Network


Latest external collaboration on country level. Dive into details by clicking on the dots.

Hotspot


Dive into the research topics where Silvio Bruni Herdade is active.

Publication


Featured researches published by Silvio Bruni Herdade.


Medical Physics | 2004

X-ray spectroscopy in mammography with a silicon PIN photodiode with application to the measurement of tube voltage

Roseli Künzel; Silvio Bruni Herdade; Ricardo Andrade Terini; Paulo Roberto Costa

In this work a silicon PIN photodiode was employed in mammographic x-ray spectroscopy under clinical and nonclinical conditions. Measurements have been performed at a constant potential tungsten anode tube, adapted in this work with molybdenum filters to produce a beam like that used in mammography, and at a clinical equipment with a molybdenum anode tube by using an additional aluminum filtration. The corrected x-ray spectra were in full agreement with those generated by theoretical models published in the literature and agree well with those measured with a CdZnTe detector for tube voltages less than 30 kV. The half value layer and the relative exposure values calculated from the corrected silicon PIN photodiode spectra were in agreement with those measured with an ionization chamber. These results indicate that a silicon PIN photodiode are very suitable for mammographic x-ray spectroscopy. As an application, the voltage (kV) applied to mammographic x-ray equipment has been measured through the evaluation of the spectra high energy cut off. Uncertainties evaluated for the voltage values calculated from the measured spectra are less than 0.13% for voltages in the range 20-35 kV. The low uncertainties associated with the obtained results in this work point out that the method employed can be accurately used for calibration of noninvasive mammographic kVp meters.


Medical Physics | 2000

Determination of the voltage applied to x-ray tubes from the bremsstrahlung spectrum obtained with a silicon PIN photodiode

Márcia C. Silva; Silvio Bruni Herdade; Patrı́cia Lammoglia; Paulo Roberto Costa; Ricardo Andrade Terini

This work describes a methodology for the obtainment of the electron accelerating potential (kVp) applied to an x-ray tube, through the determination of the end point of the energy spectrum of the radiation emitted by the tube. The measurements have been performed utilizing alternatively two silicon PIN photodiodes, directly irradiated by the x-ray beam. Both were operated at room temperature, with low bias, so avoiding the drawbacks presented by photomultiplier tubes and germanium detectors. The energy calibration of the system was performed with X- and gamma-emitter radioactive sources, which makes the method absolute. Each kVp value was determined by means of a linear regression in the end of the spectrum, to give, simultaneously, a good fit of the straight line to the experimental data and a low standard deviation for the kVp value. Results of the measurements carried out with an x-ray tube connected to a three-phase generator, using additional filtration between 1.5 and 4.0 mm of Cu, are presented. This filtration was used in order to minimize the contribution of low energy photons and to reduce pulse pile-up. Errors determined for the values of kVp are between 0.06 and 0.16 kV, in the potential range from 50 to 100 kV. As an example, the methodology has been applied in the verification of the secondary calibration of a voltage divider, utilized, by its turn, in the tertiary calibration of noninvasive kVp meters. All intrinsic sources of errors involved in the process are discussed and evaluated.


Physics in Medicine and Biology | 2006

Ambient dose equivalent and effective dose from scattered x-ray spectra in mammography for Mo/Mo, Mo/Rh and W/Rh anode/filter combinations

Roseli Künzel; Silvio Bruni Herdade; Paulo Roberto Costa; Ricardo Andrade Terini; R S Levenhagen

In this study, scattered x-ray distributions were produced by irradiating a tissue equivalent phantom under clinical mammographic conditions by using Mo/Mo, Mo/Rh and W/Rh anode/filter combinations, for 25 and 30 kV tube voltages. Energy spectra of the scattered x-rays have been measured with a Cd(0.9)Zn(0.1)Te (CZT) detector for scattering angles between 30 degrees and 165 degrees . Measurement and correction processes have been evaluated through the comparison between the values of the half-value layer (HVL) and air kerma calculated from the corrected spectra and measured with an ionization chamber in a nonclinical x-ray system with a W/Mo anode/filter combination. The shape of the corrected x-ray spectra measured in the nonclinical system was also compared with those calculated using semi-empirical models published in the literature. Scattered x-ray spectra measured in the clinical x-ray system have been characterized through the calculation of HVL and mean photon energy. Values of the air kerma, ambient dose equivalent and effective dose have been evaluated through the corrected x-ray spectra. Mean conversion coefficients relating the air kerma to the ambient dose equivalent and to the effective dose from the scattered beams for Mo/Mo, Mo/Rh and W/Rh anode/filter combinations were also evaluated. Results show that for the scattered radiation beams the ambient dose equivalent provides an overestimate of the effective dose by a factor of about 5 in the mammography energy range. These results can be used in the control of the dose limits around a clinical unit and in the calculation of more realistic protective shielding barriers in mammography.


Radiologia Brasileira | 2009

A medição da grandeza practical peak voltage na prática radiológica

Ricardo Andrade Terini; M.P.A. Potiens; Silvio Bruni Herdade; Marco A. Pereira; João dos Santos Justo Pires; Heber Simões Videira

OBJECTIVE: The present study was aimed at evaluating the practical peak voltage (PPV) determined from the voltage waveform applied to x-ray tubes and comparing it with some kVp definitions for different types of x-ray equipment: single-phase (full-wave) and three-phase (six-pulse) clinical x-ray generators, and an industrial constant potential apparatus. MATERIALS AND METHODS: The study involved the comparison between invasively measured PPV (with voltage dividers) and values obtained with two commercial noninvasive meters, besides values of other quantities utilized for measuring the x-ray tube peak voltage. The PPV variation with the voltage ripple was also analyzed in the present study. RESULTS: The authors observed that the difference between PPV and the most common peak voltage definitions increases with the ripple. PPV values varied up to 3% and 5%, respectively, in the comparison between invasive and non-invasive measurements with single-phase and three-phase devices. CONCLUSION: The results demonstrated that voltage ripple is the main quantity influencing the invasive or non-invasive PPV determination. Additionally, non-invasively measured PPV values should be evaluated taking into consideration their dependence on the data sample rate and waveform obtained by the device.


Medical Physics | 2008

X-ray spectroscopy applied to radiation shielding calculation in mammography

Roseli Künzel; Ronaldo Savarino Levenhagen; Silvio Bruni Herdade; Ricardo Andrade Terini; Paulo Roberto Costa

The protective shielding design of a mammography facility requires the knowledge of the scattered radiation by the patient and image receptor components. The shape and intensity of secondary x-ray beams depend on the kVp applied to the x-ray tube, target/filter combination, primary x-ray field size, and scattering angle. Currently, shielding calculations for mammography facilities are performed based on scatter fraction data for Mo/Mo target/filter, even though modern mammography equipment is designed with different anode/filter combinations. In this work we present scatter fraction data evaluated based on the x-ray spectra produced by a Mo/Mo, Mo/Rh and W/Rh target/filter, for 25, 30 and 35 kV tube voltages and scattering angles between 30 and 165 degrees. Three mammography phantoms were irradiated and the scattered radiation was measured with a CdZnTe detector. The primary x-ray spectra were computed with a semiempirical model based on the air kerma and HVL measured with an ionization chamber. The results point out that the scatter fraction values are higher for W/Rh than for Mo/Mo and Mo/Rh, although the primary and scattered air kerma are lower for W/Rh than for Mo/Mo and Mo/Rh target/filter combinations. The scatter fractions computed in this work were applied in a shielding design calculation in order to evaluate shielding requirements for each of these target/filter combinations. Besides, shielding requirements have been evaluated converting the scattered air kerma from mGy/week to mSv/week adopting initially a conversion coefficient from air kerma to effective dose as 1 Sv/Gy and then a mean conversion coefficient specific for the x-ray beam considered. Results show that the thickest barrier should be provided for Mo/Mo target/filter combination. They also point out that the use of the conversion coefficient from air kerma to effective dose as 1 Sv/Gy is conservatively high in the mammography energy range and overestimate the barrier thickness.


Applied Radiation and Isotopes | 2011

A portable Compton spectrometer for clinical X-ray beams in the energy range 20-150 keV.

A.A. Vieira; A. Linke; E.M. Yoshimura; Ricardo Andrade Terini; Silvio Bruni Herdade

Primary beam spectra were obtained for an X-ray industrial equipment (40-150 kV), and for a clinical mammography apparatus (25-35 kV) from beams scattered at angles close to 90°, measured with a CdTe Compton spectrometer. Actual scattering angles were determined from the Compton energy shift of characteristic X-rays or spectra end-point energy. Evaluated contribution of coherent scattering amounts to more than 15% of fluence in mammographic beams. This technique can be used in clinical environments.


Radiologia Brasileira | 2013

Doses monitoring in radiology: calibration of air kerma-area product (PKA) meters

Ricardo Andrade Terini; Maria Carolina de Santana Campelo; José N. Almeida; Silvio Bruni Herdade; Marco A. Pereira

Objective The authors have sought to study the calibration of a clinical PKA meter (Diamentor E2) and a calibrator for clinical meters (PDC) in the Laboratory of Ionizing Radiation Metrology at Instituto de Energia e Ambiente - Universidade de Sao Paulo. Materials and Methods Different qualities of both incident and transmitted beams were utilized in conditions similar to a clinical setting, analyzing the influence from the reference dosimeter, from the distance between meters, from the filtration and from the average beam energy. Calibrations were performed directly against a standard 30 cm3 cylindrical chamber or a parallel-plate monitor chamber, and indirectly against the PDC meter. Results The lowest energy dependence was observed for transmitted beams. The cross calibration between the Diamentor E2 and the PDC meters, and the PDC presented the greatest propagation of uncertainties. Conclusion The calibration coefficient of the PDC meter showed to be more stable with voltage, while the Diamentor E2 calibration coefficient was more variable. On the other hand, the PDC meter presented greater uncertainty in readings (5.0%) than with the use of the monitor chamber (3.5%) as a reference.


Archive | 2009

Evaluation of Radiation Doses Inside a Phantom of Mammography Utilizing Compton Spectrometry

José N. Almeida; Ricardo Andrade Terini; Silvio Bruni Herdade; T. C. Furquim

Absorbed dose in breast cannot be measured directly in mammography equipments. Therefore, methods based on Compton spectrometry are alternatives to evaluate dose distributions in a simulator of human breast material. In this work, a CdTe detector was used for the Compton spectrometry measurements of radiation scattered at approximately 90° by a PMMA cylinder. To reach the objective, the necessary reconstruction of primary spectra (in units of photon fluence) has been made using Klein-Nishina and Compton theories, followed by a determination of air kerma and, finally, of absorbed dose values in the breast phantom. Incident X-ray spectra and depth-dose distributions in a BR-12 phantom were determined. Results presented here were obtained for mammography range (28 to 35 kV) and show good agreement with literature data.


Radiologia Brasileira | 2005

Protocolo de testes de aceitação em equipamentos de imagem por ressonância magnética

Alessandro André Mazzola; Silvio Bruni Herdade; Hilton Augusto Koch; Antonio Carlos Pires Carvalho

OBJETIVO: Este trabalho tem como objetivo criar um protocolo de testes de aceitacao para equipamentos de imagem por ressonância magnetica e demonstrar como e quais tipos de dispositivos de teste podem ser usados para a coleta de dados. MATERIAIS E METODOS: Para cada um dos 15 testes selecionados foram elaborados a definicao, o procedimento, a forma de analise e o criterio de aceitacao. RESULTADOS: Atraves dos testes de aceitacao descritos e possivel verificar caracteristicas tecnicas que constam nas propostas de venda dos fabricantes, assim como estabelecer valores de referencias para serem utilizados em posteriores testes de constância. CONCLUSAO: Futuros programas de garantia da qualidade em imagem por ressonância magnetica devem considerar testes semelhantes ou iguais aos descritos neste trabalho.


Revista Brasileira de Física Médica | 2010

Espectrometria Compton aplicada à medição de doses em mamografia para combinações W/Mo e Mo/Mo

José N. Almeida; Ricardo Andrade Terini; Silvio Bruni Herdade; Tânia A. C. Furquim

A dose glandular media (DGM) nao pode ser medida diretamente em equipamentos mamograficos. Metodos baseados na espectrometria Compton sao alternativas para avaliar distribuicoes de dose em um simulador de mamas, tornando possivel a estimativa da DGM. Neste trabalho, um detector de CdTe foi usado para as medicoes espectrometricas de feixes de raios X espalhados a aproximadamente 90o , por efeito Compton, por um cilindro de PMMA, inserido entre as placas de um simulador de mama. A reconstrucao do espectro dos feixes primarios, a partir dos medidos, foi feita utilizando a teoria de Klein-Nishina e o formalismo de Compton, seguida pela determinacao do kerma no ar incidente na superficie do simulador, da dose absorvida e, finalmente, da DGM. Os espectros dos feixes de raios X incidentes e as distribuicoes de dose em profundidade em um simulador de BR-12 foram determinados para a faixa de mamografia (28 a 35kV). Os resultados mostraram bom acordo com dados da literatura, obtidos com TLD.

Collaboration


Dive into the Silvio Bruni Herdade's collaboration.

Top Co-Authors

Avatar
Top Co-Authors

Avatar
Top Co-Authors

Avatar

José N. Almeida

Pontifícia Universidade Católica de São Paulo

View shared research outputs
Top Co-Authors

Avatar

Roseli Künzel

University of São Paulo

View shared research outputs
Top Co-Authors

Avatar
Top Co-Authors

Avatar
Top Co-Authors

Avatar
Top Co-Authors

Avatar
Top Co-Authors

Avatar

R S Levenhagen

University of São Paulo

View shared research outputs
Top Co-Authors

Avatar

A. Linke

University of São Paulo

View shared research outputs
Researchain Logo
Decentralizing Knowledge