M.D. Falco
University of Rome Tor Vergata
Network
Latest external collaboration on country level. Dive into details by clicking on the dots.
Publication
Featured researches published by M.D. Falco.
Medical Physics | 2013
C. Di Venanzio; M. Marinelli; E. Milani; G. Prestopino; C. Verona; G. Verona-Rinati; M.D. Falco; Paolo Bagalà; R. Santoni; M. Pimpinella
PURPOSE To investigate the dosimetric properties of synthetic single crystal diamond based Schottky diodes under irradiation with therapeutic electron beams from linear accelerators. METHODS A single crystal diamond detector was fabricated and tested under 6, 8, 10, 12, and 15 MeV electron beams. The detector performances were evaluated using three types of commercial detectors as reference dosimeters: an Advanced Markus plane parallel ionization chamber, a Semiflex cylindrical ionization chamber, and a p-type silicon detector. Preirradiation, linearity with dose, dose rate dependence, output factors, lateral field profiles, and percentage depth dose profiles were investigated and discussed. RESULTS During preirradiation the diamond detector signal shows a weak decrease within 0.7% with respect to the plateau value and a final signal stability of 0.1% (1σ) is observed after about 5 Gy. A good linear behavior of the detector response as a function of the delivered dose is observed with deviations below ±0.3% in the dose range from 0.02 to 10 Gy. In addition, the detector response is dose rate independent, with deviations below 0.3% in the investigated dose rate range from 0.17 to 5.45 Gy∕min. Percentage depth dose curves obtained from the diamond detector are in good agreement with the ones from the reference dosimeters. Lateral beam profile measurements show an overall good agreement among detectors, taking into account their respective geometrical features. The spatial resolution of solid state detectors is confirmed to be better than that of ionization chambers, being the one from the diamond detector comparable to that of the silicon diode. A good agreement within experimental uncertainties was also found in terms of output factor measurements between the diamond detector and reference dosimeters. CONCLUSIONS The observed dosimetric properties indicate that the tested diamond detector is a suitable candidate for clinical electron beam dosimetry.
Physica Medica | 2016
Francesca Romana Giglioli; Lidia Strigari; Riccardo Ragona; Giuseppina R. Borzì; E. Cagni; C. Carbonini; Stefania Clemente; Rita Consorti; Randa El Gawhary; Marco Esposito; M.D. Falco; David Fedele; C. Fiandra; Maria Cristina Frassanito; Valeria Landoni; Gianfranco Loi; Elena Lorenzini; Maria Rosa Malisan; Carmelo Marino; Enrico Menghi; Barbara Nardiello; Roberta Nigro; Caterina Oliviero; Gabriella Pastore; Mariagrazia Quattrocchi; R. Ruggieri; Irene Redaelli; G. Reggiori; S. Russo; E. Villaggi
PURPOSE A large-scale multi-institutional planning comparison on lung cancer SABR is presented with the aim of investigating possible criticism in carrying out retrospective multicentre data analysis from a dosimetric perspective. METHODS Five CT series were sent to the participants. The dose prescription to PTV was 54Gy in 3 fractions of 18Gy. The plans were compared in terms of PTV-gEUD2 (generalized Equivalent Uniform Dose equivalent to 2Gy), mean dose to PTV, Homogeneity Index (PTV-HI), Conformity Index (PTV-CI) and Gradient Index (PTV-GI). We calculated the maximum dose for each OAR (organ at risk) considered as well as the MLD2 (mean lung dose equivalent to 2Gy). The data were stratified according to expertise and technology. RESULTS Twenty-six centers equipped with Linacs, 3DCRT (4% - 1 center), static IMRT (8% - 2 centers), VMAT (76% - 20 centers), CyberKnife (4% - 1 center), and Tomotherapy (8% - 2 centers) collaborated. Significant PTV-gEUD2 differences were observed (range: 105-161Gy); mean-PTV dose, PTV-HI, PTV-CI, and PTV-GI were, respectively, 56.8±3.4Gy, 14.2±10.1%, 0.70±0.15, and 4.9±1.9. Significant correlations for PTV-gEUD2 versus PTV-HI, and MLD2 versus PTV-GI, were observed. CONCLUSIONS The differences in terms of PTV-gEUD2 may suggest the inclusion of PTV-gEUD2 calculation for retrospective data inter-comparison.
BMC Cancer | 2013
Grazia Tortorelli; Luana Di Murro; Rosaria Barbarino; Sara Cicchetti; Daniela di Cristino; M.D. Falco; Dahlia Fedele; Gianluca Ingrosso; Dania Janniello; Pasquale Morelli; Alessandra Murgia; Elisabetta Ponti; Sara Terenzi; Barbara Tolu; Riccardo Santoni
BackgroundTo identify predictive factors of radiation-induced skin toxicity in breast cancer patients by the analysis of dosimetric and clinical factors.Methods339 patients treated between January 2007 and December 2010 are included in the present analysis. Whole breast irradiation was delivered with Conventional Fractionation (CF) (50Gy, 2.0/day, 25 fractions) and moderate Hypofractionated Schedule (HS) (44Gy, 2.75Gy/day, 16 fractions) followed by tumour bed boost. The impact of patient clinical features, systemic treatments and, in particular, dose inhomogeneities on the occurrence of different levels of skin reaction has been retrospectively evaluated.ResultsG2 and G3 acute skin toxicity were 42% and 13% in CF patients and 30% and 7.5% in HS patients respectively. The retrieval and revaluation of 200 treatment plans showed a strong correlation between areas close to the skin surface, with inhomogeneities >107% of the prescribed dose, and the desquamation areas as described in the clinical records.ConclusionsIn our experience dose inhomogeneity underneath G2 – G3 skin reactions seems to be the most important predictor for acute skin damage and in these patients more complex treatment techniques should be considered to avoid skin damage. Genetic polymorphisms too have to be investigated as possible promising candidates for predicting acute skin reactions.
Physica Medica | 2016
S. Russo; Laura Masi; Paolo Francescon; Maria Cristina Frassanito; M.L. Fumagalli; M. Marinelli; M.D. Falco; Anna Martinotti; M. Pimpinella; G. Reggiori; Gianluca Verona Rinati; Sabrina Vigorito; P. Mancosu
PURPOSE The aim of the present work was to evaluate small field size output factors (OFs) using the latest diamond detector commercially available, PTW-60019 microDiamond, over different CyberKnife systems. OFs were measured also by silicon detectors routinely used by each center, considered as reference. METHODS Five Italian CyberKnife centers performed OFs measurements for field sizes ranging from 5 to 60mm, defined by fixed circular collimators (5 centers) and by Iris(™) variable aperture collimator (4 centers). Setup conditions were: 80cm source to detector distance, and 1.5cm depth in water. To speed up measurements two diamond detectors were used and their equivalence was evaluated. MonteCarlo (MC) correction factors for silicon detectors were used for comparing the OF measurements. RESULTS Considering OFs values averaged over all centers, diamond data resulted lower than uncorrected silicon diode ones. The agreement between diamond and MC corrected silicon values was within 0.6% for all fixed circular collimators. Relative differences between microDiamond and MC corrected silicon diodes data for Iris(™) collimator were lower than 1.0% for all apertures in the totality of centers. The two microDiamond detectors showed similar characteristics, in agreement with the technical specifications. CONCLUSIONS Excellent agreement between microDiamond and MC corrected silicon diode detectors OFs was obtained for both collimation systems fixed cones and Iris(™), demonstrating the microDiamond could be a suitable detector for CyberKnife commissioning and routine checks. These results obtained in five centers suggest that for CyberKnife systems microDiamond can be used without corrections even at the smallest field size.
Medical Dosimetry | 2011
M.D. Falco; Davide Fontanarosa; Roberto Miceli; Alessandra Carosi; Riccardo Santoni; M. D'Andrea
Cone-beam X-ray volumetric imaging in the treatment room, allows online correction of set-up errors and offline assessment of residual set-up errors and organ motion. In this study the registration algorithm of the X-ray volume imaging software (XVI, Elekta, Crawley, United Kingdom), which manages a commercial cone-beam computed tomography (CBCT)-based positioning system, has been tested using a homemade and an anthropomorphic phantom to: (1) assess its performance in detecting known translational and rotational set-up errors and (2) transfer the transformation matrix of its registrations into a commercial treatment planning system (TPS) for offline organ motion analysis. Furthermore, CBCT dose index has been measured for a particular site (prostate: 120 kV, 1028.8 mAs, approximately 640 frames) using a standard Perspex cylindrical body phantom (diameter 32 cm, length 15 cm) and a 10-cm-long pencil ionization chamber. We have found that known displacements were correctly calculated by the registration software to within 1.3 mm and 0.4°. For the anthropomorphic phantom, only translational displacements have been considered. Both studies have shown errors within the intrinsic uncertainty of our system for translational displacements (estimated as 0.87 mm) and rotational displacements (estimated as 0.22°). The resulting table translations proposed by the system to correct the displacements were also checked with portal images and found to place the isocenter of the plan on the linac isocenter within an error of 1 mm, which is the dimension of the spherical lead marker inserted at the center of the homemade phantom. The registration matrix translated into the TPS image fusion module correctly reproduced the alignment between planning CT scans and CBCT scans. Finally, measurements on the CBCT dose index indicate that CBCT acquisition delivers less dose than conventional CT scans and electronic portal imaging device portals. The registration software was found to be accurate, and its registration matrix can be easily translated into the TPS and a low dose is delivered to the patient during image acquisition. These results can help in designing imaging protocols for offline evaluations.
Physics in Medicine and Biology | 2013
Paolo Bagalà; C. Di Venanzio; M.D. Falco; A.S. Guerra; M. Marinelli; E. Milani; M. Pimpinella; F. Pompili; G. Prestopino; R. Santoni; Alessia Tonnetti; C. Verona; G. Verona-Rinati
High-energy electron beams generated by linear accelerators, typically in the range 6 to 20 MeV, are used in small field sizes for radiotherapy of localized superficial tumors. Unshielded silicon diodes (Si-D) are commonly considered suitable detectors for relative dose measurements in small electron fields due to their high spatial resolution. Recently, a novel synthetic single crystal diamond diode (SCDD) showed suitable properties for standard electron beams and small photon beams dosimetry. The aim of the present study is twofold: to characterize 6 to 15 MeV small electron beams shaped by using commercial tubular applicators with 2, 3, 4 and 5 cm diameter and to assess the dosimetric performance under such irradiation conditions of the novel SCDD dosimeter by comparison with commercially available dosimeters, namely a Si-D and a plane–parallel ionization chamber. Percentage depth dose curves, beam profiles and output factors (OFs) were measured. A good agreement among the dosimeters was observed in all of the performed measurements. As for the tubular applicators, two main effects were evidenced: (i) OFs larger than unity were measured for a number of field sizes and energies, with values up to about 1.3, that is an output 30% greater than that obtained at the 10 × 10 cm2 reference field; (ii) for each diameter of the tubular applicator a noticeable increase of the OF values was observed with increasing beam energy, up to about 100% in the case of the smaller applicator. This OF behavior is remarkably different from what typically observed for small blocked fields having the same size and energy as those used in this study. OFs for tubular applicators depend considerably on the field size, so interpolation is unadvisable to predict the linear accelerator output for such applicators whereas reliable high-resolution detectors, as the silicon and diamond diodes used in this work allow OF measurements with uncertainties of about 1%.
Physica Medica | 2015
C. Di Venanzio; M. Marinelli; Alessia Tonnetti; G. Verona-Rinati; M.D. Falco; M. Pimpinella; Alessia Ciccotelli; S. De Stefano; Giuseppe Felici; F. Marangoni
PURPOSE To characterize a synthetic diamond dosimeter (PTW Freiburg microDiamond 60019) in high dose-per-pulse electron beams produced by an Intra Operative Radiation Therapy (IORT) dedicated accelerator. METHODS The dosimetric properties of the microDiamond were assessed under 6, 8 and 9 MeV electron beams by a NOVAC11 mobile accelerator (Sordina IORT Technologies S.p.A.). The characterization was carried out with dose-per-pulse ranging from 26 to 105 mGy per pulse. The microDiamond performance was compared with an Advanced Markus ionization chamber and a PTW silicon diode E in terms of dose linearity, percentage depth dose (PDD) curves, beam profiles and output factors. RESULTS A good linearity of the microDiamond response was verified in the dose range from 0.2 Gy to 28 Gy. A sensitivity of 1.29 nC/Gy was measured under IORT electron beams, resulting within 1% with respect to the one obtained in reference condition under (60)Co gamma irradiation. PDD measurements were found in agreement with the ones by the reference dosimeters, with differences in R50 values below 0.3 mm. Profile measurements evidenced a high spatial resolution of the microDiamond, slightly worse than the one of the silicon diode. The penumbra widths measured by the microDiamond resulted approximately 0.5 mm larger than the ones by the Silicon diode. Output factors measured by the microDiamond were found within 2% with those obtained by the Advanced Markus down to 3 cm diameter field sizes. CONCLUSIONS The microDiamond dosimeter was demonstrated to be suitable for precise dosimetry in IORT applications under high dose-per-pulse conditions.
Medical Physics | 2012
M.D. Falco; M. D'Andrea; Lidia Strigari; Daniela D'Alessio; Francesco Quagliani; Riccardo Santoni; Alessia Lo Bosco
PURPOSE During radiological interventional procedures (RIP) the skin of a patient under examination may undergo a prolonged x-ray exposure, receiving a dose as high as 5 Gy in a single session. This paper describes the use of the OneDose(TM) cable-free system based on p-type MOSFET detectors to determine the entrance skin dose (ESD) at selected points during RIP. METHODS At first, some dosimetric characteristics of the detector, such as reproducibility, linearity, and fading, have been investigated using a C-arc as a source of radiation. The reference setting (RS) was: 80 kV energy, 40 cm × 40 cm field of view (FOV), current-time product of 50 mAs and source to skin distance (SSD) of 50 cm. A calibrated PMX III solid state detector was used as the reference detector and Gafchromic(®) films have been used as an independent dosimetric system to test the entire procedure. A calibration factor for the RS and correction factors as functions of tube voltage and FOV size have been determined. RESULTS Reproducibility ranged from 4% at low doses (around 10 cGy as measured by the reference detector) to about 1% for high doses (around 2 Gy). The system response was found to be linear with respect to both dose measured with the PMX III and tube voltage. The fading test has shown that the maximum deviation from the optimal reading conditions (3 min after a single irradiation) was 9.1% corresponding to four irradiations in one hour read 3 min after the last exposure. The calibration factor in the RS has shown that the system response at the kV energy range is about four times larger than in the MV energy range. A fifth order and fourth order polynomial functions were found to provide correction factors for tube voltage and FOV size, respectively, in measurement settings different than the RS. ESDs measured with the system after applying the proper correction factors agreed within one standard deviation (SD) with the corresponding ESDs measured with the reference detector. The ESDs measured with Gafchromic(®) films were in agreement within one SD compared to the ESDs measured using the OneDose(TM) system, as well. The global uncertainty associated to the OneDose(TM) system established in our experiments, ranged from 7% to 10%, depending on the duration of the RIP due to fading. These values are much lower than the uncertainty commonly accepted for general diagnostic practices (20%) and of about the same size of the uncertainty recommended for practices with high risk of deterministic side effects (7%). CONCLUSIONS The OneDose(TM) system has shown a high sensitivity in the kV energy range and has been found capable of measuring the entrance skin dose in RIP.
Journal of Applied Clinical Medical Physics | 2015
M.D. Falco; Salvatore Masala; Matteo Stefanini; Roberto Fiori; Roberto Gandini; Paolo Bagalà; Daniele Morosetti; Eros Calabria; Alessia Tonnetti; G. Verona-Rinati; Riccardo Santoni; Giovanni Simonetti
The purpose of this work has been to dosimetrically investigate four fluoroscopically guided interventions: the percutaneous vertebroplasty (PVP), the percutaneous disc decompression (PDD), the radiofrequency medial branch neurolysis (RF) (hereafter named spine procedures), and the endovascular treatment for the critical limb ischemia (CLI). The X‐ray equipment used was a Philips Integris Allura Xper FD20 imaging system provided with a dose‐area product (DAP) meter. The parameters investigated were: maximum skin dose (MSD), air kerma (Ka,r), DAP, and fluoroscopy time (FT). In order to measure the maximum skin dose, we employed a system based on MOSFET detectors. Before using the system on patients, a calibration factor Fc and correction factors for energy (CkV) and field size (CFD) dependence were determined. Ka,r, DAP, and FT were extrapolated from the X‐ray equipment. The analysis was carried out on 40 patients, 10 for each procedure. The average fluoroscopy time and DAP values were compared with the reference levels (RLs) proposed in literature. Finally, the correlations between MSD, FT, Ka,r, and DAP values, as well as between DAP and FT values, were studied in terms of Pearsons product‐moment coefficients for spine procedures only. An Fc value of 0.20 and a very low dependence of CFD on field size were found. A third‐order polynomial function was chosen for CkV. The mean values of MSD ranged from 2.3 to 10.8 cGy for CLI and PVP, respectively. For these procedures, the DAP and FT values were within the proposed RL values. The statistical analysis showed little correlation between the investigated parameters. The interventional procedures investigated were found to be both safe with regard to deterministic effects and optimized for stochastic ones. In the spine procedures, the observed correlations indicated that the estimation of MSD from Ka,r or DAP was not accurate and a direct measure of MSD is therefore recommended. PACS number: 87
Journal of Applied Clinical Medical Physics | 2015
Cristina Di Venanzio; M. Marinelli; Alessia Tonnetti; G. Verona-Rinati; Paolo Bagalà; M.D. Falco; A.S. Guerra; M. Pimpinella
The purpose of this study was to compare the dosimetric properties of small field electron beams shaped by circular Cerrobend blocks and stainless steel tubular applicators. Percentage depth dose curves, beam profiles, and output factors of small‐size circular fields from 2 to 5 cm diameter, obtained either by tubular applicators and Cerrobend blocks, were measured for 6, 10, and 15 MeV electron beam energies. All measurements were performed using a PTW microDiamond 60019 premarket prototype. An overall similar behavior between the two collimating systems can be observed in terms of PDD and beam profiles. However, Cerrobend collimators produce a higher bremsstrahlung background under irradiation with high‐energy electrons. In such irradiation condition, larger output factors are observed for tubular applicators. Similar dosimetric properties are observed using circular Cerrobend blocks and stainless steel tubular applicators at lower beam energies. However, Cerrobend collimators allow the delivery of specific beam shapes, conformed to the target area. On the other hand, in high‐energy irradiation conditions, tubular applicators produce a lower bremsstrahlung contribution, leading to lower doses outside the target volume. In addition, the higher output factors observed at high energies for tubular applicators lead to reduced treatment times. PACS number: 87.53.Bn, 87.55.Qr, 87.56.Fc