Network


Latest external collaboration on country level. Dive into details by clicking on the dots.

Hotspot


Dive into the research topics where Sumche Man is active.

Publication


Featured researches published by Sumche Man.


Heart Rhythm | 2012

Implantable cardioverter-defibrillator longevity under clinical circumstances: An analysis according to device type, generation, and manufacturer

Joep Thijssen; C. Jan Willem Borleffs; Johannes B. van Rees; Sumche Man; Mihály K. de Bie; Jeroen Venlet; Enno T. van der Velde; Lieselot van Erven; Martin J. Schalij

BACKGROUND One of the major drawbacks of implantable cardioverter-defibrillator (ICD) treatment is the limited device service life. Thus far, data concerning ICD longevity under clinical circumstances are scarce. In this study, the ICD service life was assessed in a large cohort of ICD recipients. OBJECTIVE To assess the battery longevity of ICDs under clinical circumstances. METHODS All patients receiving an ICD in the Leiden University Medical Center were included in the analysis. During prospectively recorded follow-up visits, reasons for ICD replacement were assessed and categorized as battery depletion and non-battery depletion. Device longevity and battery longevity were calculated. The impact of device type, generation, manufacturer, the percentage of pacing, the pacing output, and the number of shocks on the battery longevity was assessed. RESULTS Since 1996, 4673 ICDs were implanted, of which 1479 ICDs (33%) were replaced. Mean device longevity was 5.0 ± 0.1 years. A total of 1072 (72%) ICDs were replaced because of battery depletion. Mean battery longevity of an ICD was 5.5 ± 0.1 years. When divided into different types, mean battery longevity was 5.5 ± 0.2 years for single-chamber ICDs, 5.8 ± 0.1 for dual-chamber ICDs, and 4.7 ± 0.1 years for cardiac resynchronization therapy-defibrillators (P <.001). Devices implanted after 2002 had a significantly better battery longevity as compared with devices implanted before 2002 (5.6 ± 0.1 years vs 4.9 ± 0.2 years; P <.001). In addition, large differences in battery longevity between manufacturers were noted (overall log-rank test, P <.001). CONCLUSIONS The majority of ICDs were replaced because of battery depletion. Large differences in longevity exist between different ICD types and manufacturers. Modern ICD generations demonstrated improved longevity.


Journal of Electrocardiology | 2014

Normal values of the electrocardiogram for ages 16-90 years

Peter R. Rijnbeek; Gerard van Herpen; Michiel L. Bots; Sumche Man; Niek Verweij; Albert Hofman; Hans L. Hillege; Matthijs E. Numans; Cees A. Swenne; Jacqueline C. M. Witteman; Jan A. Kors

INTRODUCTION To establish an up-to-date and comprehensive set of normal values for the clinically current measurements in the adult ECG, covering all ages for both sexes. METHODS The study population included 13,354 individuals, taken from four population studies in The Netherlands, ranging in age from 16 to 90 years (55% men) and cardiologically healthy by commonly accepted criteria. Standard 12-lead ECGs were available for all participants. The ECGs were processed by a well-validated computer program. Normal limits were taken as the 2nd and 98th percentiles of the measurement distribution per age group. RESULTS Our study corroborates many findings of previous studies, but also provides more differentiated results, in particular for the older age groups. Age trends were apparent for the QTc interval, QRS axis, and indices of left ventricular hypertrophy. Amplitudes in the left precordial leads showed a substantial increase in the older age groups for women, but not for men. Sex-dependent differences were apparent for most ECG parameters. All results are available on the Website www.normalecg.org, both in tabular and in graphical format. CONCLUSIONS We determined age- and sex-dependent normal values of the adult ECG. Our study distinguishes itself from other studies by the large size of the study population, comprising both sexes, the broad range of ages, and the exhaustive set of measurements. Our results emphasize that most diagnostic ECG criteria should be age- and sex-specific.


Journal of Electrocardiology | 2010

The spatial QRS-T angle in the Frank vectorcardiogram: accuracy of estimates derived from the 12-lead electrocardiogram

Charlotte Schreurs; Annemijn M. Algra; Sumche Man; Suzanne C. Cannegieter; Ernst E. van der Wall; Martin J. Schalij; Jan A. Kors; Cees A. Swenne

BACKGROUND AND PURPOSE The spatial QRS-T angle (SA), a predictor of sudden cardiac death, is a vectorcardiographic variable. Gold standard vertorcardiograms (VCGs) are recorded by using the Frank electrode positions. However, with the commonly available 12-lead ECG, VCGs must be synthesized by matrix multiplication (inverse Dower matrix/Kors matrix). Alternatively, Rautaharju proposed a method to calculate SA directly from the 12-lead ECG. Neither spatial angles computed by using the inverse Dower matrix (SA-D) nor by using the Kors matrix (SA-K) or by using Rautaharjus method (SA-R) have been validated with regard to the spatial angles as directly measured in the Frank VCG (SA-F). Our present study aimed to perform this essential validation. METHODS We analyzed SAs in 1220 simultaneously recorded 12-lead ECGs and VCGs, in all data, in SA-F-based tertiles, and after stratification according to pathology or sex. RESULTS Linear regression of SA-K, SA-D, and SA-R on SA-F yielded offsets of 0.01 degree, 20.3 degrees, and 28.3 degrees and slopes of 0.96, 0.86, and 0.79, respectively. The bias of SA-K with respect to SA-F (mean +/- SD, -3.2 degrees +/- 13.9 degrees) was significantly (P < .001) smaller than the bias of both SA-D and SA-R with respect to SA-F (8.0 degrees +/- 18.6 degrees and 9.8 degrees +/- 24.6 degrees, respectively); tertile analysis showed a much more homogeneous behavior of the bias in SA-K than of both the bias in SA-D and in SA-R. In pathologic ECGs, there was no significant bias in SA-K; bias in men and women did not differ. CONCLUSION SA-K resembled SA-F best. In general, when there is no specific reason either to synthesize VCGs with the inverse Dower matrix or to calculate the spatial QRS-T angle with Rautaharjus method, it seems prudent to use the Kors matrix.


Journal of Electrocardiology | 2008

Reconstruction of standard 12-lead electrocardiograms from 12-lead electrocardiograms recorded with the Mason-Likar electrode configuration.

Sumche Man; Arie C. Maan; Eunhyo Kim; Harmen H.M. Draisma; Martin J. Schalij; Ernst E. van der Wall; Cees A. Swenne

Electrocardiograms (ECGs) made with Mason-Likar electrode configuration (ML-ECGs) show well-known differences from standard 12-lead ECGs (Std-ECGs). We recorded, simultaneously, Std-ECGs and ML-ECGs in 180 subjects. Using these ECGs, 8 x 8 individual and general conversion matrices were created by linear regression, and standard ECGs were reconstructed from ML-ECGs using these matrices. The performance of the matrices was assessed by the root mean square differences between the original Std-ECGs and the reconstructed standard ECGs, by the differences in major ECG parameters, and by comparison of computer-generated diagnostic statements. As a result, we conclude that, based on the root mean square differences, reconstructions with 8 x 8 individual matrices perform significantly better than reconstructions with the group matrix and perform equally well with respect to the calculation of major electrocardiographic parameters, which gives an improved reliability of the QRS frontal axis and the maximal QRS and T amplitudes. Both types of matrices were able to reverse the underdiagnosis of inferior myocardial infarctions and the erroneous statements about the QRS frontal axis that arose in the ECGs that were made by using the Mason-Likar electrode positions.


Journal of Electrocardiology | 2011

Influence of the vectorcardiogram synthesis matrix on the power of the electrocardiogram-derived spatial QRS-T angle to predict arrhythmias in patients with ischemic heart disease and systolic left ventricular dysfunction.

Sumche Man; Annemijn M. Algra; Charlotte Schreurs; C. Jan Willem Borleffs; Roderick W.C. Scherptong; Lieselot van Erven; Ernst E. van der Wall; Suzanne C. Cannegieter; Martin J. Schalij; Cees A. Swenne

BACKGROUND AND PURPOSE Several studies have demonstrated that the spatial mean QRS-T angle (SA) predicts cardiac events and mortality. Spatial mean QRS-T angle is a vectorcardiographic variable. Because in clinical practice, 12-lead standard electrocardiograms (ECGs) are recorded rather than vectorcardiograms (VCGs) according to Frank, VCGs are commonly obtained by synthesizing them from 12-lead ECGs, by using a VCG synthesis matrix. Hence, the thus computed SA is an estimate of the real SA measured in the Frank VCG. Recent studies have shown that Kors VCG synthesis matrix yields better estimates of SA than the inverse Dower VCG synthesis matrix. Our current study aims to compare the predictive power of these SA variants for the occurrence of potentially lethal arrhythmias. METHODS The study group consisted of patients with ischemic heart disease and left ventricular systolic dysfunction who received an implantable cardioverter-defibrillator (ICD) for primary prevention. During follow-up, the occurrence of appropriate device therapy (occurrence of ventricular arrhythmia) was noted. Alternative SAs were computed in VCGs synthesized from standard 12-lead ECGs by using either the inverse Dower matrix (SA-Dower) or the Kors matrix (SA-Kors). Comparison of the predictive power of SA-Dower and SA- Kors was performed by receiver operating characteristic analysis, by Kaplan-Meier analysis, and by univariate and multivariate Cox regression analysis, using every 10th percentile of SA as a cutoff value. RESULTS The study group consisted of 412 patients (361 men; mean ± SD age 63 ± 11 years), in which 56 patients had appropriate ICD therapy during follow-up. Receiver operating characteristic analysis revealed that the area under the curve of SA-Kors was significantly larger than area under the curve of SA-Dower (0.646 vs 0.607, P = .043). The discriminative power of SA-Kors for the absence/presence of appropriate ICD therapy in patients during follow-up was generally superior to SA-Dower over a wide range of cutoff values in the Kaplan-Meier analysis and generally yielded stronger hazard ratios in the univariate and multivariate Cox regression analyses. CONCLUSION If there is no specific reason to use the inverse Dower matrix, VCG synthesis from standard 12-lead ECGs should preferably be done by using the Kors matrix. It is likely to assume that already published studies in which the predictive value of SA-Dower was demonstrated would yield stronger results if the SA-Dower angles were substituted by SA-Kors angles.


Annals of Noninvasive Electrocardiology | 2012

Comparison of standard versus orthogonal ECG leads for T-wave alternans identification.

Laura Burattini; Sumche Man; Roberto Burattini; Cees A. Swenne

T‐wave alternans (TWA), an electrophysiologic phenomenon associated with ventricular arrhythmias, is usually detected from selected ECG leads. TWA amplitude measured in the 12‐standard and the 3‐orthogonal (vectorcardiographic) leads were compared here to identify which lead system yields a more adequate detection of TWA as a noninvasive marker for cardiac vulnerability to ventricular arrhythmias. Our adaptive match filter (AMF) was applied to exercise ECG tracings from 58 patients with an implanted cardiac defibrillator, 29 of which had ventricular tachycardia or fibrillation during follow‐up (cases), while the remaining 29 were used as controls. Two kinds of TWA indexes were considered, the single‐lead indexes, defined as the mean TWA amplitude over each lead (MTWAA), and lead‐system indexes, defined as the mean and the maximum MTWAA values over the standard leads and over the orthogonal leads. Significantly (P < 0.05) higher TWA in the cases versus controls was identified only occasionally by the single‐lead indexes (odds ratio: 1.0–9.9, sensitivity: 24–76%, specificity: 76–86%), and consistently by the lead‐system indexes (odds ratio: 4.5–8.3, sensitivity: 57–72%, specificity: 76%). The latter indexes also showed a significant correlation (0.65–0.83) between standard and orthogonal leads. Hence, when using the AMF, TWA should be detected in all leads of a system to compute the lead‐system indexes, which provide a more reliable TWA identification than single‐lead indexes, and a better discrimination of patients at increased risk of cardiac instability. The standard and the orthogonal leads can be considered equivalent for TWA identification, so that TWA analysis can be limited to one‐lead system.


Journal of Electrocardiology | 2011

Predictive power of T-wave alternans and of ventricular gradient hysteresis for the occurrence of ventricular arrhythmias in primary prevention cardioverter-defibrillator patients

Sumche Man; Priscilla V. De Winter; Arie C. Maan; Joep Thijssen; C. Jan Willem Borleffs; Wilbert Pm Van Meerwijk; Marianne Bootsma; Lieselot van Erven; Ernst E. van der Wall; Martin J. Schalij; Laura Burattini; Roberto Burattini; Cees A. Swenne

BACKGROUND AND PURPOSE Left ventricular ejection fraction lacks specificity to predict sudden cardiac death in heart failure. T-wave alternans (TWA; beat-to-beat T-wave instability, often measured during exercise) is deemed a promising noninvasive predictor of major cardiac arrhythmic event. Recently, it was demonstrated that TWA during recovery from exercise has additional predictive value. Another mechanism that potentially contributes to arrhythmogeneity is exercise-recovery hysteresis in action potential morphology distribution, which becomes apparent in the spatial ventricular gradient (SVG). In the current study, we investigated the performance of TWA amplitude (TWAA) during a complete exercise test and of exercise-recovery SVG hysteresis (SVGH) as predictors for lethal arrhythmias in a population of heart failure patients with cardioverter-defibrillators (ICDs) implanted for primary prevention. METHODS We performed a case-control study with 34 primary prevention ICD patients, wherein 17 patients (cases) and 17 patients (controls) had no ventricular arrhythmia during follow-up. We computed, in electrocardiograms recorded during exercise tests, TWAA (maximum over the complete test) and the exercise-recovery hysteresis in the SVG. Statistical analyses were done by using the Student t test, Spearman rank correlation analysis, receiver operating characteristics analysis, and Kaplan-Meier analysis. Significant level was set at 5%. RESULTS Both SVGH and TWAA differed significantly (P < .05) between cases (mean ± SD, SVGH: -18% ± 26%, TWAA: 80 ± 46 μV) and controls (SVGH: 5% ± 26%, TWAA: 49 ± 20 μV). Values of TWAA and SVGH showed no significant correlation in cases (r = -0.16, P = .56) and in controls (r = -0.28, P = .27). Receiver operating characteristics of SVGH (area under the curve = 0.734, P = .020) revealed that SVGH less than 14.8% discriminated cases and controls with 94.1% sensitivity and 41.2% specificity; hazard ratio was 3.34 (1.17-9.55). Receiver operating characteristics of TWA (area under the curve = 0.699, P = .048) revealed that TWAA greater than 32.5 μV discriminated cases and controls with 93.8% sensitivity and 23.5% specificity; hazard ratio was 2.07 (0.54-7.91). DISCUSSION AND CONCLUSION Spatial ventricular gradient hysteresis bears predictive potential for arrhythmias in heart failure patients with an ICD for primary prevention, whereas TWA analysis seems to have lesser predictive value in our pilot group. Spatial ventricular gradient hysteresis is relatively robust for noise, and, as it rests on different electrophysiologic properties than TWA, it may convey additional information. Hence, joint analysis of TWA and SVGH may, possibly, improve the noninvasive identification of high-risk patients. Further research, in a large group of patients, is required and currently carried out by our group.


Journal of Electrocardiology | 2014

Acute coronary syndrome with a totally occluded culprit artery: relation of the ST injury vector with ST-elevation and non-ST elevation ECGs

Sumche Man; Chinar Rahmattulla; Arie C. Maan; Niek H J J van der Putten; W. Arnold Dijk; Erik W. van Zwet; Ernst E. van der Wall; Martin J. Schalij; Anton P.M. Gorgels; Cees A. Swenne

BACKGROUND In acute coronary syndrome (ACS), ST-segment elevation (STE), often associated with a completely occluded culprit artery, is an important ECG criterion for primary percutaneous coronary intervention (PCI). However, several studies showed that in ACS a completely occluded culprit artery can also occur with a non-ST-elevation (NSTE) ECG. In order to elucidate reasons for this discrepancy we examined ST injury vector orientation and magnitude in ACS patients with and without STE, all admitted for primary PCI and having a completely occluded culprit artery. METHODS We studied the ECGs of 300 ACS patients (214/86 STE/NSTE; 228/72 single/multivessel disease) who had a completely occluded culprit artery during angiography prior to primary PCI. The J+60 injury vector orientation and magnitude were computed from Frank XYZ leads derived from the 10-s standard 12-lead ECG. RESULTS Demographic and anthropomorphic characteristics of the STE and NSTE patients did not differ. STE patients had a higher rate of right coronary artery occlusions, and a lower rate of left circumflex occlusions than NSTE patients (43 vs. 31%, and 13 vs. 22%, respectively; P<0.05). Injury vector elevation and magnitude were larger in STE than in NSTE patients (32° ± 37° vs. 6° ± 39°, and 304 ± 145 μV vs. 134 ± 72 μV, respectively; P<0.0001). CONCLUSION STE criteria favor certain injury vector directions and larger injury vector magnitudes. Obviously, several ACS patients with complete culprit artery occlusions requiring primary PCI do not fulfill these criteria. Our study suggests that STE-NSTE-based ACS stratification needs further enhancement.


Journal of Electrocardiology | 2015

Comparison of model-based and expert-rule based electrocardiographic identification of the culprit artery in patients with acute coronary syndrome

Vivian P. Kamphuis; Galen S. Wagner; Olle Pahlm; Sumche Man; Charles W. Olson; Ljuba Bacharova; Cees A. Swenne

BACKGROUND AND PURPOSE Culprit coronary artery assessment in the triage ECG of patients with suspected acute coronary syndrome (ACS) is relevant a priori knowledge preceding percutaneous coronary intervention (PCI). We compared a model-based automated method (Olson method) with an expert-rule based method for the culprit artery assessment. METHODS In each of the 53 patients who were admitted with the working diagnosis of suspected ACS, scheduled for emergent angiography with a view on revascularization as initial treatment and subsequently found to have an angiographically documented completely occluded culprit artery, culprit artery location was assessed in the preceding ECG by both the model-based Olson method and the expert-rule based method that considered either visual or computer-measured J-point amplitudes. ECG culprit artery estimations were compared with the angiographic culprit lesion locations. Proportions of correct classifications were compared by a Z test at the 5% significance level. RESULTS The Olson method performed slightly, but not significantly, better, when the expert-rule based method used visual assessment of J-point amplitudes (88.7% versus 81.1% correct; P=0.28). However, the Olson method performed significantly better when the expert-rule based method used computer-measured J-point amplitudes (88.7% versus 71.7% correct; P<0.05). CONCLUSION The automated model-based Olson method performed at least at the level of expert cardiologists using a manual rule-based method.


American Journal of Physiology-heart and Circulatory Physiology | 2009

Biventricular pacing-induced acute response in baroreflex sensitivity has predictive value for midterm response to cardiac resynchronization therapy

Maaike G. J. Gademan; Rutger J. van Bommel; C. Jan Willem Borleffs; Sumche Man; Joris C. W. Haest; Martin J. Schalij; Ernst E. van der Wall; Jeroen J. Bax; Cees A. Swenne

In a previous study we demonstrated that the institution of biventricular pacing in chronic heart failure (CHF) acutely facilitates the arterial baroreflex. The arterial baroreflex has important prognostic value in CHF. We hypothesized that the acute response in baroreflex sensitivity (BRS) after the institution of cardiac resynchronization therapy (CRT) has predictive value for midterm response. One day after implantation of a CRT device in 33 CHF patients (27 male/6 female; age, 66.5 +/- 9.5 yr; left ventricular ejection fraction, 28 +/- 7%) we measured noninvasive BRS and heart rate variability (HRV) in two conditions: CRT device switched on and switched off (on/off order randomized). Echocardiography was performed before implantation (baseline) and 6 mo after implantation (follow-up). CRT responders were defined as patients in whom left ventricular end-systolic volume at follow-up had decreased by > or =15%. Responders (69.7%) and nonresponders (30.3%) had similar baseline characteristics. In responders, CRT increased BRS by 30% (P = 0.03); this differed significantly (P = 0.02) from the average BRS change (-2%) in the nonresponders. CRT also increased HRV by 30% in responders (P = 0.02), but there was no significant difference found compared with the increase in HRV (8%) in the nonresponders. Receiver-operating characteristic curve analysis revealed that the percent BRS increase had predictive value for the discrimination of responders and nonresponders (area under the curve, 0.69; 95% confidence interval, 0.51-0.87; maximal accuracy, 0.70). Our study demonstrates that a CRT-induced acute BRS increase has predictive value for the echocardiographic response to CRT. This finding suggests that the autonomic nervous system is actively involved in CRT-related reverse remodeling.

Collaboration


Dive into the Sumche Man's collaboration.

Top Co-Authors

Avatar

Cees A. Swenne

Leiden University Medical Center

View shared research outputs
Top Co-Authors

Avatar

Martin J. Schalij

Leiden University Medical Center

View shared research outputs
Top Co-Authors

Avatar

Arie C. Maan

Leiden University Medical Center

View shared research outputs
Top Co-Authors

Avatar

Laura Burattini

Marche Polytechnic University

View shared research outputs
Top Co-Authors

Avatar

Ernst E. van der Wall

Leiden University Medical Center

View shared research outputs
Top Co-Authors

Avatar

C. Cato ter Haar

Leiden University Medical Center

View shared research outputs
Top Co-Authors

Avatar

C. Jan Willem Borleffs

Leiden University Medical Center

View shared research outputs
Top Co-Authors

Avatar

Lieselot van Erven

Leiden University Medical Center

View shared research outputs
Top Co-Authors

Avatar

Marjolein C. De Jongh

Leiden University Medical Center

View shared research outputs
Top Co-Authors

Avatar

Sandro Fioretti

Marche Polytechnic University

View shared research outputs
Researchain Logo
Decentralizing Knowledge