Network


Latest external collaboration on country level. Dive into details by clicking on the dots.

Hotspot


Dive into the research topics where Yusuke Chikata is active.

Publication


Featured researches published by Yusuke Chikata.


Respiratory Care | 2014

Humidification Performance of Two High-Flow Nasal Cannula Devices: A Bench Study

Yusuke Chikata; Masayo Izawa; Nao Okuda; Taiga Itagaki; Emiko Nakataki; Mutsuo Onodera; Hideaki Imanaka; Masaji Nishimura

INTRODUCTION: Delivering heated and humidified medical gas at 20–60 L/min, high-flow nasal cannula (HFNC) creates low levels of PEEP and ameliorates respiratory mechanics. It has become a common therapy for patients with respiratory failure. However, independent measurement of heat and humidity during HFNC and comparison of HFNC devices are lacking. METHODS: We evaluated 2 HFNC (Airvo 2 and Optiflow system) devices. Each HFNC was connected to simulated external nares using the manufacturers standard circuit. The Airvo 2 outlet-chamber temperature was set at 37°C. The Optiflow system incorporated an O2/air blender and a heated humidifier, which was set at 40°C/−3. For both systems, HFNC flow was tested at 20, 40, and 50 L/min. Simulating spontaneous breathing using a mechanical ventilator and TTL test lung, we tested tidal volumes (VT) of 300, 500, and 700 mL, and breathing frequencies of 10 and 20 breaths/min. The TTL was connected to the simulated external nares with a standard ventilator circuit. To prevent condensation, the circuit was placed in an incubator maintained at 37°C. Small, medium, and large nasal prongs were tested. Absolute humidity (AH) of inspired gas was measured at the simulated external nares. RESULTS: At 20, 40, and 50 L/min of flow, respective AH values for the Airvo 2 were 35.3 ± 2.0, 37.1 ± 2.2, and 37.6 ± 2.1 mg/L, and for the Optiflow system, 33.1 ± 1.5, 35.9 ± 1.7, and 36.2 ± 1.8 mg/L. AH was lower at 20 L/min of HFNC flow than at 40 and 50 L/min (P < .01). While AH remained constant at 40 and 50 L/min, at 20 L/min of HFNC flow, AH decreased as VT increased for both devices. CONCLUSIONS: During bench use of HFNC, AH increased with increasing HFNC flow. When the inspiratory flow of spontaneous breathing exceeded the HFNC flow, AH was influenced by VT. At all experimental settings, AH remained > 30 mg/L.


Respiratory Care | 2013

Humidification Performance of Humidifying Devices for Tracheostomized Patients With Spontaneous Breathing: A Bench Study

Yusuke Chikata; Jun Oto; Mutsuo Onodera; Masaji Nishimura

BACKGROUND: Heat and moisture exchangers (HMEs) are commonly used for humidifying respiratory gases administered to mechanically ventilated patients. While they are also applied to tracheostomized patients with spontaneous breathing, their performance in this role has not yet been clarified. We carried out a bench study to investigate the effects of spontaneous breathing parameters and oxygen flow on the humidification performance of 11 HMEs. METHODS: We evaluated the humidification provided by 11 HMEs for tracheostomized patients, and also by a system delivering high-flow CPAP, and an oxygen mask with nebulizer heater. Spontaneous breathing was simulated with a mechanical ventilator, lung model, and servo-controlled heated humidifier at tidal volumes of 300, 500, and 700 mL, and breathing frequencies of 10 and 20 breaths/min. Expired gas was warmed to 37°C. The high-flow CPAP system was set to deliver 15, 30, and 45 L/min. With the 8 HMEs that were equipped with ports to deliver oxygen, and with the high-flow CPAP system, measurements were taken when delivering 0 and 3 L/min of dry oxygen. After stabilization we measured the absolute humidity (AH) of inspired gas with a hygrometer. RESULTS: AH differed among HMEs applied to tracheostomized patients with spontaneous breathing. For all the HMEs, as tidal volume increased, AH decreased. At 20 breaths/min, AH was higher than at 10 breaths/min. For all the HMEs, when oxygen was delivered, AH decreased to below 30 mg/L. With an oxygen mask and high-flow CPAP, at all settings, AH exceeded 30 mg/L. CONCLUSIONS: None of the HMEs provided adequate humidification when supplemental oxygen was added. In the ICU, caution is required when applying HME to tracheostomized patients with spontaneous breathing, especially when supplemental oxygen is required.


Artificial Organs | 2015

Re-evaluation of Pre-pump Arterial Pressure to Avoid Inadequate Dialysis and Hemolysis: Importance of Prepump Arterial Pressure Monitoring in Hemodialysis Patients.

Eriko Shibata; Kojiro Nagai; Risa Takeuchi; Yasuhiro Noda; Tomomi Makino; Yusuke Chikata; Michael Hann; Sakiya Yoshimoto; Hiroyuki Ono; Sayo Ueda; Masanori Tamaki; Taichi Murakami; Motokazu Matsuura; Hideharu Abe; Toshio Doi

Prepump arterial pressure (PreAP) is monitored to avoid generating excessive negative pressure. The National Kidney Foundation K/DOQI clinical practice guidelines for vascular access recommend that PreAP should not fall below -250 mm Hg because excessive negative PreAP can lead to a decrease in the delivery of blood flow, inadequate dialysis, and hemolysis. Nonetheless, these recommendations are consistently disregarded in clinical practice and pressure sensors are often removed from the dialysis circuit. Thus far, delivered blood flow has been reported to decrease at values more negative than -150 mm Hg of PreAP. These values have been analyzed by an ultrasonic flowmeter and not directly measured. Furthermore, no known group has evaluated whether PreAP-induced hemolysis occurs at a particular threshold. Therefore, the aim of this study was to clarify the importance of PreAP in the prediction of inadequate dialysis and hemolysis. By using different diameter needles, human blood samples from healthy volunteers were circulated in a closed dialysis circuit. The relationship between PreAP and delivered blood flow or PreAP and hemolysis was investigated. We also investigated the optimal value for PreAP using several empirical monitoring methods, such as a pressure pillow. Our investigation indicated that PreAP is a critical factor in the determination of delivered blood flow and hemolysis, both of which occured at pressure values more negative than -150 mm Hg. With the exception of direct pressure monitoring, commonly used monitoring methods for PreAP were determined to be ineffective. We propose that the use of a vacuum monitor would permit regular measurement of PreAP.


Pediatric Anesthesia | 2009

Humidification during high-frequency oscillation ventilation is affected by ventilator circuit and ventilatory setting

Yusuke Chikata; Hideaki Imanaka; Yoshiaki Onishi; Masahiko Ueta; Masaji Nishimura

Background:  High‐frequency oscillation ventilation (HFOV) is an accepted ventilatory mode for acute respiratory failure in neonates. As conventional mechanical ventilation, inspiratory gas humidification is essential. However, humidification during HFOV has not been clarified. In this bench study, we evaluated humidification during HFOV in the open circumstance of ICU. Our hypothesis is that humidification during HFOV is affected by circuit design and ventilatory settings.


Respiratory Care | 2016

Inspiratory Tube Condensation During High-Flow Nasal Cannula Therapy: A Bench Study

Yusuke Chikata; Kazuaki Unai; Masayo Izawa; Nao Okuda; Jun Oto; Masaji Nishimura

BACKGROUND: High-flow nasal cannula (HFNC) therapy provides better humidification than conventional oxygen therapy. To allay loss of vapor as condensation, a servo-controlled heating wire is incorporated in the inspiratory tube, but condensation is not completely avoidable. We investigated factors that might affect condensation: thermal characteristics of the inspiratory tube, HFNC flow, and ambient temperature. METHODS: We evaluated 2 types of HFNC tubes, SLH Flex 22-mm single tube and RT202. Both tubes were connected to a heated humidifier with water reservoir. HFNC flow was set at 20, 40, and 60 L/min, and FIO2 was set at 0.21. Air conditioning was used maintain ambient temperature at close to either 20 or 25°C. We weighed the tubes on a digital scale before (0 h) and at 3, 6, and 24 h after, turning on the heated humidifier, and calculated the amount of condensation by simple subtraction. The amount of distilled water used during 24 h was also recorded. RESULTS: At 25°C, there was little condensation, but at 20°C and HFNC flow of 20, 40, and 60 L/min for 24 h, the amount of condensation with the SLH was 50.2 ± 10.7, 44.3 ± 17.7, and 56.6 ± 13.9 mg, and the amount with the RT202 was 96.0 ± 35.1, 72.8 ± 8.2, and 64.9 ± 0.8 mg. When ambient temperature was set to 20°C, condensation with the RT202 was statistically significantly greater than with the SLH at all flow settings (P < .001). CONCLUSIONS: Ambient temperature statistically significantly influenced the amount of condensation in the tubes.


Journal of intensive care | 2013

Temperature of gas delivered from ventilators

Yusuke Chikata; Mutsuo Onodera; Hideaki Imanaka; Masaji Nishimura

BackgroundAlthough heated humidifiers (HHs) are the most efficient humidifying device for mechanical ventilation, some HHs do not provide sufficient humidification when the inlet temperature to the water chamber is high. Because portable and home-care ventilators use turbines, blowers, pistons, or compressors to inhale in ambient air, they may have higher gas temperature than ventilators with piping systems. We carried out a bench study to investigate the temperature of gas delivered from portable and home-care ventilators, including the effects of distance from ventilator outlet, fraction of inspiratory oxygen (FIO2), and minute volume (MV).MethodsWe evaluated five ventilators equipped with turbine, blower, piston, or compressor system. Ambient air temperature was adjusted to 24°C ± 0.5°C, and ventilation was set at FIO2 0.21, 0.6, and 1.0, at MV 5 and 10 L/min. We analyzed gas temperature at 0, 40, 80, and 120 cm from ventilator outlet and altered ventilator settings.ResultsWhile temperature varied according to ventilators, the outlet gas temperature of ventilators became stable after, at the most, 5 h. Gas temperature was 34.3°C ± 3.9°C at the ventilator outlet, 29.5°C ± 2.2°C after 40 cm, 25.4°C ± 1.2°C after 80 cm and 25.1°C ± 1.2°C after 120 cm (P < 0.01). FIO2 and MV did not affect gas temperature.ConclusionGas delivered from portable and home-care ventilator was not too hot to induce heated humidifier malfunctioning. Gas soon declined when passing through the limb.


Critical Care Research and Practice | 2012

Humidification Performance of Heat and Moisture Exchangers for Pediatric Use

Yusuke Chikata; Chihiro Sumida; Jun Oto; Hideaki Imanaka; Masaji Nishimura

Background. While heat and moisture exchangers (HMEs) have been increasingly used for humidification during mechanical ventilation, the efficacy of pediatric HMEs has not yet been fully evaluated. Methods. We tested ten pediatric HMEs when mechanically ventilating a model lung at respiratory rates of 20 and 30 breaths/min and pressure control of 10, 15, and 20 cmH2O. The expiratory gas passed through a heated humidifier. We created two rates of leakage: 3.2 L/min (small) and 5.1 L/min (large) when pressure was 10 cmH2O. We measured absolute humidity (AH) at the Y-piece. Results. Without leakage, eight of ten HMEs maintained AH at more than 30 mg/L. With the small leak, AH decreased below 30 mg/L (26.6 to 29.5 mg/L), decreasing further (19.7 to 27.3 mg/L) with the large leak. Respiratory rate and pressure control level did not affect AH values. Conclusions. Pediatric HMEs provide adequate humidification performance when leakage is absent.


Respiratory Care | 2017

FIO2 in an Adult Model Simulating High-Flow Nasal Cannula Therapy.

Yusuke Chikata; Mutsuo Onodera; Jun Oto; Masaji Nishimura

BACKGROUND: High-flow nasal cannula therapy (HFNC) is widely used for patients with acute respiratory failure. HFNC has a number of physiological effects. Although FIO2 is considered to be constant, because HFNC is an open system, FIO2 varies according to inspiratory flow, tidal volume (VT), and HFNC gas flow. We investigated the influence of HFNC gas flow and other respiratory parameters on FIO2 during HFNC. METHODS: We evaluated an HFNC system and, for comparison, a conventional oxygen therapy system. The HFNC apparatus was composed of an air/oxygen blender, a heated humidifier, an inspiratory limb, and small, medium, and large nasal prongs. HFNC gas flow was set at 20, 40, and 60 L/min, and FIO2 was set at 0.3, 0.5, and 0.7. We measured FIO2 for 1-min intervals using an oxygen analyzer and extracted data for the final 3 breaths of each interval. Spontaneous breathing was simulated using a mechanical ventilator connected to the muscle compartment of a model lung. The lung compartment passively moved with the muscle compartment, thus inspiring ambient air via a ventilator limb. With a decelerating flow waveform, simulated VT was set at 300, 500, and 700 mL, breathing frequency at 10 and 20 breaths/min, and inspiratory time at 1.0 s. RESULTS: With HFNC gas flow of 20 and 40 L/min, at all set FIO2 values, inspiratory oxygen concentration varied with VT (P < .001). As the set value for FIO2 increased, the difference between set FIO2 and measured FIO2 increased. Neither breathing frequency nor prong size influenced FIO2. CONCLUSIONS: During HFNC with simulated spontaneous breathing, when HFNC gas flow was 60 L/min, measured FIO2 was similar to set FIO2 at 0.3 and 0.5, whereas at 0.7, as VT increased, measured FIO2 decreased slightly. However, at 20 or 40 L/min, changes in VT related with deviation from set FIO2.


Respiratory Care | 2017

Humidity and Inspired Oxygen Concentration During High-Flow Nasal Cannula Therapy in Neonatal and Infant Lung Models

Yusuke Chikata; Saki Ohnishi; Masaji Nishimura

BACKGROUND: High-flow nasal cannula therapy (HFNC) for neonate/infants can deliver up to 10 L/min of heated and humidified gas, and FIO2 can be adjusted to between 0.21 and 1.0. With adults, humidification and actual FIO2 are known to vary according to inspiratory and HFNC gas flow, tidal volume (VT), and ambient temperature. There have been few studies focused on humidification and FIO2 in HFNC settings for neonates/infants, so we performed a bench study to investigate the influence of gas flow, ambient temperature, and respiratory parameters on humidification and actual FIO2 in a neonate/infant simulation. METHODS: HFNC gas flow was set at 3, 5, and 7 L/min, and FIO2 was set at 0.3, 0.5, and 0.7. Spontaneous breathing was simulated using a 2-bellows-in-a-box model of a neonate lung. Tests were conducted with VT settings of 20, 30, and 40 mL and breathing frequencies of 20 and 30 breaths/min. Inspiratory time was 0.8 s with decelerating flow waveform. The HFNC tube was placed in an incubator, which was either set at 37°C or turned off. Absolute humidity (AH) and actual FIO2 were measured for 1 min using a hygrometer and an oxygen analyzer, and data for the final 3 breaths were extracted. RESULTS: At all settings, when the incubator was turned on, AH was greater than when it was turned off (P < .001). When the incubator was turned off, as gas flow increased, AH increased (P < .001); however, VT did not affect AH (P = .16). As gas flow increased, actual FIO2 more closely corresponded to set FIO2. When gas flow was 3 L/min, measured FIO2 decreased proportionally more at each FIO2 setting increment (P < .001). CONCLUSIONS: AH was affected by ambient temperature and HFNC gas flow. Actual FIO2 depended on VT when gas flow was 3 L/min.


Respiratory Care | 2015

Performance of Ventilators Compatible With Magnetic Resonance Imaging: A Bench Study

Yusuke Chikata; Nao Okuda; Masayo Izawa; Mutsuo Onodera; Masaji Nishimura

BACKGROUND: Magnetic resonance imaging (MRI) is indispensable for diagnosing brain and spinal cord abnormalities. Magnetic components cannot be used during MRI procedures; therefore, patient support equipment must use MRI-compatible materials. However, little is known of the performance of MRI-compatible ventilators. METHODS: At commonly used settings, we tested the delivered tidal volume (VT), FIO2, PEEP, and operation of the high-inspiratory-pressure-relief valves of 4 portable MRI-compatible ventilators (Pneupac VR1, ParaPAC 200DMRI, CAREvent MRI, iVent201) and one ICU ventilator (Servo-i). Each ventilator was set in volume control/continuous mandatory ventilation mode. Breathing frequency and VT were tested at 10 breaths/min and 300, 500, and 700 mL, respectively. The Pneupac VR1 has fixed VT and frequency combinations, so it was tested at VT = 300 mL and 20 breaths/min, VT = 500 mL and 12 breaths/min, and VT = 800 mL and 10 breaths/min. FIO2 was 0.6 and 1.0. At the air-mix setting, FIO2 was fixed at 0.5 with the Pneupac VR1, 0.45 with the ParaPAC 200DMRI, and 0.6 with the CAREvent MRI. PEEP was set at 5 and 10 cm H2O, and pressure relief was set at 30 and 40 cm H2O. RESULTS: VT error varied widely among ventilators (−28.1 to 25.5%). As VT increased, error decreased with the Pneupac VR1, ParaPAC 200DMRI, and CAREvent MRI (P < .05). FIO2 error ranged from −13.3 to 25.3% at 0.6 (or air mix). PEEP error varied among ventilators (−29.2 to 42.5%). Only the Servo-i maintained VT, FIO2, and PEEP at set levels. The pressure-relief valves worked in all ventilators. CONCLUSIONS: None of the MRI-compatible ventilators maintained VT, FIO2, and PEEP at set levels. Vital signs of patients with unstable respiratory mechanics should be monitored during transport and MRI.

Collaboration


Dive into the Yusuke Chikata's collaboration.

Top Co-Authors

Avatar
Top Co-Authors

Avatar
Top Co-Authors

Avatar
Top Co-Authors

Avatar

Jun Oto

University of Tokushima

View shared research outputs
Top Co-Authors

Avatar

Masayo Izawa

University of Tokushima

View shared research outputs
Top Co-Authors

Avatar

Nao Okuda

University of Tokushima

View shared research outputs
Top Co-Authors

Avatar

Kazuaki Unai

University of Tokushima

View shared research outputs
Top Co-Authors

Avatar
Top Co-Authors

Avatar
Top Co-Authors

Avatar
Researchain Logo
Decentralizing Knowledge