L.S. Fog
Copenhagen University Hospital
Network
Latest external collaboration on country level. Dive into details by clicking on the dots.
Publication
Featured researches published by L.S. Fog.
Medical Physics | 2013
Anna Fredh; Jonas Bengtsson Scherman; L.S. Fog; Per Munck af Rosenschöld
PURPOSE The purpose of the present study was to investigate the ability of commercial patient quality assurance (QA) systems to detect linear accelerator-related errors. METHODS Four measuring systems (Delta(4®), OCTAVIUS(®), COMPASS, and Epiqa™) designed for patient specific quality assurance for rotational radiation therapy were compared by measuring four clinical rotational intensity modulated radiation therapy plans as well as plans with introduced intentional errors. The intentional errors included increasing the number of monitor units, widening of the MLC banks, and rotation of the collimator. The measurements were analyzed using the inherent gamma evaluation with 2% and 2 mm criteria and 3% and 3 mm criteria. When applicable, the plans with intentional errors were compared with the original plans both by 3D gamma evaluation and by inspecting the dose volume histograms produced by the systems. RESULTS There was considerable variation in the type of errors that the various systems detected; the failure rate for the plans with errors varied between 0% and 72%. When using 2% and 2 mm criteria and 95% as a pass rate the Delta(4®) detected 15 of 20 errors, OCTAVIUS(®) detected 8 of 20 errors, COMPASS detected 8 of 20 errors, and Epiqa™ detected 20 of 20 errors. It was also found that the calibration and measuring procedure could benefit from improvements for some of the patient QA systems. CONCLUSIONS The various systems can detect various errors and the sensitivity to the introduced errors depends on the plan. There was poor correlation between the gamma evaluation pass rates of the QA procedures and the deviations observed in the dose volume histograms.
Acta Oncologica | 2014
Daniel Gasic; Lars Ohlhues; N. Patrik Brodin; L.S. Fog; Tobias Pommer; J.P. Bangsgaard; Per Munck af Rosenschöld
Abstract Background. Flattening filter-free (FFF) beams are an emerging technology that has not yet been widely implemented as standard practice in radiotherapy centers. To facilitate the clinical implementation of FFF, we attempted to elucidate the difference in plan quality and treatment delivery time compared to flattening filter beams (i.e. standard, STD) for several patient groups. We hypothesize that the treatment plan quality is comparable while the treatment delivery time of volumetric modulated arc therapy (VMAT) is considerably shorter using FFF beams, especially for stereotactic treatments. Methods. A total of 120 patients treated for head and neck (H&N) tumors, high-grade glioma, prostate cancer, early stage lung cancer and intra-cranial metastatic disease (both single and multiple metastases) were included in the study. For each cohort, 20 consecutive patients were selected. The plans were generated using STD- and FFF-VMAT for both 6 MV and 10 MV, and were compared with respect to plan quality, monitor units and delivery time using Wilcoxon signed rank tests. Results. For H&N and high-grade gliomas, there was a significant difference in homogeneity index in favor for STD-VMAT (p < 0.001). For the stereotactic sites there were no differences in plan conformity. Stereotactic FFF-VMAT plans required significantly shorter delivery time compared to STD-VMAT plans (p < 0.001) for higher dose per fraction, on average 54.5% for 6 MV and 71.4% for 10 MV. FFF-VMAT generally required a higher number of MU/Gy (p < 0.001), on average 7.0% for 6 MV and 8.4% for 10 MV. Conclusion. It was generally possible to produce FFF-VMAT plans with the same target dose coverage and doses to organs at risk as STD-VMAT plans. Target dose homogeneity tended to be somewhat inferior for FFF-VMAT for the larger targets investigated. For stereotactic radiotherapy, FFF-VMAT resulted in a considerable time gain while maintaining similar plan quality compared to STD beams.
Physics in Medicine and Biology | 2011
L.S. Fog; Jens F B Rasmussen; Marianne C. Aznar; Flemming Kjær-Kristoffersen; Ivan R. Vogelius; Svend Aage Engelholm; J.P. Bangsgaard
RapidArc® has become the treatment of choice for an increasing number of treatment sites in many clinics. The extensive use of multiple subfields in RapidArc® treatments presents unique challenges, especially for small targets treated in few fractions. In this work, very small static fields and subsequently RapidArc® and conventional plans for two targets (0.4 and 9.9 cm(3)) were investigated. Doses from static fields 1-4 MLC leaves (0.25-1.00 cm) wide, and larger fields with 1-4 MLC leaves closed in their centres, were measured using the portal dosimeter-based QA system EPIQA (v 1.3) and gafchromic film. RapidArc and conventional plans for two tumours were then measured using EPIQA, gafchromic EBT2 film and the phantom-based QA system Delta4. Eclipse 8.6 and 8.9, grid spacings of 1.25 and 2.50 mm and a Varian HD linac were used. For static fields one MLC leaf wide, the dose was underestimated by Eclipse by as much as 53% (v 8.6, 2.5 mm grid). Eclipse underestimated the dose downstream from a few MLC leaves closed in the centre of a large MLC field by as much as 30%. Eclipse consistently overestimated the width of the penumbra by about 100%. For the conventional plans, there was good agreement between the calculated and measured dose for the 9.9 cm(3) PTV, but a 10% underdose was observed for the 0.4 cm(3) PTV. For the RapidArc® plans, the measured dose for the 9.9 cm(3) PTV was in good agreement with the calculated one. However, for the 0.4 cm(3) PTV about 10% overdosing was detected (Eclipse v 8.6, 2.5 mm grid spacing). EPIQA data indicated that the measured dose profiles were overmodulated compared to the calculated one. The use of small subfields, typically a few MLC leaves wide, or larger fields with one or a few MLC leaves closed in its centre can result in significant errors in the dose calculation. The detector systems used vary in their ability to detect the discrepancies. Using a smaller grid size and newer version of Eclipse reduces the discrepancies observed in this work but does not eliminate them.
Radiotherapy and Oncology | 2016
C.A. Espensen; L.S. Fog; M. Aznar; Lena Specht; J.F. Kiilgaard
S69 ______________________________________________________________________________________________________ dose rate brachytherapy with Ir192 source used to deliver 10 Gy/1fr at 1 cm radius from the center of source. PTBD tube was replaced by 10 mm, non sheathed self expandable metallic modified Giantruco Z stent. In EBRT group, stenting, followed by EBRT(dose of 45Gy/25fr/ 5 weeks) by conformal technique to primary tumour and stent area. All the patients were given single agent 5-Fluorouracil chemotherapy 370 mg/m2 Day1-5 at 4 weekly for 6 cycles.
Radiotherapy and Oncology | 2018
C.A. Espensen; L.S. Fog; M. Aznar; A. Gothelf; J.F. Kiilgaard; Ane L Appelt
Radiotherapy and Oncology | 2018
M. Hemer; V. Gram; Marianne C. Aznar; H. Pappot; P. Sjøgren; L.S. Fog; A.L. Appelt
Radiotherapy and Oncology | 2018
S. Damkjaer; N.K. Jensen; L.S. Fog; Mirjana Josipovic
Physica Medica | 2018
Charlotte Espensen; L.S. Fog; Juliette Thariat; J. Hérault; Marianne C. Aznar; Celia Maschi; Jean-Pierre Caujolle; Jens Folke Kiilgaard; Ane L Appelt
Physica Medica | 2018
Sidsel Marie Skov Damkjær; Nikolaj K.G. Jensen; L.S. Fog; Mirjana Josipovic
Radiotherapy and Oncology | 2017
E. Ravnkilde; I.M. Jakobsen; Ane L Appelt; J.P. Bangsgaard; L.S. Fog