Network


Latest external collaboration on country level. Dive into details by clicking on the dots.

Hotspot


Dive into the research topics where Martin Hupfer is active.

Publication


Featured researches published by Martin Hupfer.


Radiology | 2014

Estimates of Effective Dose for CT Scans of the Lower Extremities

Natalia Saltybaeva; Mary Ellen Jafari; Martin Hupfer; Willi A. Kalender

PURPOSE To determine the dose-length product (DLP)-effective dose (ED) (DLP/ED) conversion coefficient (k) tables for the lower extremities that can be used for calculating ED. MATERIALS AND METHODS Dose calculations were performed on standard phantoms using a validated Monte Carlo calculation tool. Calculations were performed to obtain ED values for tube voltages from 80 kV to 140 kV in steps of 20 kV for the following examinations: hip (femur), knee, ankle, and computed tomographic (CT) angiography of the lower extremities. Values of the DLP were calculated by multiplying measured CT dose index values by the scan length; k values resulted as the quotients of the ED and DLP values. DLP/ED coefficients averaged over the range of voltage values and their standard deviations were determined for the given lower-extremity CT examinations for all age groups and for both sexes. RESULTS Coefficients depend strongly on the phantom age and size, but little on the kilovolt value. In the case of the newborn, for example, k values were 0.0612, 0.0046, 0.0014, and 0.047 for hip, knee, ankle, and CT angiography, respectively, while in the case of the adult, these respective values were 0.0110, 0.0004, 0.0002, and 0.0062. A substantial difference up to 20% between coefficients in male and female phantoms was observed for CT angiographic examination. CONCLUSION DLP/ED conversion coefficients are provided for lower extremities and allow estimation of ED for commonly used clinical musculoskeletal CT and CT angiographic protocols.


Medical Physics | 2011

Potential of high-Z contrast agents in clinical contrast-enhanced computed tomography

Tristan Nowak; Martin Hupfer; Robert Brauweiler; Fabian Eisa; Willi A. Kalender

PURPOSE Currently, only iodine- and barium-based contrast media (CM) are used in clinical contrast-enhanced computed tomography (CE-CT). High-Z metals would produce a higher contrast at equal mass density for the x-ray spectra used in clinical CT. Using such materials might allow for significant dose reductions in CE-CT. The purpose of this study was to quantify the potential for dose reduction when using CM based on heavy metals. METHODS The contrast-to-noise ratio weighted by dose (CNRD) was determined as a function of scan protocol by means of measurements and simulations on a clinical CT scanner. For simulations, water cylinders with diameters 160, 320, 480, and 640 mm were used to cover a broad range of patient sizes. Measurements were conducted with 160 and 320 mm water-equivalent plastic cylinders. A central bore of 13 mm diameter was present in all phantoms. The tube voltage was varied from 80 to 140 kV for measurements and from 60 to 180 kV for simulations. Additional tin filtration of thicknesses 0.4, 0.8, and 1.2 mm was applied in the simulation to evaluate a range of spectral hardness. The bore was filled with a mixture of water and 10 mg/ml of pure iodine, holmium, gadolinium, ytterbium, osmium, tungsten, gold, and bismuth for the simulations and with aqueous solutions of ytterbium, tungsten, gold, and bismuth salts as well as Iopromid containing 10 mg/ml of the pure materials for the measurements. CNRDs were compared to iodine at phantom size-dependent reference voltages for all high-Z materials and the resulting dose reduction was calculated for equal contrast-to-noise ratio. RESULTS Dose reduction potentials strongly depended on phantom size, spectral hardness, and tube voltage. Depending on the added filtration, a dose reduction of 19%-60% could be reached at 80 kV with gadolinium for the 160 mm phantom, 52%-69% at 100 kV with holmium for the 320 mm phantom, 62%-78% with 120 kV for hafnium and the 480 mm phantom and 74%-86% with 140 kV for gold and the 640 mm phantom. While gadolinium might be considered at 160 mm diameter, hafnium showed the best overall performance for phantom sizes of 320 mm and above. The measurements conducted on the clinical CT scanner showed very good agreement with simulations with deviations in the order of 5 to 10%. CONCLUSIONS The results of this study encourage the development and use of CM based on high-Z materials, especially for adipose patients, where high tube voltages are necessary to reach sufficiently short scan times. Hafnium proved to be the best compromise for average-size and for adipose patients. Even higher-Z materials such as gold and bismuth showed a good overall performance in conjunction with high tube voltage, large patients or strong added filtration and may be recommended for scans under these conditions.


Medical Physics | 2012

Dosimetry concepts for scanner quality assurance and tissue dose assessment in micro‐CT

Martin Hupfer; Daniel Kolditz; Tristan Nowak; Fabian Eisa; Robert Brauweiler; Willi A. Kalender

PURPOSE At present, no established methods exist for dosimetry in micro computed tomography (micro-CT). The purpose of this study was therefore to investigate practical concepts for both dosimetric scanner quality assurance and tissue dose assessment for micro-CT. METHODS The computed tomography dose index (CTDI) was adapted to micro-CT and measurements of the CTDI both free in air and in the center of cylindrical polymethyl methacrylate (PMMA) phantoms of 20 and 32 mm diameter were performed in a 6 month interval with a 100 mm pencil ionization chamber calibrated for low tube voltages. For tissue dose assessment, z-profile measurements using thermoluminescence dosimeters (TLDs) were performed and both profile and CTDI measurements were compared to Monte Carlo (MC) dose calculations to validate an existing MC tool for use in micro-CT. The consistency of MC calculations and TLD measurements was further investigated in two mice cadavers. RESULTS CTDI was found to be a reproducible quantity for constancy tests on the micro-CT system under study, showing a linear dependence on tube voltage and being by definition proportional to mAs setting and z-collimation. The CTDI measured free in air showed larger systematic deviations after the 6 month interval compared to the CTDI measured in PMMA phantoms. MC calculations were found to match CTDI measurements within 3% when using x-ray spectra measured at our micro-CT installation and better than 10% when using x-ray spectra calculated from semi-empirical models. Visual inspection revealed good agreement for all z-profiles. The consistency of MC calculations and TLD measurements in mice was found to be better than 10% with a mean deviation of 4.5%. CONCLUSIONS Our results show the CTDI implemented for micro-CT to be a promising candidate for dosimetric quality assurance measurements as it linearly reflects changes in tube voltage, mAs setting, and collimation used during the scan, encouraging further studies on a variety of systems. For tissue dose assessment, MC calculations offer an accurate and fast alternative to TLD measurements allowing for dose calculations specific to any geometry and scan protocol.


Investigative Radiology | 2016

High Atomic Number Contrast Media Offer Potential for Radiation Dose Reduction in Contrast-Enhanced Computed Tomography.

Ann-Christin Roessler; Martin Hupfer; Daniel Kolditz; Gregor Jost; Hubertus Pietsch; Willi A. Kalender

ObjectivesSpectral optimization of x-ray computed tomography (CT) has led to substantial radiation dose reduction in contrast-enhanced CT studies using standard iodinated contrast media. The purpose of this study was to analyze the potential for further dose reduction using high-atomic-number elements such as hafnium and tungsten. As in previous studies, spectra were determined for which the patient dose necessary to provide a given contrast-to-noise ratio (CNR) is minimized. Materials and MethodsWe used 2 different quasi-anthropomorphic phantoms representing the liver cross-section of a normal adult and an obese adult patient with the lateral widths of 360 and 460 mm and anterior-posterior heights of 200 and 300 mm, respectively. We simulated and measured on 2 different scanners with x-ray spectra from 80 to 140 kV and from 70 to 150 kV, respectively. We determined the contrast for iodine-, hafnium-, and tungsten-based contrast media, the noise, and 3-dimensional dose distributions at all available tube voltages by measurements and by simulations. The dose-weighted CNR was determined as optimization parameter. ResultsSimulations and measurements were in good agreement regarding their dependence on energy for all parameters investigated. Hafnium provided the best performance for normal and for obese patient phantoms, indicating a dose reduction potential of 30% for normal and 50% for obese patients at 120 kV compared with iodine; this advantage increased further with higher kV values. Dose-weighted CNR values for tungsten were always slightly below the hafnium results. Iodine proved to be the superior choice at voltage values of 80 kV and below. DiscussionHafnium and tungsten both seem to be candidates for contrast-medium-enhanced CT of normal and obese adult patients with strongly reduced radiation dose at unimpaired image quality. Computed tomography examinations of obese patients will decrease in dose for higher kV values.


Physica Medica | 2015

Comparative investigation of the detective quantum efficiency of direct and indirect conversion detector technologies in dedicated breast CT.

Jan Kuttig; Christian Steiding; Daniel Kolditz; Martin Hupfer; Marek Karolczak; Willi A. Kalender

PURPOSE To investigate the dose saving potential of direct-converting CdTe photon-counting detector technology for dedicated breast CT. MATERIALS AND METHODS We analyzed the modulation transfer function (MTF), the noise power spectrum (NPS) and the detective quantum efficiency (DQE) of two detector technologies, suitable for breast CT (BCT): a flat-panel energy-integrating detector with a 70 μm and a 208 μm thick gadolinium oxysulfide (GOS) and a 150 μm thick cesium iodide (CsI) scintillator and a photon-counting detector with a 1000 μm thick CdTe sensor. RESULTS The measurements for GOS scintillator thicknesses of 70 μm and 208 μm delivered 10% pre-sampled MTF values of 6.6 mm(-1) and 3.2 mm(-1), and DQE(0) values of 23% and 61%. The 10% pre-sampled MTF value for the 150 μm thick CsI scintillator 6.9 mm(-1), and the DQE(0) value was 49%. The CdTe sensor reached a 10% pre-sampled MTF value of 8.5 mm(-1) and a DQE(0) value of 85%. CONCLUSION The photon-counting CdTe detector technology allows for significant dose reduction compared to the energy-integrating scintillation detector technology used in BCT today. Our comparative evaluation indicates that a high potential dose saving may be possible for BCT by using CdTe detectors, without loss of spatial resolution.


Physics in Medicine and Biology | 2012

Optical tracking of contrast medium bolus to optimize bolus shape and timing in dynamic computed tomography

Fabian Eisa; Robert Brauweiler; Alexander Peetz; Martin Hupfer; Tristan Nowak; Willi A. Kalender

One of the biggest challenges in dynamic contrast-enhanced CT is the optimal synchronization of scan start and duration with contrast medium administration in order to optimize image contrast and to reduce the amount of contrast medium. We present a new optically based approach, which was developed to investigate and optimize bolus timing and shape. The time-concentration curve of an intravenously injected test bolus of a dye is measured in peripheral vessels with an optical sensor prior to the diagnostic CT scan. The curves can be used to assess bolus shapes as a function of injection protocols and to determine contrast medium arrival times. Preliminary results for phantom and animal experiments showed the expected linear behavior between dye concentration and absorption. The kinetics of the dye was compared to iodinated contrast medium and was found to be in good agreement. The contrast enhancement curves were reliably detected in three mice with individual bolus shapes and delay times of 2.1, 3.5 and 6.1 s, respectively. The optical sensor appears to be a promising approach to optimize injection protocols and contrast enhancement timing and is applicable to all modalities without implying any additional radiation dose. Clinical tests are still necessary.


Medical Physics | 2012

Spectral optimization for micro-CT

Martin Hupfer; Tristan Nowak; Robert Brauweiler; Fabian Eisa; Willi A. Kalender

PURPOSE To optimize micro-CT protocols with respect to x-ray spectra and thereby reduce radiation dose at unimpaired image quality. METHODS Simulations were performed to assess image contrast, noise, and radiation dose for different imaging tasks. The figure of merit used to determine the optimal spectrum was the dose-weighted contrast-to-noise ratio (CNRD). Both optimal photon energy and tube voltage were considered. Three different types of filtration were investigated for polychromatic x-ray spectra: 0.5 mm Al, 3.0 mm Al, and 0.2 mm Cu. Phantoms consisted of water cylinders of 20, 32, and 50 mm in diameter with a central insert of 9 mm which was filled with different contrast materials: an iodine-based contrast medium (CM) to mimic contrast-enhanced (CE) imaging, hydroxyapatite to mimic bone structures, and water with reduced density to mimic soft tissue contrast. Validation measurements were conducted on a commercially available micro-CT scanner using phantoms consisting of water-equivalent plastics. Measurements on a mouse cadaver were performed to assess potential artifacts like beam hardening and to further validate simulation results. RESULTS The optimal photon energy for CE imaging was found at 34 keV. For bone imaging, optimal energies were 17, 20, and 23 keV for the 20, 32, and 50 mm phantom, respectively. For density differences, optimal energies varied between 18 and 50 keV for the 20 and 50 mm phantom, respectively. For the 32 mm phantom and density differences, CNRD was found to be constant within 2.5% for the energy range of 21-60 keV. For polychromatic spectra and CMs, optimal settings were 50 kV with 0.2 mm Cu filtration, allowing for a dose reduction of 58% compared to the optimal setting for 0.5 mm Al filtration. For bone imaging, optimal tube voltages were below 35 kV. For soft tissue imaging, optimal tube settings strongly depended on phantom size. For 20 mm, low voltages were preferred. For 32 mm, CNRD was found to be almost independent of tube voltage. For 50 mm, voltages larger than 50 kV were preferred. For all three phantom sizes stronger filtration led to notable dose reduction for soft tissue imaging. Validation measurements were found to match simulations well, with deviations being less than 10%. Mouse measurements confirmed simulation results. CONCLUSIONS Optimal photon energies and tube settings strongly depend on both phantom size and imaging task at hand. For in vivo CE imaging and density differences, strong filtration and voltages of 50-65 kV showed good overall results. For soft tissue imaging of animals the size of a rat or larger, voltages higher than 65 kV allow to greatly reduce scan times while maintaining dose efficiency. For imaging of bone structures, usage of only minimum filtration and low tube voltages of 40 kV and below allow exploiting the high contrast of bone at very low energies. Therefore, a combination of two filtrations could prove beneficial for micro-CT: a soft filtration allowing for bone imaging at low voltages, and a variable stronger filtration (e.g., 0.2 mm Cu) for soft tissue and contrast-enhanced imaging.


Investigative Radiology | 2012

Development and Evaluation of a Phantom for Dynamic Contrast-enhanced Imaging

Robert Brauweiler; Fabian Eisa; Martin Hupfer; Tristan Nowak; Daniel Kolditz; Willi A. Kalender

ObjectivesDynamic contrast-enhanced imaging allows assessing functional information in addition to morphology using various modalities. Several applications have been established in clinical practice; however, there is no standard with respect to injection protocols or postprocessing algorithms. The purpose of this study was to develop a phantom for generating reproducible contrast-enhancement curves and providing a standard for comparison of different protocols and modalities in dynamic imaging. Materials and MethodsOur experimental setup consists of a peristaltic pump to generate a water flow through the phantom and a contrast injection pump. The phantom holds a sequence of layers allowing for assessment of perfusion, signal-to-noise ratio, and spatiotemporal resolution; the latter is the spatial resolution of structures with temporally changing contrast. Reproducibility was evaluated by the functional parameters time to peak, mean transit time, and peak enhancement by 24 scans over 4 weeks on a clinical computed tomography scanner. In addition, the area under the curve was evaluated for different injection durations at constant injection volume. Spatiotemporal resolution was assessed by spatial profiles on perfused bore patterns and compared for standard reconstructions, smooth reconstructions, and highly constrained backprojection for local reconstruction (HYPR LR). ResultsThe phantom showed good reproducibility in repeated measurements, with maximal deviations of 4% for time to peak, 9% for mean transit time, and 8% for peak enhancement. Area under the curve was constant within 3.5% for different injection protocols. For the static case, HYPR LR maintained spatial resolution. For dynamic objects, however, HYPR LR reduced spatial resolution dependent on temporal dynamics by up to 19% for highest dynamics, which was still superior to smooth reconstructions (27%). ConclusionsThe proposed phantom showed good reproducibility and therefore allows for comparing injection protocols or modalities in dynamic imaging. Assessment of spatiotemporal resolution under measurement conditions provides means for assessing postprocessing methods and reconstruction techniques in dynamic imaging.


Investigative Radiology | 2017

Dose Optimization for Computed Tomography Localizer Radiographs for Low-Dose Lung Computed Tomography Examinations.

Bernhard Schmidt; Martin Hupfer; Natalia Saltybaeva; Daniel Kolditz; Willi A. Kalender

Introduction Recent studies have shown a substantial reduction of radiation dose from computed tomography (CT) scans down to 0.1 mSv for lung cancer screening and cardiac examinations, when applying optimization techniques. Hence, CT localizer radiographs (LRs) might now be considered a significant contributor to the total dose of the CT examination. We investigated in our study the potential for reducing dose of the LRs by adapting the patient-specific acquisition parameters of the LR. Materials and Methods Localizer radiographs covering the lungs were acquired on 2 clinical scanners (64 slices, conventional detector [CD]; 96 slices, fully integrated detector [ID]) for 3 semianthropomorphic phantoms, representing a slim, a normal, and an obese adult. Starting at 120-kV tube voltage and 250-mA current were reduced until the image quality of the LR, and thereby the accuracy of the automatic exposure control was compromised; this was defined as a deviation of measured attenuation values in the center of the LR of more than 5% from the reference values measured at the highest tube voltage and current. Subsequent Monte Carlo calculations on anthropomorphic phantoms were performed to calculate organ and effective dose values for the respective optimal settings. In addition, effective dose values normalized to CTDIvol for tube voltages ranging from 60 to 160 kV were determined for the different combinations of phantom sizes, sexes, and LR views to evaluate dose efficiency. Results For the CD scanner, the optimal LR settings depended strongly on phantom size. Higher tube voltage and current were necessary for the larger phantoms. The ID scanner showed uncompromised LR quality for all phantoms using the lowest possible tube voltage-tube current combination of 80 kV and 20 mA. Depending on patient size and LR direction, effective dose values for the optimal settings ranged from 6 to 53 &mgr;Sv and 3 to 11 &mgr;Sv for the CD and ID scanner, respectively. For the example of an anterior-posterior LR on a normal patient, using the optimal settings instead of the standard settings on the ID scanner reduced LR dose from 53 &mgr;Sv (120 kV, 30 mA) to 10 &mgr;Sv (80 kV, 20 mA). The simulations for the different tube voltages show that effective dose and CTDIvol behave similarly for different views and patient sizes. However, the tube voltage level itself impacts the relationship between CTDIvol and effective dose, by up to a factor of 2. Discussion Dose from LRs may contribute significantly to the total effective dose of low-dose CT examinations such as lung cancer screening. Optimal LR settings can reduce LR dose substantially, but adaptations have to consider scanner characteristics, detector technology, and patient size. Thus, for low-dose CT examinations, such as cardiac examinations and lung cancer screening, LR optimization may result in a significant dose reduction and thereby in a substantial reduction of total dose.


Medical Physics | 2012

Time‐delayed summation as a means of improving resolution on fast rotating computed tomography systems

Tristan Nowak; Martin Hupfer; Felix Althoff; Robert Brauweiler; Fabian Eisa; Christian Steiding; Willi A. Kalender

PURPOSE Modern computed tomography (CT) systems are supporting increasingly fast rotation speeds, which are a prerequisite for fast dynamic acquisition, e.g. in perfusion imaging, and for new modalities such as dedicated breast CT, where breathhold scanning is indicated. However, not all detector technologies are supporting the high frame rates that are necessary to retain high resolution for objects far away from the isocenter. Even on systems that would support a sufficiently high frame rate, the necessary bandwidth of the data transfer from the rotating gantry stills remains challenging. The authors evaluated a pixel shifting technique termed time-delayed summation (TDS) as a method of increasing resolution on fast rotating CT systems without the need to increase the frame rate. METHODS In TDS mode, detector pixel values are shifted along rows during image acquisition to compensate for detector motion. In order to fully exploit TDS, focal spot position control (FSC) was used in combination with TDS. FSC applies a counter movement to the x-ray focal spot during image acquisition such that it is kept fixed in space. As a proof of concept, measurements were performed on a prototype photon counting detector capable of TDS. The detector was mounted on a movable table and a gold wire phantom was imaged with different TDS settings and detector velocities. Additionally, simulations of a broad range of TDS and FSC settings on two different modalities, a clinical CT scanner and a breast CT scanner, and two different detector geometries, flat and cylindrical, were performed to assess the gain in resolution and contrast in cylindrical water phantoms containing a small wire at distances from the phantom center varied from 5% to 90% of the phantom radius. As figures of merit, the modulation transfer function (MTF) at 10% and the maximum contrast were used and compared against the respective values when using step-and-shoot acquisition, which means stopping the rotation when a projection image is acquired. RESULTS Measurements showed that detector movement and the resulting blurring of the wire projections were compensated to the expected degree when using the appropriate number of TDS shifts per frame (TDS factor). Using simulations it was found that when using the optimal TDS factor, over 90% of the resolution achieved in step-and-shot mode was reached for all investigated wire positions. TDS showed better performance on a cylindrical detector that on the same system with a flat detector. TDS factors that were deviating from the optimum by more than 1 shift led to a performance below that of standard continuous acquisition. CONCLUSIONS The findings of this study encourage the combined usage of TDS and FSC in systems that require fast rotation. The integration of TDS in state-of-the-art x-ray detectors is feasible.

Collaboration


Dive into the Martin Hupfer's collaboration.

Top Co-Authors

Avatar

Willi A. Kalender

University of Erlangen-Nuremberg

View shared research outputs
Top Co-Authors

Avatar

Daniel Kolditz

University of Erlangen-Nuremberg

View shared research outputs
Top Co-Authors

Avatar

Fabian Eisa

University of Erlangen-Nuremberg

View shared research outputs
Top Co-Authors

Avatar

Robert Brauweiler

University of Erlangen-Nuremberg

View shared research outputs
Top Co-Authors

Avatar

Tristan Nowak

University of Erlangen-Nuremberg

View shared research outputs
Top Co-Authors

Avatar

Christian Steiding

University of Erlangen-Nuremberg

View shared research outputs
Top Co-Authors

Avatar

Felix Althoff

University of Erlangen-Nuremberg

View shared research outputs
Top Co-Authors

Avatar

Ferdinand Lück

University of Erlangen-Nuremberg

View shared research outputs
Top Co-Authors

Avatar
Top Co-Authors

Avatar

Jan Kuttig

University of Erlangen-Nuremberg

View shared research outputs
Researchain Logo
Decentralizing Knowledge