P von Voigts-Rhetz
Technische Hochschule Mittelhessen
Network
Latest external collaboration on country level. Dive into details by clicking on the dots.
Publication
Featured researches published by P von Voigts-Rhetz.
Physics in Medicine and Biology | 2013
Mathias Anton; R-P Kapsch; Achim Krauss; P von Voigts-Rhetz; Klemens Zink; M McEwen
In order to increase the usefulness of the alanine dosimeter as a tool for quality assurance measurements in radiotherapy using MV x-rays, the response with respect to the dose to water needs to be known accurately. This quantity is determined experimentally relative to (60)Co for 4, 6, 8, 10, 15 and 25 MV x-rays from two clinical accelerators. For the calibration, kQ factors for ionization chambers with an uncertainty of 0.31% obtained from calorimetric measurements were used. The results, although not inconsistent with a constant difference in response for all MV x-ray qualities compared to (60)Co, suggest a slow decrease from approximately 0.996 at low energies (4-6 MV) to 0.989 at the highest energy, 25 MV. The relative uncertainty achieved for the relative response varies between 0.35% and 0.41%. The results are confirmed by revised experimental data from the NRC as well as by Monte Carlo simulations using a density correction for crystalline alanine. By comparison with simulated and measured data, also for MeV electrons, it is demonstrated that the weak energy dependence can be explained by a transition of the alanine dosimeter (with increasing MV values) from a photon detector to an electron detector. An in-depth description of the calculation of the results and the corresponding uncertainty components is presented in an appendix for the interested reader. With respect to previous publications, the uncertainty budget had to be modified due to new evidence and to changes of the measurement and analysis method used at PTB for alanine/ESR.
Physics in Medicine and Biology | 2016
P von Voigts-Rhetz; Mathias Anton; H Vorwerk; Klemens Zink
In modern radiotherapy the verification of complex treatments plans is often performed in inhomogeneous or even anthropomorphic phantoms. For dose verification small detectors are necessary and therefore alanine detectors are most suitable. Though the response of alanine for a wide range of clinical photon energies in water is well know, the knowledge about the influence of the surrounding phantom material on the response of alanine is sparse. Therefore we investigated the influence of twenty different surrounding/phantom materials for alanine dosimeters in clinical photon fields via Monte Carlo simulations. The relative electron density of the used materials was in the range [Formula: see text] up to 1.69, covering almost all materials appearing in inhomogeneous or anthropomorphic phantoms used in radiotherapy. The investigations were performed for three different clinical photon spectra ranging from 6 to 25 MV-X and Co-60 and as a result a perturbation correction [Formula: see text] depending on the environmental material was established. The Monte Carlo simulation show, that there is only a small dependence of [Formula: see text] on the phantom material and the photon energy, which is below ±0.6%. The results confirm the good suitability of alanine detectors for in-vivo dosimetry.
Medical Physics | 2016
D Czarnecki; P von Voigts-Rhetz; B Poppe; Klemens Zink
PURPOSE The impact of removing the flattening filter on absolute dosimetry based on IAEAs TPR-398 and AAPMs TG-51 was investigated in this study using Monte Carlo simulations. METHODS The EGSnrc software package was used for all Monte Carlo simulations performed in this work. Five different ionization chambers and nine linear accelerator heads have been modeled according to technical drawings. To generate a flattening filter free radiation field the flattening filter was replaced by a 2 mm thick aluminum layer. Dose calculation in a water phantom were performed to calculate the beam quality correction factor kQ as a function of the beam quality specifiers %dd(10)x , TPR20,10 and mean photon and electron energies at the point of measurement in photon fields with (WFF) and without flattening filter (FFF). RESULTS The beam quality correction factor as a function of %dd(10)x differs systematically between FFF and WFF beams for all investigated ionization chambers. The largest difference of 1.8% was observed for the largest investigated Farmer-type ionization chamber with a sensitive volume of 0.69 cm3 . For ionization chambers with a smaller nominal sensitive volume (0.015 - 0.3 cm3 ) the deviation was less than 0.4% between WFF and FFF beams for %dd(10)x > 62%. The specifier TPR20,10 revealed only a good correlation between WFF and FFF beams (< 0.3%) for low energies. CONCLUSION The results confirm that %dd(10)x is a suitable beam quality specifier for FFF beams with an acceptable bias. The deviation depends on the volume of the ionization chamber. Using %dd(10)x to predictkQ for a large volume chamber in a FFF photon field may lead to not acceptable errors according to the results of this study. This bias may be caused by the volume effect due to the inhomogeneous photon fields of FFF linear accelerators.
Medical Physics | 2016
P von Voigts-Rhetz; T Pretzsch; D Czarnecki; Hilke Vorwerk; Klemens Zink
PURPOSE National and international dosimetry protocols assume a position accuracy for ionization chambers of less than 0.2mm. To follow this precept the manufacturer PTW-Freiburg introduced a positioning assistance system (TRUFIX) for their particular ion chambers. Aim of this study is an experimental investigation of the positioning uncertainties for ROOS-type ionization chambers. METHODS For all measurements a linear accelerator Elekta Synergie was used. The experiments were performed in a water-phantom. To collimate the electron beam a 10×10cm2 applicator was installed. All measured depth dose curves were normalized to their maximum. In all cases the TRUFIX system was applied for chamber positioning. For the first measurement series, to determine the positioning reproducibility of a ROOS chamber, one person placed the chamber three times in a 6 MeV electron beam. The mean value of this three measurements was the reference for further six random persons who repeated this procedure. The results were compared for different depths (R50 , zref and Rp ). To investigate the impact of different individual chambers of the same type 10 different ROOS chambers were placed by the same person in a 6, 12 and 18MeV electron beam and the measured reference depths zref were compared. RESULTS The absolute positioning reproducibility is less than 0.1mm for the same person. The positioning uncertainties are increasing up to +/-0.3mm if different persons perform the chambers positioning within the water phantom. The comparison of the 10 different ROOS chambers resulted in reference depths zref with deviations in the range of +/-0.45mm for all energies. CONCLUSION The position accuracy of 0.2mm can be fulfilled with the TRUFIX system. The comparison of the 10 different ROOS ionization chambers showed noticeable deviations in the determined reference depth. The impact of a positioning uncertainty of about 0.3-0.4mm on the total perturbation correction will be considered.
Medical Physics | 2015
N Langner; D Czarnecki; P von Voigts-Rhetz; Klemens Zink
Purpose: This paper aims to determine the effective point of measurement and the total perturbation correction p of parallel-plate chambers for clinical photon dosimetry. Methods: The effective point of measurement (EPOM) was calculated using the EGSnrc Monte Carlo code system with the EGSnrc user code egs_ chamber. Depth dose curves of the ionization chambers were calculated in a water phantom for several high energy photon spectra (4, 6, 10, 15, 18 MV-X). Different normalization criterions (normalization to the maximum of the depth dose curve and normalization to the value in 10 cm depth) have been applied. The EPOM was determined by shifting the normalized depth dose curve of a small water voxel against the depth ionization curve until the disagreement (calculated by the root mean square deviation) reaches a minimum. In addition, the total perturbation correction p was calculated by the ratio of the dose to water and the product of the dose determined in the chamber and the water to air stopping power ratio. Results: The EPOM varied slightly depending on the chosen normalization criterion. For all chambers the necessary shift of the EPOM decreased linearly with increasing beam quality specifier TPR20/10. For the Roos and NACP chamber, the results were positive suggesting that the chambers need to be shifted towards the focus. For the Markus chamber, the required shift was negative and for the Advanced Markus chamber partly negative and partly positive. The total perturbation correction p was almost independent of the depth. Only for regions below 1 cm the perturbation correction deviated significantly from unity. Conclusion: In the present study, the effective point of measurement and the total perturbation correction p was determined for four parallel-plate ionization chambers and five clinical relevant photon spectra. Applying the calculated EPOM, the residual perturbation correction p was mostly depth independent.
Medical Physics | 2015
D Czarnecki; P von Voigts-Rhetz; Klemens Zink
Purpose: Changing the characteristic of a photon beam by mechanically removing the flattening filter may impact the dose response of ionization chambers. Thus, perturbation factors of cylindrical ionization chambers in conventional and flattening filter free photon beams were calculated by Monte Carlo simulations. Methods: The EGSnrc/BEAMnrc code system was used for all Monte Carlo calculations. BEAMnrc models of nine different linear accelerators with and without flattening filter were used to create realistic photon sources. Monte Carlo based calculations to determine the fluence perturbations due to the presens of the chambers components, the different material of the sensitive volume (air instead of water) as well as the volume effect were performed by the user code egs_chamber. Results: Stem, central electrode, wall, density and volume perturbation factors for linear accelerators with and without flattening filter were calculated as a function of the beam quality specifier TPR20/10. A bias between the perturbation factors as a function of TPR20/10 for flattening filter free beams and conventional linear accelerators could not be observed for the perturbations caused by the components of the ionization chamber and the sensitive volume. Conclusion: The results indicate that the well-known small bias between the beam quality correction factor as a function of TPR20/10 for the flattening filter free and conventional linear accelerators is not caused by the geometry of the detector but rather by the material of the sensitive volume. This suggest that the bias for flattening filter free photon fields is only caused by the different material of the sensitive volume (air instead of water).
Medical Physics | 2014
D Czarnecki; P von Voigts-Rhetz; Klemens Zink
PURPOSE In current dosimetry protocols [AAPM TG51, IAEA TRS-389] the beam quality correction factor kQ and the water-to-air restricted mass collision stopping-power ratio SPR are related to beam quality specifiers %dd(10)x respectively TPR20,10 Determining kQ and SPR using these regular beam quality specifiers for conventional accelerators (WFF) and flattening filter free accelerators (FFF) similarly could lead to systemic bias. The influence of the flattening filter on the relationship between various beam quality specifiers and SPR respectively kQ was studied using Monte Carlo simulations with realistic beam sources. METHODS All Monte Carlo simulations were performed using the BEAMnrc/EGSnrc code system. Radiation transport through nine linear accelerator heads modeled according to technical drawings given by the manufactures and a 6 0 Co therapy source was simulated with BEAMnrc and then used as a radiation source for further simulations. FFF beam sources were implemented by removing the fattening filter from the WFF model. SPR was calculated applying the user code SPRRZnrc. The mean photon energy below the accelerator head and the mean energies of photons and electrons at the measuring point within the water phantom were calculated using FLURZnrc. Dose calculations within a small water voxel and the thimble ionization chamber PTW-31010 in a water depth of 10 cm were made using the egs_chamber code. RESULTS SPR and kQ as a function of fluence spectra based beam quality specifiers as well as conventional beam quality specifiers differ systematically between FFF and WFF beams. According to the results the specifier %dd(10)x revealed the smallest deviation (max. 0.4%) between FFF and WFF beams. CONCLUSION The results show that %dd(10)x is an acceptable beam quality specifier for FFF beams. Nevertheless the results confirm the expected bias between FFF and WFF beams which must by further investigated.
Medical Physics | 2014
Klemens Zink; D Czarnecki; H. K. Looe; P von Voigts-Rhetz; Dietrich Harder
Physics in Medicine and Biology | 2015
Mathias Anton; Thomas Hackel; Klemens Zink; P von Voigts-Rhetz; H-J Selbach
Medical Physics | 2015
D Czarnecki; P von Voigts-Rhetz; D Uchimura Shishechian; Klemens Zink