncertainties in the seed positions in a COMS eye plaque on the central axis (CAX) dose. Methods: A Silastic insert was placed into a photopolymer 3D printed 16 mm COMS plaque, which was then positioned onto a custom-designed PMMA eye phantom. High resolution 3D images were acquired of the setup using a Siemens Inveon microPET/CT scanner. Images were acquired with the plaque unloaded and loaded with IsoAid I-125 seed shells (lack of silver core to minimize metal artifacts). Seed positions and Silastic thickness beneath each slot were measured. The measured seed coordinates were used to alter the seed positions within a standard 16 mm COMS plaque in Plaque Simulator v5.7.3 software. Doses along the plaque CAX were compared for the original and modified plaque coordinates using 3.5 mCi seeds with treatment times set to deliver 70 Gy to tumour apexes of 3.5, 5, and 10 mm height. Results: The majority of seeds showed length-wise displacement, and all seeds showed displacement radially outward from the eye center. The average radial displacement was 0.15 mm larger than the expected 1.4 mm offset, approximately half of which was due to increased Silastic thickness beneath each slot. The CAX doses for the modified seed positions were consistently lower for all tumour heights due to geometric displacement of the seeds; dose differences were found to increase to a maximum of 2.6% at a depth of ∼10 mm, after which they decreased due to the inverse square dose fall-off minimizing this effect. Conclusion: This work presents initial results of a broader dosimetric uncertainty evaluation for fully loaded COMS eye plaques and demonstrates the effects of seed positioning uncertainties. The small shifts in seed depths had noticeable effects on the CAX doses indicating the importance of careful Silastic loading. Funding provided by Alberta Cancer Foundation Grant #26655, Vanier Canada Graduate Scholarship, and Alberta Innovates Health Sciences Graduate Studentship
Medical Physics | 2014
Hali Morrison; Geetha Menon; Ron S. Sloboda
The purpose of this study was to investigate the accuracy of radiochromic film calibration procedures used in external beam radiotherapy when applied to I-125 brachytherapy sources delivering higher doses, and to determine any necessary modifications to achieve similar accuracy in absolute dose measurements. GafChromic EBT3 film was used to measure radiation doses upwards of 35 Gy from 6 MV, 75 kVp and (∼28 keV) I-125 photon sources. A custom phantom was used for the I-125 irradiations to obtain a larger film area with nearly constant dose to reduce the effects of film heterogeneities on the optical density (OD) measurements. RGB transmission images were obtained with an Epson 10000XL flatbed scanner, and calibration curves relating OD and dose using a rational function were determined for each colour channel and at each energy using a non-linear least square minimization method. Differences found between the 6 MV calibration curve and those for the lower energy sources are large enough that 6 MV beams should not be used to calibrate film for low-energy sources. However, differences between the 75 kVp and I-125 calibration curves were quite small; indicating that 75 kVp is a good choice. Compared with I-125 irradiation, this gives the advantages of lower type B uncertainties and markedly reduced irradiation time. To obtain high accuracy calibration for the dose range up to 35 Gy, two-segment piece-wise fitting was required. This yielded absolute dose measurement accuracy above 1 Gy of ∼2% for 75 kVp and ∼5% for I-125 seed exposures.
Medical Physics | 2018
Hali Morrison; Geetha Menon; Matthew P. Larocque; Bob van Veelen; Yury Niatsetski; Ezekiel Weis; Ron S. Sloboda