M. Falk
Copenhagen University Hospital
Network
Latest external collaboration on country level. Dive into details by clicking on the dots.
Publication
Featured researches published by M. Falk.
Radiotherapy and Oncology | 2010
M. Falk; Per Munck af Rosenschöld; P Keall; Herbert Cattell; Byung Chul Cho; P.R. Poulsen; Sergey Povzner; Amit Sawant; Jens Zimmerman; S. Korreman
BACKGROUND AND PURPOSE Motion compensation with MLC tracking was tested for inversely optimized arc radiotherapy with special attention to the impact of the size of the target displacements and the angle of the leaf trajectory. MATERIALS AND METHODS An MLC-tracking algorithm was used to adjust the MLC positions according to the target movements using information from an optical real-time positioning management system. Two plans with collimator angles of 45 degrees and 90 degrees , respectively, were delivered and measured using the Delta(4)(R) dosimetric device moving in the superior-inferior direction with peak-to-peak displacements of 5, 10, 15, 20 and 25 mm and a cycle time of 6s. RESULTS Gamma index evaluation for plan delivery with MLC tracking gave a pass rate higher than 98% for criteria 3% and 3 mm for both plans and for all sizes of the target displacement. With no motion compensation, the average pass rate was 75% for plan 1 and 70% for plan 2 for 25 mm peak-to-peak displacement. CONCLUSION MLC tracking improves the accuracy of inversely optimized arc delivery for the cases studied. With MLC tracking, the dosimetric accuracy was independent of the magnitude of the peak-to-peak displacement of the target and not significantly affected by the angle between the leaf trajectory and the target movements.
Medical Physics | 2012
M. Falk; Tobias Larsson; P Keall; Byung Chul Cho; Marianne C. Aznar; S. Korreman; P.R. Poulsen; Per Munck af Rosenschöld
PURPOSE Real-time dynamic multileaf collimator (MLC) tracking for management of intrafraction tumor motion can be challenging for highly modulated beams, as the leaves need to travel far to adjust for target motion perpendicular to the leaf travel direction. The plan modulation can be reduced by using a leaf position constraint (LPC) that reduces the difference in the position of adjacent MLC leaves in the plan. The purpose of this study was to investigate the impact of the LPC on the quality of inversely optimized arc radiotherapy plans and the effect of the MLC motion pattern on the dosimetric accuracy of MLC tracking delivery. Specifically, the possibility of predicting the accuracy of MLC tracking delivery based on the plan modulation was investigated. METHODS Inversely optimized arc radiotherapy plans were created on CT-data of three lung cancer patients. For each case, five plans with a single 358° arc were generated with LPC priorities of 0 (no LPC), 0.25, 0.5, 0.75, and 1 (highest possible LPC), respectively. All the plans had a prescribed dose of 2 Gy × 30, used 6 MV, a maximum dose rate of 600 MU/min and a collimator angle of 45° or 315°. To quantify the plan modulation, an average adjacent leaf distance (ALD) was calculated by averaging the mean adjacent leaf distance for each control point. The linear relationship between the plan quality [i.e., the calculated dose distributions and the number of monitor units (MU)] and the LPC was investigated, and the linear regression coefficient as well as a two tailed confidence level of 95% was used in the evaluation. The effect of the plan modulation on the performance of MLC tracking was tested by delivering the plans to a cylindrical diode array phantom moving with sinusoidal motion in the superior-inferior direction with a peak-to-peak displacement of 2 cm and a cycle time of 6 s. The delivery was adjusted to the target motion using MLC tracking, guided in real-time by an infrared optical system. The dosimetric results were evaluated using gamma index evaluation with static target measurements as reference. RESULTS The plan quality parameters did not depend significantly on the LPC (p ≥ 0.066), whereas the ALD depended significantly on the LPC (p < 0.001). The gamma index failure rate depended significantly on the ALD, weighted to the percentage of the beam delivered in each control point of the plan (ALD(w)) when MLC tracking was used (p < 0.001), but not for delivery without MLC tracking (p ≥ 0.342). The gamma index failure rate with the criteria of 2% and 2 mm was decreased from > 33.9% without MLC tracking to <31.4% (LPC 0) and <2.2% (LPC 1) with MLC tracking. CONCLUSIONS The results indicate that the dosimetric robustness of MLC tracking delivery of an inversely optimized arc radiotherapy plan can be improved by incorporating leaf position constraints in the objective function without otherwise affecting the plan quality. The dosimetric robustness may be estimated prior to delivery by evaluating the ALD(w) of the plan.
Medical Physics | 2014
M. Falk; Tobias Pommer; P Keall; S. Korreman; G.F. Persson; P.R. Poulsen; Per Munck af Rosenschöld
PURPOSE To compare real-time dynamic multileaf collimator (MLC) tracking, respiratory amplitude and phase gating, and no compensation for intrafraction motion management during intensity modulated arc therapy (IMAT). METHODS Motion management with MLC tracking and gating was evaluated for four lung cancer patients. The IMAT plans were delivered to a dosimetric phantom mounted onto a 3D motion phantom performing patient-specific lung tumor motion. The MLC tracking system was guided by an optical system that used stereoscopic infrared (IR) cameras and five spherical reflecting markers attached to the dosimetric phantom. The gated delivery used a duty cycle of 35% and collected position data using an IR camera and two reflecting markers attached to a marker block. RESULTS The average gamma index failure rate (2% and 2 mm criteria) was <0.01% with amplitude gating for all patients, and <0.1% with phase gating and <3.7% with MLC tracking for three of the four patients. One of the patients had an average failure rate of 15.1% with phase gating and 18.3% with MLC tracking. With no motion compensation, the average gamma index failure rate ranged from 7.1% to 46.9% for the different patients. Evaluation of the dosimetric error contributions showed that the gated delivery mainly had errors in target localization, while MLC tracking also had contributions from MLC leaf fitting and leaf adjustment. The average treatment time was about three times longer with gating compared to delivery with MLC tracking (that did not prolong the treatment time) or no motion compensation. For two of the patients, the different motion compensation techniques allowed for approximately the same margin reduction but for two of the patients, gating enabled a larger reduction of the margins than MLC tracking. CONCLUSIONS Both gating and MLC tracking reduced the effects of the target movements, although the gated delivery showed a better dosimetric accuracy and enabled a larger reduction of the margins in some cases. MLC tracking did not prolong the treatment time compared to delivery with no motion compensation while gating had a considerably longer delivery time. In a clinical setting, the optical monitoring of the patients breathing would have to be correlated to the internal movements of the tumor.
Medical Physics | 2013
Tobias Pommer; M. Falk; P.R. Poulsen; P Keall; Ricky O’Brien; Per Munck af Rosenschöld
PURPOSE Intensity modulated arc therapy (IMAT) is commonly used to treat prostate cancer. The purpose of this study was to evaluate the impact of leaf width and plan complexity on dynamic multileaf collimator (DMLC) tracking for prostate motion management during IMAT treatments. METHODS Prostate IMAT plans were delivered with either a high-definition MLC (HDMLC) or a Millennium MLC (M-MLC) (0.25 and 0.50 cm central leaf width, respectively), with and without DMLC tracking, to a dosimetric phantom that reproduced four prostate motion traces. The plan complexity was varied by applying leaf position constraints during plan optimization. A subset of the M-MLC plans was converted for delivery with the HDMLC, isolating the effect of the different leaf widths. The gamma index was used for evaluation. Tracking errors caused by target localization, leaf fitting, and leaf adjustment were analyzed. RESULTS The gamma pass rate was significantly improved with DMLC tracking compared to no tracking (p < 0.001). With DMLC tracking, the average gamma index pass rate was 98.6% (range 94.8%-100%) with the HDMLC and 98.1% (range 95.4%-99.7%) with the M-MLC, using 3%, 3 mm criteria and the planned dose as reference. The corresponding pass rates without tracking were 87.6% (range 76.2%-94.7%) and 91.1% (range 81.4%-97.6%), respectively. Decreased plan complexity improved the pass rate when static target measurements were used as reference, but not with the planned dose as reference. The main cause of tracking errors was leaf fitting errors, which were decreased by 42% by halving the leaf width. CONCLUSIONS DMLC tracking successfully compensated for the prostate motion. The finer leaf width of the HDMLC improved the tracking accuracy compared to the M-MLC. The tracking improvement with limited plan complexity was small and not discernible when using the planned dose as reference.
7th International Conference on 3D Radiation Dosimetry (IC3DDose); 444, pp 012098-012098 (2013) | 2013
Sofie Ceberg; Crister Ceberg; M. Falk; P. Munk af Rosenschold; Sven Bäck
Respiratory motion during dynamic radiotherapy may affect the absorbed dose distribution both by dose-reducing smoothing and by more complicated interplay effects. In this study we present a novel method to determine the relative importance of these two effects. For the two dynamic deliveries studied in this work, the expected target dose reduction due to the smoothing effect was estimated by measurements convolved by the motion function. Remaining absorbed dose differences were attributed to interplay effects between the motion of the gel phantom and the movement of the modulating MLC leaves during modulated arc radiotherapy. The total dosimetric effect due to breathing motion and dynamic MLC motion during VMAT delivery resulted in an average of about 4% target dose reduction. Comparing with only the smoothing effect, the average difference was decreased to around 1%, and the remaining distribution was attributed to interplay effects. Although the interplay effects were small compared to the smoothing effect, the standard deviations of 1.4-2.3% (1SD) were larger than the narrow distribution for repeated stationary measurement with a standard deviation between 0.5-0.9% (1SD). (Less)
Journal of Physics: Conference Series | 2010
Sofie Ceberg; M. Falk; Per Munck af Rosenschöld; Herbert Cattell; Helen Gustafsson; P Keall; S. Korreman; Joakim Medin; Fredrik Nordström; G.F. Persson; Amit Sawant; Michelle Marie Svatos; Jens Zimmerman; Sven Bäck
The aim of this study was to carry out a dosimetric verification of a dynamic multileaf collimator (DMLC)-based tumor-tracking delivery during respiratory-like motion. The advantage of tumor-tracking radiation delivery is the ability to allow a tighter margin around the target by continuously following and adapting the dose delivery to its motion. However, there are geometric and dosimetric uncertainties associated with beam delivery system constraints and output variations, and several investigations have to be accomplished before a clinical integration of this tracking technique. Two types of delivery were investigated in this study I) a single beam perpendicular to a target with a one dimensional motion parallel to the MLC moving direction, and II) an intensity modulated arc delivery (RapidArc®) with a target motion diagonal to the MLC moving direction. The feasibility study (I) was made using an 2D ionisation chamber array and a true 3D polymer gel. The arc delivery (II) was verified using polymer gel and a biplanar diode array. Good agreement in absorbed dose was found between delivery to a static target and to a moving target with DMLC tracking using all three detector systems. However, due to the limited spatial resolution of the 2D array a detailed comparison was not possible. The RapidArc® plan delivery was successfully verified using the biplanar diode array and true 3D polymer gel, and both detector systems could verify that the DMLC-based tumor-tracking delivery system has a very good ability to account for respiratory target motion.
Physics in Medicine and Biology | 2013
Tobias Pommer; M. Falk; P.R. Poulsen; P Keall; R. O'Brien; P.M. Petersen; Per Munck af Rosenschöld
Radiotherapy and Oncology | 2011
M. Falk; P Keall; Byung Chul Cho; S. Korreman; P. Munck af Rosenschöld
Radiotherapy and Oncology | 2013
Sofie Ceberg; Crister Ceberg; M. Falk; P. Munck af Rosenschöld; Sven Bäck
Radiotherapy and Oncology | 2012
T. Larsson; M. Falk; P Keall; R. O'Brien; P. Munck af Rosenschöld