J Barbiere
Hackensack University Medical Center
Network
Latest external collaboration on country level. Dive into details by clicking on the dots.
Publication
Featured researches published by J Barbiere.
Medical Physics | 2016
G Beninati; J Barbiere; L Godfrey; A Ndlovu
PURPOSE To demonstrate that Volumetric Modulated Arc Therapy (VMAT) can be an alternative technique to Brachytherapy Accelerated Partial Breast Irradiation (APBI) for treating large breasted women. The non-coplanar VMAT technique uses a commercially available couch and a small number of angles. This technique with the patient in the prone position can reduce high skin and critical structure doses in large breasted women, which are usually associated with Brachytherapy APBI. METHODS Philips Pinnacle treatment planning system with Smart Arc was used to plan a left sided laterally located excision cavity on a standard prone breast patient setup. Three thirty-degree arcs entered from the lateral side at respective couch angles of 345, 0, and 15 degrees. A fourth thirty degree arc beam entered from the medial side at a couch angle of 0 degrees. The arcs were selected to avoid critical structures as much as possible. A test run was then performed to verify that the beams did not collide with the patient nor support structures. NSABP B-39/RTOG 0413 protocol guidelines were used for dose prescription, normal tissue, and target definition. RESULTS Dose Volume Histogram analysis indicated that all parameters were equal or better than RTOG recommendations. Of particular note regarding the plan quality:1.(a) For a prescribed dose of 3850cGy the PTV-EVAL target volume receiving 100 percent of the dose(V100) was 93; protocol recommendation is V90 > 90 percent. (b) Maximum dose was 110 percent versus the allowed 120 percent .2. Uninvolved percentage of normal breast V100 and V50 were 17 and 47 versus allowed 35 and 60 percent respectively.3. For the skin, V100 was 5.7cc and the max dose to 0.1 cc was 4190cGy. CONCLUSION Prone Breast non-coplanar VMAT APBI can achieve better skin cosmesis and lower critical structure doses than Brachytherapy APBI.
Medical Physics | 2016
J Barbiere; A Ndlovu
PURPOSE This work demonstrates clinical differences in a prostate IMRT plan DVH depending on which of four common algorithms is used for computation. Previous comparisons were often based on simple static or re-optimized IMRT plans which may not be clearly indicative of the clinical end result due solely to the computation algorithm. METHODS A step and shoot prostate multiple field IMRT plan was created in Pinnacle with a Collapsed Cone Convolution Superposition (CCCS) algorithm. The computed patient plan with corresponding dose and the control points without dose were both independently exported to Eclipse. Using the same patient image and structure data sets the control points were recomputed with the Anisotropic Analytical Algorithm (AAA) and Acuros XB for dose-to-water in medium (AXBwater) and dose-to-medium in medium (AXBmedium). All DVH values were measured in Eclipse to eliminated known differences between various DVH systems. A 0.3cm3 volume was created at the isocenter for point dose (PD) comparison. RESULTS The prostate target Acuros dose-to-medium in medium is clearly different than the others with a mean value of 101.7 cGy versus 103.8, 103.5 and 103.1 cGy for AAA, CCCS and AXBwater respectively. The difference in other clinical parameters such as volume at 105% dose (V105) is also unexpectedly different with values of 1.6, 23.9, 16.0 and 11.6 % respectively. The mean value in a small volume around the isocenter was 99.6, 102.4, 102.2 and 101.5 respectively which though not a great difference could affect the plan normalization and add to other effects in all regions of interest. The V102 is 25% larger for AAA compared to AXBwater. The variation in V50 for the bladder and rectum was approximately 5%. CONCLUSION This work demonstrates how an individual plan can be recomputed and compared directly between various algorithms and that clinical parameters can indeed vary according to which algorithm is used for computation.
Medical Physics | 2015
J Barbiere; G Beninati; A Ndlovu
Purpose: It has been argued that a 3D-conformal technique (3DCRT) is suitable for SBRT due to its simplicity for non-coplanar planning and delivery. It has also been hypothesized that a high dose delivered in a short time can enhance indirect cell death due to vascular damage as well as limiting intrafraction motion. Flattening Filter Free (FFF) photon beams are ideal for high dose rate treatment but their conical profiles are not ideal for 3DCRT. The purpose of our work is to present a method to efficiently segment an FFF beam for standard 3DCRT planning. Methods: A 10×10 cm Varian True Beam 6X FFF beam profile was analyzed using segmentation theory to determine the optimum segmentation intensity required to create an 8 cm uniform dose profile. Two segments were automatically created in sequence with a Varian Eclipse treatment planning system by converting isodoses corresponding to the calculated segmentation intensity to contours and applying the “fit and shield” tool. All segments were then added to the FFF beam to create a single merged field. Field blocking can be incorporated but was not used for clarity. Results: Calculation of the segmentation intensity using an algorithm originally proposed by Xia and Verhey indicated that each segment should extend to the 92% isodose. The original FFF beam with 100% at the isocenter at a depth of 10 cm was reduced to 80% at 4cm from the isocenter; the segmented beam had +/−2.5 % uniformity up to 4.4cm from the isocenter. An additional benefit of our method is a 50% decrease in the 80%-20% penumbra of 0.6cm compared to 1.2cm in the original FFF beam. Conclusion: Creation of two optimum segments can flatten a FFF beam and also reduce its penumbra for clinical 3DCRT SBRT treatment.
Medical Physics | 2015
J Barbiere; J Napoli; A Ndlovu
Purpose: Commonly Ir-192 HDR treatment planning system commissioning is only based on a single absolute measurement of source activity supplemented by tabulated parameters for multiple factors without independent verification that the planned distribution corresponds to the actual delivered dose. The purpose on this work is to present a methodology using Gafchromic film with a statistically valid calibration curve that can be used to validate clinical HDR vaginal cylinder cases by comparing the calculated plan dose distribution in a plane with the corresponding measured planar dose. Methods: A vaginal cylinder plan was created with Oncentra treatment planning system. The 3D dose matrix was exported to a Varian Eclipse work station for convenient extraction of a 2D coronal dose plane corresponding to the film position. The plan was delivered with a sheet of Gafchromic EBT3 film positioned 1mm from the catheter using an Ir-192 Nucletron HDR source. The film was then digitized with an Epson 10000 XL color scanner. Film analysis is performed with MatLab imaging toolbox. A density to dose calibration curve was created using TG43 formalism for a single dwell position exposure at over 100 points for statistical accuracy. The plan and measured film dose planes were registered using a known dwell position relative to four film marks. The plan delivered 500 cGy to points 2 cm from the sources. Results: The distance to agreement of the 500 cGy isodose between the plan and film measurement laterally was 0.5 mm but can be as much as 1.5 mm superior and inferior. The difference between the computed plan dose and film measurement was calculated per pixel. The greatest errors up to 50 cGy are near the apex. Conclusion: The methodology presented will be useful to implement more comprehensive quality assurance to verify patient-specific dose distributions
Medical Physics | 2013
J Barbiere; A Ndlovu
PURPOSE The purpose of this work is to present a technique by which we are able to measure the error produced by computing the delivered dose to a patient based on the dose delivered to a uniform density cylindrical phantom. METHODS ArcCHECK by Sun Nuclear (3DVH, Sun Nuclear Corp., Melbourne, FL) has the desirable ability to perform VMAT QA and configure the results in the form of a DVH similar to the original plan evaluation. In this work we examine the hypothesis that measured errors in a phantom translate very closely to those in a 3D patient. It is possible to create Verification Plans (VP) whereby the treatment fields for a lung patient are applied to both the patient data set (VPpatient) and a predefined phantom data set (VPphantom). The VP 360 arcs are subdivided into 30 degree sectors. The most useful feature of the VP is that we now have the ability to make changes. Each of the individual sectors is essentially an independent control point with alterable treatment aperture and Monitor Units. We modified a control point in the phantom (VPphantomModified) to see the effect on the patient (VPpatientModified). We tested the hypothesis in the two dimensional isocenter plane by exporting the four relevant dose matrices to MATLAB and performing the calculations as described by Sun Nuclear. RESULTS The amount of error greater then the traditional acceptance vale of 3% can be significant. Most of this error is concentrated in the lung and rib regions as would be expected from their greatest deviation from the uniform density in the phantom. CONCLUSION This work shows a technique whereby the user of Sun Nuclear ArcCHECK can verify and quantify the hypothesis that measured dose error in a homogenous phantom is applicable to a heterogeneous patient.
Medical Physics | 2013
J Barbiere; A Ndlovu
PURPOSE Present an algorithm that is capable of deconvoluting a single radiochromic film exposed in a cylindrical phantom into the basic control point components. Changes noted in the measured control points can be incorporated into a delivered treatment plan using a commercial treatment planning system to compare the delivered dose with the planned dose using DVH tools. METHODS A 180 degree VMAT verification plan was subdivided into control points every 5 degrees. Dose matrices for the individual control points were exported for analysis to a PC with MATLAB software. The dose in the center ring of the cylindrical phantom represents the simulated film measurement. The plan dose is equal to the sum of the control points. Mathematically, the control points are the basis vectors of thedelivered plan. The deconvolution algorithm utilizes linear optimization to first determine the coefficients of the vectors and then to check for changes in the individual control points. Changes in the control points were introduced to test the deconvolution algorithm. For development purposes only a single plane was investigated. Measurement of both the entrance and exit doses improves the accuracy and is also used to compute the control point angle. RESULTS (1) Measured film has superior resolution compared to electronic detectors, (2) the deconvolution algorithm can determine the individual delivered control point parameters, (3) the measured control points can be used to calculate the delivered dose distribution, and (4) the treatment planning system can be used to compare the plan and delivered dose distribution within the patient using its DVH tools. CONCLUSION Deconvolution of an integral film in a cylindrical phantom for a VMAT plan represents the delivered control points that can be entered into the treatment planning system for patient specific VMAT QA utilizing DVH tools.
Medical Physics | 2012
J Barbiere; A Ingenito; G Beninati; A Ndlovu
PURPOSE We present an analysis of the variation in acceptable SBRT lung plans with beam parameters. A figure of merit encompassing standard metrics is used for analytical comparison to determine the optimum plan quality. METHODS A set of optimization dose-volume constraints was formulated that consistently produced acceptable plans. Plans were normalized to deliver a prescription dose (PD) of 5000 cGy to 95% of the PTV volume. The Conformity Index (CI), Conformity Number (CN), and Gradient Index (GI), and mean GTV dose (MDgtv) were calculated. In SBRT hotspots near the target center are often deemed acceptable. The ratio MDgtv/PD is greater then 1.0 and larger values indicate that more dose is delivered where desired within the PTV. We combine the indices into a single figure of merit, FOM = (1/CI)*CN*(1/GI)*(MDgtv/PD), for which larger values indicate better plan quality dosimetrically. FOM values were normalized to 1.0 for the best plan. Twenty four plans were calculated for 6X, 6X flattening filter free (FFF), 10X, and 10X FFF photon beams. The gantry arc rotations were 0°-180° (180arc), 135°-30° (255arc), and 181°-179° (360arc). The couch angle was either 0° (coplanar) or +/- 15° (non-coplanar). RESULTS For the normal lung volume there was no significant variation in either mean dose or percent volume receiving 2000 cGy. However, the percent volume receiving 500 cGy varies significantly with energy and couch angle. Ninety six plan quality indices were tabulated. Overall, the 6X FFF non-coplanar beam with a 255 degree arc gave the best the figure of merit; it was 6.5% higher then nearest competitor largely due to superior conformality. CONCLUSIONS Individual plan quality indices were combined into a single figure of merit for various beam parameters that can be used to analytically select the optimum dosimetric plan.
Medical Physics | 2010
J Barbiere; J Napoli; J Hanley
Purpose: To present a stand‐alone RadiotherapyTreatment Planning simulator (RTPsim) that physicists can use for interactive classroom instruction and provide to students for independent study. Method and Materials: A PC with MATLAB programming language and the imaging toolbox is required. The RTPsim components are (1) a set of grayscale images for dose computation with regions of interest for dose area histogram analysis; (2) a double Gaussian pencil beam algorithm to create treatment fields; (3) dose distribution calculation routines for individual open or wedged beams and for a composite of weighted beams; and (4) a Graphical User Interface (GUI) to run the system with such user friendly features as command buttons to run algorithms, text or cursor variable input, list boxes for patient selection, and radio buttons for beam modifiers such as wedges. Results: The percent depth dose (%DD) for a typical 10×10 cm 6MV beam created by pencil beams closely approximates the actual clinical data. By computing and storing a correction factor table, the %DD values match exactly. The incorporation of divergence and a low dose region beyond the geometrical field edge result in excellent agreement in calculated versus clinical data for open beam profiles. Wedged field profiles show reasonable agreement with our clinical system. Several test cases from texts were reproduced accurately for interactive teaching purposes. As an example, an RTPsim wedged tangent field breast plan comparable to a clinical plan was produced. Variations in wedge, weighting, and normalization point can be demonstrated in real time. Use of a MATLAB GUI is suitable for planning demonstration; physicists and dosimetrists found the GUI convenient and easy to use. Conclusion: The user‐friendly RadiotherapyTreatment Planning simulator can simulate a clinical treatment planningcomputer with sufficient accuracy and speed for interactive educational purposes and independent learning.
Medical Physics | 2010
J Barbiere; J Hanley
Purpose: Derive a novel solution to the initial premise that IMRT fluence can be easily calculated using dose image backprojection. Method and Materials: Calculations were performed using MATLAB with the imaging toolbox using parallel beam geometry. A phantom matrix P was constructed with a concave target, simultaneous integrated boost, and organ at risk with respective desired percent doses of 100, 110, and 40. Let R:SP be the backprojection of sinogram S of P. Iterative reconstruction using standard Ratio Method converges on a solution by computing Si+1 = Si * (P / R:SiP). For the novel Corona Method, we compute the corona using iterative backprojection reconstruction using ratios on only the target volumes T, T’ = R:S0T, and then superimpose the organ at risk, P’ = T’ + OAR. Iterative backprojection on P’ is then performed by successive addition correction, Si+1 = Si‐1 + (SO − Si), where SO is the sinogram of P’. Any negative intensities are set to zero during the iteration process. Target coverage is improved using the Compensated Phantom Method by computing a revised target based on the results Z of the process described thus far. We then define a new target T” = T’/Z and superimpose the OAR, such that P” = T” + OAR. The final procedure is to perform additive iteration constrained to contain only positive intensities. Results: Ratio Method: Target and boost areas receive their respective dose goals but the organ at risk unacceptable. Corona Method : OAR acceptable but poor target coverage. Compensated Phantom Method : Acceptable target coverage and the OAR goal is achieved. Conclusion: Dose image backprojection with initial corona calculation significantly reduces the dose to the organ at risk while maintaining acceptable target coverage.
Medical Physics | 2009
J Hanley; J Napoli; J Barbiere; M Tuna; D Smith; L Godfrey
Purpose: To investigate the dosimetric aspects of robotically‐assisted minimally invasive brachytherapy.Materials and Methods: Using the Intuitive Surgical DaVinci S System, the following were inserted into a porcine abdominopelvic cavity through a 12 mm auxiliary port and sutured directly to the pelvic sidewall: three 6‐French catheters; the Xoft Axxent delivery system; a prototype minimally invasive applicator (MIA). Plans were generated in Nucletrons treatment planning software for an Ir‐192 source and the Xoft 50 kV source. The ability of the prototype MIA to reproducibly position catheters was tested using filmdosimetry. Other applicators were designed to allow an en face application of the source, and various catheter arrangements were investigated. For the en face application, the sensitivity of the delivered dose to random deviations from normal incidence of the catheters (⩽ 5 degrees) and regular catheter spacing were tested. Results: All three trials for introducing the catheters were performed with ease by the surgeon. After the apparatus was sutured into place, it maintained a fixed geometry. The dosimetric measurements using GafChromic EBT film showed a uniform distribution of dose. The treatment plan for the en‐face arrangement using the Xoft source with 16 catheters at random angles showed a deviation from the planned dose of ⩽ 0.39%. Conclusions: The feasibility of performing robotically‐assisted minimally invasive brachytherapy was demonstrated for techniques utilizing both parallel and en face geometries. A customized applicator was designed to deliver a uniform dose to the target. The reproducibility of the device was confirmed with filmdosimetry. For the geometries tested, the en face application was found to be insensitive to random deviations of the catheters of up to 5 degrees from normal incidence. These techniques may be used for minimally invasive intra‐operative or post‐operative brachytherapy procedures using real‐time or atlas‐based planning.